MDO3000 Programmer Manual.pdf - 第182页

Commands Listed in A lphabetical Order Conditions A vailable for M SO/DPO4000B series models, 2-channel MDO3000 models, and MDO4000C models without option SA3 or SA6 only . Group V ertical Syntax AUXin:PRObe:C OMMAND &lt…

100%1 / 1158
Commands Listed in Alphabetical Order
AUXin:PRObe:
AUTOZero (No Query Form)
This command executes the attached probe’s Auto Zero function, for probes that
support this feature. See your probe documentation for more details.
Conditions
Available for MSO/DPO4000B series models, 2-channel MDO3000 models, and
MDO4000C models without option SA3 or SA6 only.
Group
Vertical
Syntax
AUXin:PRObe:AUTOZero EXECute
Arguments
EXECute sets the probe to autozero.
Examples
AUXin:PRObe:AUTOZero EXECute performs the AutoZero function in the
probe.
AUXin:PRObe:CALibrate:CALIBRATABLe? (Query Only)
This query indicates whethe
r the attached probe is calibratable. It returns a
boolean value.
Conditions
Available for MSO/DPO4000B series models, 2-channel MDO3000 models, and
MDO4000C models without option SA3 or SA6 only.
Group
Vertical
Syntax
AUXin:PRObe:CALibrate:CALIBRATABLe?
Examples
AUXin:PRObe:CALibrate:CALIBRATABLe? might return 0, indicating that the
probe attached to the Aux Input connector is not calibratable.
AUXin:PRObe:COMMAND (No Query Form)
Sets the state of the probe control specified with the first argument to the state
specified with the second argument. The commands and states are unique to the
attached probe type. Only certain VPI probes support this command. See the
probe documentation for how to set these string arguments.
MDO4000/B/C, MSO/DPO4000B and MDO3000 Series Oscilloscopes Programmer Manual 2-151
Commands Listed in Alphabetical Order
Conditions
Available for M
SO/DPO4000B series models, 2-channel MDO3000 models, and
MDO4000C models without option SA3 or SA6 only.
Group
Vertical
Syntax
AUXin:PRObe:COMMAND <QString>, <QS tring>
Arguments
<QString> are quoted strings specifying the probe command and value to set in
the probe attached to the Aux Input connector.
Examples
AUXIN:PROBE:COMMAND “OUTPUT”, “ON” turns the output of a Tektronix
VPI-DPG probe on.
AUXIN:PROBE:COMMAND “MODE”, “4–4V1MHz” sets a Tektronix VPI-DPG
probe to the 4-4V1MHz mode.
AUXIN:PROBE:COMMAND? might return AUXIN :PROBE:COMMAND
“MODE”,“4-4V1MHZ”
.
AUXin:PRObe:DEGAUss (No Query Form)
Starts a degauss/autozero cycle on a TekVPI current probe attached to the Aux
Input connector. If you send this command to a probe that does not support this
function, it is ignored.
Conditions
Available for MSO/DPO4000B series mod
els, 2-channel MDO3000 models, and
MDO4000C models without option SA3 or SA6 only.
Group
Vertical
Syntax
AUXin:PRObe:DEGAUss {EXECute}
Arguments
EXECute starts a probe degauss cycle.
Examples
AUXin:PROBE:DEGAUSS EXECUTE degausses the probe attached to the Aux
Input connector.
2-152 MDO4000/B/C, MSO/DPO4000B and MDO3000 Series Oscilloscopes Programmer Manual
Commands Listed in Alphabetical Order
AUXin:PRObe:
DEGAUss:STATE? (Query Only)
Returns the state of the probe degauss (NEEDED, RECOMMENDE D, PASSED,
FAILED, RUNNING). The command will return PASSED for probes that do not
support dega
uss operations.
Conditions
Available for MSO/DPO4000B series models, 2-channel MDO3000 models, and
MDO4000C m
odels without option SA3 or SA6 only.
Group
Vertical
Syntax
AUXin:PRObe:DEGAUss:STATE?
Example
s
AUXin:PROBE:DEGAUSS:STATE? might return:
AUXin:PROBE:DEGAUSS:STATE PASSED indicating that the probe has been
degaussed.
AUXin:PRObe:FORCEDRange
Changes or returns the range on a TekVPI probe attached to the A
ux Input
connector.
Conditions
Available for MSO/DPO4000B series models, 2-channel MDO30
00 models, and
MDO4000C models without option SA3 or SA6 only.
Group
Vertical
Syntax
AUXin:PRObe:FORCEDRange <NR3>
AUXin:PRObe:FORCEDRange?
Arguments
<NR3> is a floating point number that specifies the probe range, which is probe
dependent.
AUXin:PRObe:GAIN
Specifies the gain factor of a probe that is attached to the Aux Input connector.
Conditions
Available for MSO/DPO4000B series models, 2-channel MDO3000 models, and
MDO4000C models without option SA3 or SA6 only.
MDO4000/B/C, MSO/DPO4000B and MDO3000 Series Oscilloscopes Programmer Manual 2-153