Commands Listed in Alphabetical Order SEARCH:SEARCH <x>:TRIGger:A:BU S:B<x>:MIL1553B: COMMAND: ADDRess:VALue ? Related Commands SEARCH:SEARCH<x>:TRIGger:A:BU S:B<x>:MIL1553B:CONDition SEARCH:SEARC…

Commands Listed in Alphabetical Order
Syntax
SEARCH:SEARCH
<x>:TRIGger:A:BUS:B<x>:MIL1553B:COMMAND:
ADDRess:HIVALue <QString>
SEARCH:SEARC
H<x>:TRIGger:A:BUS:B<x>:MIL1553B:COMMAND:
ADDRess:HIVALue?
Related Commands
SEARCH:SEARCH<x>:TRIGger:A:BUS:B<x>:MIL1553B:CONDition
SEARCH:SEARCH<x>:TRIG:A:BUS:B<x>:MIL1553B:COMMAND:ADDR:
QUAL
SEARCH:SEARCH<x>:TRIGger:A:BUS:B<x>:MIL1553B:COMMAND:
ADDRess:VALue
Arguments
QString is a quoted string of up to 5 characters where the allowable characters are
0, 1, and X.
Examples
SEARCH:SEARCH1:TRIGger:A:BUS:B1:MIL1553B:HIVALue “01000” sets
the upper limit of the remote terminal address range to 01000 (when the search
condition has been set to
MIL1553B, and the qualifier has been set to INrange
or OUTrange).
SEARCH:SEARCH1:TRIGger:A:BUS:B1:MIL1553B:HIVALue?
mig
ht return
“XXXXX”.
SEARCH:SEARCH<x>:TRIGger:A:BUS:B<x>:MIL1553B:COMMAND:ADDRess:VALue
When the MIL-STD-1553 bus search condition is set to CO MMAND,andthe
qualifier is set to
LESSthan, MOREthan, EQual, UNEQual, LESSEQual
or MOREEQual, this command specifies the value of the 5–bit remote terminal
address to be used in the search. When the MIL-STD-1553 bus search conditionis
set to
COMMAND, and the qualifier is set to INrange or OUTrange, this command
specifies the lower limit of the remote terminal address range. The default is all
X’s (don’t care).
SEARCH<x> is the search number, which is always 1, and B<x>
Conditions
This command requires a DPO4AERO application module for DPO/MSO4000B
and MDO4000/B/C models or MDO3AERO application module for
MDO3000 models.
Group
Search
Syntax
SEARCH:SEARCH<x>:TRIGger:A:BUS:B<x>:MIL1553B:COMMAND:
ADDRess:VALue <QString >
2-672 MDO4000/B/C, MSO/DPO4000B and MDO3000 Series Oscilloscopes Programmer Manual

Commands Listed in Alphabetical Order
SEARCH:SEARCH
<x>:TRIGger:A:BUS:B<x>:MIL1553B:COMMAND:
ADDRess:VALue?
Related Commands
SEARCH:SEARCH<x>:TRIGger:A:BUS:B<x>:MIL1553B:CONDition
SEARCH:SEARCH<x>:TRIGger:A:BUS:B<x>:MIL1553B:COMMAND:
ADDRess:HIVALue
SEARCH:SEARCH<x>:TRIG:A:BUS:B<x>:MIL1553B:COMMAND:ADDR:
QUAL
Arguments
QString is a quoted string of up to 5 characters, where the allowable characters are
0, 1, and X.
Examples
SEARCH:SEARCH1:TRIGger:A:BUS:B1:MIL1553B:COMMAND:ADDRess:
VALue “ 01000”
sets the remote terminal address to be used in the search to
01000.
SEARCH:SEARCH1:TRIGger:A:BUS:B1:MIL1553B:COMMAND:ADDRess:
VALue?
might return 01000.
SEARCH:SEARCH<x>:TRIG:A:BUS:B<x>:MIL1553B:COMMAND:ADDR:QUAL
When the MIL-STD-1553 bus search condition is set to COMMAND, this command
specifies the qualifier to be used with the remote terminal address field.
SEARCH<x> is the search number, which is always 1, and B<x>
Conditions
This command requires a DPO4AERO application module for DPO/MSO4000B
and MDO4000/B/C models or MDO3AERO application module for
MDO3000 models.
Group
Search
Syntax
SEARCH:SEARCH<x>:TRIG:A:BUS:B<x>:MIL1553B:
COMMAND:ADDR:QUAL {LESSthan|MOREthan|EQual
|UNEQual|LESSEQual|MOREEQual|INrange|OUTrange}
SEARCH:SEARCH<x>:TRIG:A:BUS:B<x>:MIL1553B:COMMAND:ADDR:QUAL?
Related Commands
SEARCH:SEARCH<x>:TRIGger:A:BUS:B<x>:MIL1553B:COMMAND:
ADDRess:VALue
SEARCH:SEARCH<x>:TRIGger:A:BUS:B<x>:MIL1553B:COMMAND:
ADDRess:HIVALue
MDO4000/B/C, MSO/DPO4000B and MDO3000 Series Oscilloscopes Programmer Manual 2-673

Commands Listed in Alphabetical Order
Arguments
LESSthan sets t
he Command Address qualifier to less than.
MOREthan sets the Command Address qualifier to greater than.
EQual sets the Command Address qualifier to equal.
UNEQual sets the Command Address qualifier to not equal.
LESSEQual sets the Command Address qualifier to less than or equal.
MOREEQual sets the Command Address qualifier to greater than or equal.
INrange sets the Command Address qualifier to in range.
OUTrange sets the Command Address qualifier to out of range.
Examples
SEARCH:SEARCH1:TRIGger:A:BUS:B1:MIL1553B:COMMAND:ADDRess:
QUALifier EQual
sets the qualifier to be used with the remote terminal address
to equal.
SEARCH:SEARCH1:TRIGger:A:BUS:B1:MIL1553B:COMMAND:ADDRess:
QUALifier?
might return INRANGE.
SEARCH:SEARCH<x>:TRIGger:A:BUS:B<x>:MIL1553B:COMMAND:COUNt
When the MIL-STD-1553 bus search condition is set to COMMAND, this command
specifies the bit pattern for the 5–bit Word Count/Mode Code sub-address field
that is to be used in the search. (Use the commandSEARCH:SEARCH<x>:
TRIGger:A:BUS:B<x>:MIL1553B:COMMAND:SUBADdress to specify Word
Count or Mode Code.) In Word Count mode, this field defines the number of data
words that is to be transmitted, or received, depending on the T/R bit setting.
(Use the commandSEARCH:SEARCH<x>:TRIGger:A:BUS:B<x>:MIL1553B:
COMMAND:TRBit to set the T/R bit.) A word count value of 0 actually indicates
a transfer of 32 data words.
SEARCH<x> is the search number, which is always
1, and
B<x>
Conditions
This command requires a DPO4AERO application module for DPO/MSO4000B
and MDO4000/B/C models or MDO3AERO application module for
MDO3000 models.
Group
Search
Syntax
SEARCH:SEARCH<x>:TRIGger:A:BUS:B<x>:MIL1553B:COMMAND:COUNt
<QString>
SEARCH:SEARCH<x>:TRIGger:A:BUS:B<x>:MIL1553B:COMMAND:COUNt?
Related Commands
SEARCH:SEARCH<x>:TRIGger:A:BUS:B<x>:MIL1553B:CONDition
2-674 MDO4000/B/C, MSO/DPO4000B and MDO3000 Series Oscilloscopes Programmer Manual