MDO3000 Programmer Manual.pdf - 第722页
Commands Listed in A lphabetical Order Arguments 0 1 X sets the value to X (“don’t care”) w hich is the default. ZERO sets the value to 0. ONE sets the value to 1. NOCARE sets the value to X (“don’t care”) which is the d…

Commands Listed in Alphabetical Order
Related Commands
SEARCH:SEARCH
<x>:TRIGger:A:BUS:B<x>:MIL1553B:CONDition
Arguments
0
1
X
sets the value to X (“don’t care”) which is the default.
ZERO sets the value to 0.
ONE sets the value to 1.
NOCARE sets the value to X (“don’t care”) which is the default.
OFF sets the value to 0.
ON sets the value to 1.
Examples
SEARCH:SEARCH1:TRIGger:A:BUS:B1:MIL1553B:STATus:BIT:
SUBSF N OCARE
sets the status word subsystem flag bit value to be used in the
search to X.
SEARCH:SEARCH1:TRIGger:A:BUS:B1:MIL1553B:STATus:BIT:
SUBSF?
might r
eturn 1.
SEARCH:SEARCH<x>:TRIGger:A:BUS:B<x>:MIL1553B:STATus:BIT:TF
When the MIL-STD-1553 bus search condition is set to STATus, this command
specifies the status word terminal flag bit value (bit 19) to be used in the search.
Returned values are 0, 1, or X (don’t care, which is the default).
SEARCH<x> is
the search number, which is always 1, and
B<x>
Conditions
This command requires a DPO4AERO application module for DPO/MSO4000B
and MDO4000/B/C models or MDO3AERO application module for
MDO3000 models.
Group
Search
Syntax
SEARCH:SEARCH<x>:TRIGger:A:BUS:B<x>:MIL1553B:STATus:BIT:TF
{0|1|X|ZERo|ONE|NOCARE|OFF|ON}
SEARCH:SEARCH<x>:TRIGger:A:BUS:B<x>:MIL1553B:STATus:BIT:TF?
Related Commands
SEARCH:SEARCH<x>:TRIGger:A:BUS:B<x>:MIL1553B:CONDition
MDO4000/B/C, MSO/DPO4000B and MDO3000 Series Oscilloscopes Programmer Manual 2-691

Commands Listed in Alphabetical Order
Arguments
0
1
X
sets the value to X (“don’t care”) which is the default.
ZERO sets the value to 0.
ONE sets the value to 1.
NOCARE sets the value to X (“don’t care”) which is the default.
OFF sets the value to 0.
ON sets the value to 1.
Examples
SEARCH:SEARCH1:TRIGger:A:BUS:B1:MIL1553B:STATus:BIT:TF
NOCARE
sets the st
atus word terminal flag bit value to be used in the search to X.
SEARCH:SEARCH1:TRIGger:A:BUS:B1:MIL1553B:STATus:BIT:TF? might
return 1.
SEARCH:SEARCH<x>:TRIGger:A:BUS:B<x>:MIL1553B:STATus:PARity
When the MIL-STD-1553 bus search condition is set
to
STATus, this command
specifies the status parity bit value to be used in the search. Returned values are
0, 1, or X (don’t care, which is the default).
SEARCH<x> is the search number,
whichisalways1,and
B<x>
Conditions
This command requires a DPO4AERO application module for DPO/MSO4000B
and MDO4000/B/C models or MDO3AERO application module for
MDO3000 models.
Group
Search
Syntax
SEARCH:SEARCH<x>:TRIGger:A:BUS:B<x>:MIL1553B:STATus:PARity
{0|1|X|ZERo|ONE|NOCARE|OFF|ON}
SEARCH:SEARCH<x>:TRIGger:A:BUS:B<x>:MIL1553B:STATus:PARity?
Related Commands
SEARCH:SEARCH<x>:TRIGger:A:BUS:B<x>:MIL1553B:CONDition
Arguments
0
1
X
sets the value to X (“don’t care”) which is the default.
2-692 MDO4000/B/C, MSO/DPO4000B and MDO3000 Series Oscilloscopes Programmer Manual

Commands Listed in Alphabetical Order
ZERO sets the va
lue to 0.
ONE sets the value to 1.
NOCARE sets the value to X (“don’t care”) which is the default.
OFF sets the value to 0.
ON sets the value to 1.
Examples
SEARCH:SEARCH1:TRIGger:A:BUS:B1:MIL1553B:STATus:PARity
ONE
sets the parity bit value to be used in the search to
1.
SEARCH:SEARCH1:TRIGger:A:BUS:B1:MIL1553B:STATus:PARity?
might return X, indicating that the parity do
esn’t matter.
SEARCH:SEARCH<x>:TRIGger:A:BUS:B<x>:MIL1553B:TIMe:LESSLimit
When the MIL-STD-1553 bus search condition is set to TIMe, this command
specifies either the minimum remote terminal response time (RT) limit
for the amount of time the terminal has to transmit, or it specifies the
minimum inter-message gap (IMG). (You can specify RT or IMG using the
SEARCH:SEARCH<x>:TRIGger:A:BUS:B<x>:MIL1553B:CONDition TIMe
command.)
SEARCH<x> is the search number, which is always 1, and B<x>
Conditions
This command requires a DPO4AERO application module for DPO/MSO4000B
and MDO4000/B/C models or MDO3AERO application module for
MDO3000 models.
Group
Search
Syntax
SEARCH:SEARCH<x>:TRIGger:A:BUS:B<x>:MIL1553B:TIMe:LESSLimit
<NR3>
SEARCH:SEARCH<x>:TRIGger:A:BUS:B<x>:MIL1553B:TIMe:LESSLimit?
Related Commands
SEARCH:SEARCH<x>:TRIGger:A:BUS:B<x>:MIL1553B:CONDition
SEARCH:SEARCH<x>:TRIGger:A:BUS:B<x>:MIL1553B:TIMe:MORELimit
Arguments
<NR3> is a floating point number that specifies either the minimum remote
terminal response time (RT) or the inter-message gap (IMG) in seconds.
Examples
SEARCH:SEARCH1:TRIGger:A:BUS:B1:MIL1553B:TIMe:LESSLimit
3.0000e-6
would set either the remote terminal response time (RT) or the
inter-message gap (IMG) to 3.00 μS.
MDO4000/B/C, MSO/DPO4000B and MDO3000 Series Oscilloscopes Programmer Manual 2-693