MDO3000 Programmer Manual.pdf - 第322页

Commands Listed in A lphabetical Order Syntax DISplay:INTEN SITy:BACKLight:A UTODim:ENAble {OFF|ON|0|1} DISplay:INTEN SITy:BACKLight:A UTODim:ENAble? Arguments OFF or 0 turns off the display auto-dim feature. ON or 1 tur…

100%1 / 1158
Commands Listed in Alphabetical Order
GRId specifies a
frame and grid only.
SOLid specifies a solid graticule.
Examples
DISPLAY:GRATICULE FRAme sets the graticule type to display the frame only.
DISPLAY:GRATICULE? might return :DISPLAY:GRAT ICULE FULL indicating
that all graticule elements are selected.
DISplay:INTENSITy? (Query Onl
y)
Returns the display intensity settings.
Group
Display
Syntax
DISplay:INTENSITy?
Examples
DISPLAY:INTENSITY? might return: :D ISPLAY:INTEnsITY:WAVEFORM
30;GRATICULE 75;BACKLIG HT HIGH
DISplay:INTENSITy:BACKLight
Sets and returns the waveform backlight intensity settings.
Group
Display
Syntax
DISplay:INTENSITy:BACKLight {LOW|MEDium|HIGH}
DISplay:INTENSITy:BACKLight?
Examples
DISPLAY:INTENSITY:BACKLIGHT? might return
DISPLAY:INTEnsITY:BACKLIGHT HIGH
DISplay:INTENSITy:BACKLight:AUTODim:ENAble
Sets or returns the state of the display auto-dim feature. The default is enabled.
Once the backlight has dimmed, any button push or knob turn returns the backlight
value to the value set by DISplay:INTENSITy:BACKLight.
Group
Display
MDO4000/B/C, MSO/DPO4000B and MDO3000 Series Oscilloscopes Programmer Manual 2-291
Commands Listed in Alphabetical Order
Syntax
DISplay:INTEN
SITy:BACKLight:AUTODim:ENAble {OFF|ON|0|1}
DISplay:INTENSITy:BACKLight:AUTODim:ENAble?
Arguments
OFF or 0 turns off the display auto-dim feature.
ON or 1 turns it on.
DISplay:INTENSITy:BACKLight:AUTODim:TIMe
Sets or returns the amount of time, in minutes, to wait for no UI activity before
automatically dimming the display. The time can range from a minimum of
10 minutes to a maximum of 1440 minutes (24 hours). The default is 60 minutes.
Group
Display
Syntax
DISplay:INTENSITy:BACKLight:AUTODim:TIMe <NR1>
DISplay:INTENSITy:BACKLight:AUTODim:TIMe?
Arguments
An integer that represents the amount of time, in minutes, to wait for no UI
activity before automatically dimming the display.
DISplay:INTENSITy:GRAticule
Sets and ret
urns the display graticule intensity settings.
Group
Display
Syntax
DISplay:INTENSITy:GRAticule <NR1>
DISplay:INTENSITy:GRAticule?
Arguments
<NR1> is the graticule intensity and ranges from 0 t
o100percent.
Examples
DISPLAY:INTENSITY:GRATICULE? might return
DISPLAY:INTEnsITY:GRATICULE 30
DISplay:INTENSITy:WAVEform
Sets and returns the display waveform intensity settings.
2-292 MDO4000/B/C, MSO/DPO4000B and MDO3000 Series Oscilloscopes Programmer Manual
Commands Listed in Alphabetical Order
Group
Display
Syntax
DISplay:INTENSITy:WAVEform <NR1>
DISplay:INTENSITy:WAVEform?
Arguments
<NR1> is the waveform intensity and ranges from 1 to 100 percent.
Examples
DISPLAY:INTENSITY:WAVEFORM? might return
DISPLAY:INTEnsITY:WAVEFORM 60
as the intensity of the waveforms.
DISplay:PERSistence
This command specifies the display persistence for analog waveforms. This
affects the display only.
Group
Display
Syntax
DISplay:PERSistence {<NR3>|CLEAR|AUTO|INFInite|OFF}
DISplay:PERSistence?
Arguments
<NR3> is a floating point number that specifies the time of the persistence.
CLEAR resets the persist time count down and clears the display of acquired points.
INFInite displays waveform points until a control change resets the acquisition
system.
NOTE. When persistence is set to innite, it does not mean that the b rightness
of any pixel should never decrease. The brightness of a pixel is proportionally
dependent on the r atio between its intensity, which does NOT decrease at innite
persistence, and the maximum v alue of intensity of any pixel on the screen. Thus,
if a particular pixel gets hit less often than others, its brightness will decrease
over time. It will become less bright relative to the pixels that get hit more often.
AUTO specifies that the oscilloscope automatically determines the
best waveform persistence based on the value of waveform intensity
(DISplay:INTENSITy:WAVEform)
OFF turnsoffDPOmode(0secondsofpersistence).
MDO4000/B/C, MSO/DPO4000B and MDO3000 Series Oscilloscopes Programmer Manual 2-293