Command Groups T able 2-30: Mark Commands (cont.) Command Description MARK:TOT al? Returns how many marks are used MARK:USERLIST The command creates a single u ser mark on a waveform in the time domain. The arguments con…

Command Groups
Table 2-29: Horizontal Commands (cont.)
Command
Description
HORizontal:DIGital:SAMPLERate:
MAGnivu?
Returns the sample rate of the Magnivu digital acquisition
HORizontal:DIGital:SAMPLERate:
MAIn?
Returns the sample rate of the main digital acquisition
HORizontal:POSition This command specifies the horizontal position, in percent, that is used when delay is off
HORizontal:PREViewstate?
Returns the display system preview state
HORizontal:RECOrdlength This command specifies the record length.
HORizontal:SAMPLERate? Query to display the sample rate in seconds.
HORizontal:SCAle This command specifies the horizontal scale
Mark Command Group
Use the commands in the Mark Command Group to identify areas of the acquired
wave
form that warrant further investigation.
Table 2-30: Mark Commands
Command Description
MARK
Move to the next or previous mark on the w aveform or returns all learnable settings from
the mark commands
MARK:CREATE Creates a mark on a particular waveform or all waveforms in a column
MARK:DELEte
Deletes a mark on a particular waveform, all waveforms in a column, or all marks
MARK:FREE? Returns how many marks are free to be used
MARK:SAVEALL This command saves all current marks on waveforms in the time domain to an internal
memory location. (This is equivalent to pressing the “Save All Marks" button in the Search
button menu on the front panel.) In order to retrieve the information, use the query form of
MARK:USERLIST.
MARK:SELected:END? Returns the end of the selected mark, in terms of 0 to 100% of the waveform
MARK:SELected:FOCUS? Returns the focus of the selected mark, in terms of 0 to 100% of the waveform
MARK:SELected:
MARKSINCOLumn?
Returns how many marks are in the current zoom pixel column
MARK:SELected:OWNer? Returns the owner of the selected mark
MARK:SELected:SOURCe? Returns the source waveform of the selected mark
MARK:SELected:STARt? Returns the start of the selected mark, in terms of 0 to 100% of the waveform
MARK:SELected:STAT E? Returns the on or off state of the selected mark
MARK:SELected:ZOOm:POSition? Returns the position of the selected mark, in terms of 0 to 100% of the upper window
2-30 MDO4000/B/C, MSO/DPO4000B and MDO3000 Series Oscilloscopes Programmer Manual

Command Groups
Table 2-30: Mark Commands (cont.)
Command Description
MARK:TOTal?
Returns how many marks are used
MARK:USERLIST The command creates a single user mark on a waveform in the time domain. The arguments
consist of an enumeration specifying the source waveform, followed by 7 time mark
parameters. You can create up to 1,024 marks. To save all the marks to memory, use
the command
MARK:SAVEALL TOUSER.
Mask Command Group
Use the commands in the Mask Command Group to perform mask and limit
testing, useful for long-term signal monitoring, characterizing signals during
design, or testing on a production line. The commands simplify many tasks, such
as conducting pass/fail tests in search of glitches or other waveform anomalies,
comparing a signal against a known good signal, or against a standard or custom
mask,
counting hits against specific mask segments, and much more.
NOTE. In order to use Mask commands, you need to install an optional Limit and
mask test application module, Tektronix part number DPO4LMT. Also, ≥350 MHz
bandwidth models are recommended for telecommunication standards >55 Mb/s.
NOTE. To ensure that Mask commands are enabled, use the command
APPLication:TYPe LIMITMask.
There are three categories of masks available:
N
OTE. Standard mask support is not available for the MDO3000 series models.
Standard The instrument has a built-in set of pre-defined standard
telecommunications masks (see the
MASK:STANdard command for a complete
list). Although these masks are non-editable, they can be copied and used to
initialize user-defined custom masks.
Limit Test: Limit testing is used to compare a tested signal against a known good
or "golden" version of the same signal with user-defined vertical and horizontal
tolerances. These non-editable masks are created using a Reference waveform or
template. Use the
MASK:TEMPLate commands for limit testing.
Custom: (also referred to in this document as “user mask”). You can create
custom masks and tailor them to your own specific testing requirements. These
commands provide granular control over the points of the mask and offer a
variety of testing options. For instance, you can specify test duration either by
MDO4000/B/C, MSO/DPO4000B and MDO3000 Series Oscilloscopes Programmer Manual 2-31

Command Groups
number of wavef
orms or by time, or specify a violation threshold that must be
met before considering a test a failure, or specify an action (such as writing to a
file or sending an SRQ) that is to be performed upon a violation, a test failure,
or test completion. Use MASK:CUSTom andthenthe
:MASK:USER commands
to create user-defined custom masks.
A series of examples showing how to use mask commands for typical tasks
is included as an appendix. (See page E-1, Mask/Limit Command Sequence
Examples.)
Table 2-31
: Mask commands
Command
Descripti
on
APPLicat
ion:TYPe
When a mask/limit or power test application module is installed, one of the associated test
types is always selected by default. This command allows the test type to be changed
from the
default.
MASK:CO
Py:SOUrce
Specifie
s the standard mask that is to be copied to a user created custom mask.
MASK:C
OPy:USER
Copies
the specified standard mask to a user created custom mask.
MASK:C
OUNt
Resets
to zero the number of hits and failures for all mask segments.
MASK:
COUNt:FAILURES?
Retur
ns the number of pass/fail mask tests that have failed
MASK
:COUNt:HITS?
This
query returns the sum of all hits in all mask segments.
MASK
:COUNt:SEG<x>:HITS?
This
query returns the number of hits for the specified mask segment.
MAS
K:COUNt:TESTS?
Ret
urns the number of pass/fail tests that have occurred
MAS
K:COUNt:VIOLATIONS?
Ret
urns the number of test violations that have occurred i n the current pass/fail test.
MA
SK:COUNt:WAVEFORMS?
Re
turns the number of waveforms that have been acquired and processed during pass/fail
mask testing.
MASK:CUSTom Sets the user-defined custom mask to its initialized state, or copies the currently active
mask to the user-defined custom mask.
MASK:DISplay Controls whether a user-defined custom mask is displayed on the screen.
MASK:LOCk Locks the mask to the waveform so that any changes made to the horizontal and/or vertical
scale settings of the w aveform will redraw the mask segments in proportion.
MASK:MARgin:PERCent Sets or returns the tolerance for the mask test. A positive value expands the m ask and a
negative margin shrinks the m ask by the specified percentage.
MASK:SOUrce Specifies the mask source waveform to be used during pass/fail mask testing
MASK:STANdard Replaces the existing m ask, if any, with a specified standard mask.
MASK:STOPOnviolation Stops the waveform acquisitions upon the first occurrence of a waveform violation.
MASK:TEMPLate:CREATEmask Causes a template m ask to be created for limit testing, based on the s ettings of the
MASK:TEMPLate:SOUrce, MASK:TEMPLate:TOLerance:H ORizontal, and
MASK:TEMPLate:VERTical commands.
MASK:TEMPLate:SOUrce Specifies the source waveform to be used to create a m ask template for limit testing.
MASK:TEMPLate:TOLerance:
HORizontal
Specifies the +/- horizontal limit (tolerance) for a mask template to be used for limit testing.
MASK:TEMPLate:TOLerance:
VERTical
Specifies the +/- vertical limit (tolerance) for a mask template to be used for limit testing.
2-32 MDO4000/B/C, MSO/DPO4000B and MDO3000 Series Oscilloscopes Programmer Manual