Commands Listed in Alphabetical Order
E1Coax –E1Coax
ial Pair 2.048 Mb/s
E1Symmetrical E1 Symmetric Pair 2.048 Mb/s
E2 E2 8.448 Mb/s
E3 E3 34.368 Mb/s
USB:
HST1 HS:T1 480 Mb/s
HST2 –HS:T
1480Mb/s
HST3 HS:T1 480 Mb/s
HST4 HS:T1 480 Mb/s
HST5 HS:T1 480 Mb/s
HST6 HS:T1 480 Mb/s
In addition to the standard masks,
LIMit, USER, and NONe can be input as well.
Returns
MASK:STANdard? returns one of the following:
NONE indicates that the mask application is not active.
USER indicates one of the following:
the current mask has been recalled from a file or
the current mask has been created from scratch using the MASK commands or
the current mask has been created by modifying one of the built-in standard
telecommunications masks.
LIMIT indicates that the mask is a limit test (template) mask.
DS1, etc. indicates that the current mask is one of the built-in standard masks.
Examples
MASK:STANdard DSO Double replaces the existing mask, if any, with a
standard ITU-T DSO Double mask.
MASK:STANdard? might return NONE, indicating that the mask application
is not currently active.
MASK:STOPOnviolation
This command stops the waveform acquisitions upon the first occurrence
of a waveform violation. You can also specify an action to be performed
when acquisitions are stopped by using the commands such as MASK:TESt:
2-380 MDO4000/B/C, MSO/DPO4000B and MDO3000 Series Oscilloscopes Programmer Manual
Commands Listed in Alphabetical Order
SAVEWFM, MASK:
TESt:SAVEIMAGE, MASK:TESt:AUXout:FAILure,
MASK:TESt:HARDCopy,orMASK:TESt:SRQ:FAILure.
Aseriesofexa
mples showing how to use mask commands for typical tasks
is included in an appendix. (See page E-1, Mask/Limit Command Sequence
Examples.)
Conditions
This command requires a DPO4LMT application module for 4000 series models,
and an MDO3LMT application module for MDO3000 series models.
Group
Mask
Syntax
MASK:STOPOnviolation {ON|OFF|<NR1>}
MASK:STOPOnviolation?
Related Commands
MASK:TESt:SAVEWFM,
MASK:TESt:SAVEIMAGE,
MASK:TESt:AUXout:FAILure,
MASK:TESt:HARDCopy,
MASK:TESt:SRQ:FAILure
Arguments
ON stops waveform acquisition
on the first occurrence of a mask violation.
OFF turns this feature off.
<NR1>=0 turns this feature off ; any other value turns it on.
Examples
MASK:STOPONVIOLATION ON sets the instrument to stop waveform acquisition
on the first occurrence of a waveform violation.
MASK:STOPONVIOLATION? might return :MASK:STOPONVIOLATION ON,
indicating that the waveform acquisitions will be stopped upon the first occurrence
of a waveform violation.
MASK:TEMPLate:CREATEmask (No Query Form)
This command, used for limit mask testing, causes a template mask to be created
based on the settings of the MASK:TEMPLate:SOUrce, MASK:TEMPLate:
TOLerance:HORizontal,andMASK:TEMPLate:TOLerance:VERTical
commands. If you do not specify horizontal or vertical tolerances, the default
tolerance of one minor division will be used.
MDO4000/B/C, MSO/DPO4000B and MDO3000 Series Oscilloscopes Programmer Manual 2-381
Commands Listed in Alphabetical Order
Aseriesofexam
ples showing how to use mask commands for typical tasks
is included in an appendix. (See page E-1, Mask/Limit Command Sequence
Examples.)
NOTE. To ensure that Mask commands are enabled, use the command
APPLication:TYPe LIMITMask.
Conditions
This comma
nd requires a DPO4LMT application module for 4000 series models,
and an MDO3LMT application module for MDO3000 series models.
Group
Mask
Syntax
MASK:TEMPLate:CREATEmask
Related Commands
MASK:TEMPLate:SOUrce,
MASK:TEMPLate:TOLerance:HORizontal,
MASK:TEMPLate:TOLerance:VERTical,
SAVe:MASK
MASK:TEMPLate:SOUrce
This command, typically used for limit mask testing, specifies the source
waveform to be used to create a template mask. The source can be either
one of four channels, or one of four saved reference waveforms. See the
SAVe:WAVEform commands for more information on creating reference
waveforms.
A series of examples showing how to use mask commands for typical tasks
is included in an appendix. (See pag
eE-1,Mask/Limit Command Sequence
Examples.)
NOTE. To ensure that Mask commands are enabled, use the command
APPLication:TYPe LIMITMask.
Conditions
This command requires a DPO4LMT application module for 4000 series models,
andanMDO3LMTapplicationmodu
le for MDO3000 series models.
Group
Mask
2-382 MDO4000/B/C, MSO/DPO4000B and MDO3000 Series Oscilloscopes Programmer Manual