Commands Listed in Alphabetical Order WFMInpre:YUNi t command (YUNit units). It uses the following formula (where dl means digitizing levels; curve_in_dl is a data point value in the CURVe data): V alue in YUNit units = …

Commands Listed in Alphabetical Order
WFMINPRE:YOFF
?
might return :WFMINPRE:YOFF 25 indicating the vertical
position of the incoming waveform is 25 digitizing levels.
WFMInpre:YUNit
This command specifies the vertical units of data points in the incoming waveform
record sent to the oscilloscope using the
CURVE command. This can be any of
several string values, depending upon the vertical units of the waveform being
sent.
Supported units are: %, /Hz, A, A/A, A/V, A/W, A/dB, A/s, AA, AW, AdB, As,
B, Hz, IRE
, S/s, V, V/A, V/V, V/W, V/dB, V/s, VV, VW, VdB, volts, Vs, W,
W/A, W/V, W/W, W/dB, W/s, WA, WV, WW, WdB, Ws, dB, dB/A, dB/V, dB/W,
dB/dB, dBA, dBV, dBW, dBdB, day, degrees, div, hr, min, ohms, percent, s
NOTE. (See page 2-95, Waveform Transfer Command Group.)
Group
Waveform Transfer
Syntax
WFMInpre:YUNit <QString>
WFMInpre:YUNit?
Related Commands
WFMOutpre:YUNit?
Arguments
<QString> contains the characters that represent the vertical units for the
incoming waveform.
Examples
WFMINPRE:YUNIT? might return :WFMINPRE:YUNIT "V" indicating the vertical
units for the incoming waveform are volts.
WFMINPRE:YUNIT "A" specifies that the vertical units for the incoming waveform
are Amperes.
WFMInpre:YZEro
This command specifies the vertical offset of the destination reference waveform
in units speci
fiedbythe
WFMInpre:YUNit command. Variations in this number
are analogous to changing the vertical offset of the waveform.
The
WFMInpre
:YMUlt
, WFMIn pre:YOFf, and WFMInpre :YZEro commands
are used to convert waveform record values to units specified using the
2-958 MDO4000/B/C, MSO/DPO4000B and MDO3000 Series Oscilloscopes Programmer Manual

Commands Listed in Alphabetical Order
WFMInpre:YUNi
t command (YUNit units). It uses the following formula (where
dl means digitizing levels; curve_in_dl is a data point value in the
CURVe data):
Value in YUNit units = ((curve_in_dl - YOFf) * YMUlt) + YZEro
NOTE. For a given waveform record, YMUlt, YOFf, and YZEro have to be a
consistent set, otherwise vertical cursor readouts and vertical measurements may
give incorrect results.
(See page 2-95, Waveform Transfer Command Group.)
Group
Waveform Transfer
Syntax
WFMInpre:YZEro <NR3>
WFMInpre:YZEro?
Related Commands
WFMInpre:YUNit
WFMInpre:YZEro
Arguments
<NR3> is a floating point number that specifies the offset in in units specified by
the
WFMInpre:YUNit command (YUNits).
Examples
WFMINPRE:YZERO 1.5E+0 specifies that the zero reference point for the
incoming waveform is 1.5 V below the center of the data range (given that
WFMInpre:YUNit is set to V).
WFMINPRE:YZERO? might return :WFMINPRE:YZEro 7.5000E-6 indicating
that the zero reference fo
r the incoming waveform is 7.5 µV below the center of
the data range (given that WFMInpre:YUNit is set to V).
WFMOutpre? (Query Only)
This query returns the information needed to interpret the waveform data points
returned by the
CURVe? query. It returns the waveform transmission and
formatting parameters for the waveform specified by the DATa:SOUrce command.
If the waveform specified by the DATa:SOUrce command is not displayed, the
oscilloscope returns only the waveform transmission parame
ters (BYT_Nr,
BIT_Nr, ENCdg, BN_Fmt, BYT_Or) and sets an error event indicating that the
source waveform is not turned on.
The Waveform Transfer command group text contains more comprehensive
infor
mation. (See page 2-95, Wa veform Transfer Command Group.)
(See page D-1, Wa veform Transfer (WFMOutpre and CURVe Query) Examples.)
MDO4000/B/C, MSO/DPO4000B and MDO3000 Series Oscilloscopes Programmer Manual 2-959

Commands Listed in Alphabetical Order
NOTE. For comma
nd sequence examples, see Appendix D. (See page D-1,
Waveform Transfer (WFMOutpre and CURVe Query) Examples.)
Example 1: Analog Waveform (channel 1 - 4)
Example 2: Digital Waveform (channel DO-D15)
Example 3: The Digital Collection with 4 Bytes Per Point with MagniVu Off
Example 4: The Digital Collection with 8 Bytes Per Point with MagniVu Off
Example 5: The Digital Collection with 4 Bytes Per Point with MagniVu On
Example 6: The Digital Collection with 8 Bytes Per Point with MagniVu On
Example 7: RF Frequency Domain Waveform
Group
Waveform Transfer
Syntax
WFMOutpre?
Examples
WFMOUTPRE? ? might return the waveform formatting data as:
:WFMOUTPRE:BYT_NR 2;BIT_NR 16;ENCD G ASCII;BN_FMT RI;BYT_OR
MSB;WFID “Ch1, DC coupling, 100.0mV/div, 4.000us/div, 10000
points, Sample mode”;NR_PT 10000;PT_FMT Y;XUNIT “s”;XINCR
4.0000E-9;XZERO - 2 0.0000E-6;PT_OFF 0;YUNIT “V”;YMULT
15.6250E-6;YOFF :”6.4000E+3;YZERO 0.0000
WFMOutpre:BIT_Nr
This command specifies the number of bits per data point in the outgoing
waveform being transferred using the
CURVe? query. Changing the value of
WFMOutpre:BIT_Nr also changes the values of WFMOutpre:BYT_Nr and
DATa:WIDth. The waveform is specified by the DATa:SOUrce command.
(See page 2-95, Waveform Transfer Command Group.)
(See page D-1, Waveform Transfer (WFMOutpre and CURVe Query) Examples.)
NOTE. The
WFMOutpre:BYT _Nr
and
WFMOutpre:BIT _Nr
settings are
directly related; setting one causes the other to be set accordingly. For
example,
WFMOutp
re:BYT_Nr 2
causes
WFMOutpre_B IT_Nr
to be set to
16 (2 * 8 bits/byte). Similarly, setting
WFMOutpre: BIT_Nr
to 16 causes
WFMOutpre: BYT_Nr
to be set to 2.
Group
Waveform Transfer
2-960 MDO4000/B/C, MSO/DPO4000B and MDO3000 Series Oscilloscopes Programmer Manual