Command Groups T able 2-31: Mask commands (cont.) Command Description MASK:TESt:AUXout:COMPLetion Causes the instrument to send a TTL signal to the AUX:out port whenever a pass/fail mask test completes. MASK:TESt:AUXout:…

100%1 / 1158
Command Groups
number of wavef
orms or by time, or specify a violation threshold that must be
met before considering a test a failure, or specify an action (such as writing to a
file or sending an SRQ) that is to be performed upon a violation, a test failure,
or test completion. Use MASK:CUSTom andthenthe
:MASK:USER commands
to create user-defined custom masks.
A series of examples showing how to use mask commands for typical tasks
is included as an appendix. (See page E-1, Mask/Limit Command Sequence
Examples.)
Table 2-31
: Mask commands
Command
Descripti
on
APPLicat
ion:TYPe
When a mask/limit or power test application module is installed, one of the associated test
types is always selected by default. This command allows the test type to be changed
from the
default.
MASK:CO
Py:SOUrce
Specie
s the standard mask that is to be copied to a user created custom mask.
MASK:C
OPy:USER
Copies
the specied standard mask to a user created custom mask.
MASK:C
OUNt
Resets
to zero the number of hits and failures for all mask segments.
MASK:
COUNt:FAILURES?
Retur
ns the number of pass/fail mask tests that have failed
MASK
:COUNt:HITS?
This
query returns the sum of all hits in all mask segments.
MASK
:COUNt:SEG<x>:HITS?
This
query returns the number of hits for the specied mask segment.
MAS
K:COUNt:TESTS?
Ret
urns the number of pass/fail tests that have occurred
MAS
K:COUNt:VIOLATIONS?
Ret
urns the number of test violations that have occurred i n the current pass/fail test.
MA
SK:COUNt:WAVEFORMS?
Re
turns the number of waveforms that have been acquired and processed during pass/fail
mask testing.
MASK:CUSTom Sets the user-dened custom mask to its initialized state, or copies the currently active
mask to the user-dened custom mask.
MASK:DISplay Controls whether a user-dened custom mask is displayed on the screen.
MASK:LOCk Locks the mask to the waveform so that any changes made to the horizontal and/or vertical
scale settings of the w aveform will redraw the mask segments in proportion.
MASK:MARgin:PERCent Sets or returns the tolerance for the mask test. A positive value expands the m ask and a
negative margin shrinks the m ask by the specied percentage.
MASK:SOUrce Species the mask source waveform to be used during pass/fail mask testing
MASK:STANdard Replaces the existing m ask, if any, with a specied standard mask.
MASK:STOPOnviolation Stops the waveform acquisitions upon the rst occurrence of a waveform violation.
MASK:TEMPLate:CREATEmask Causes a template m ask to be created for limit testing, based on the s ettings of the
MASK:TEMPLate:SOUrce, MASK:TEMPLate:TOLerance:H ORizontal, and
MASK:TEMPLate:VERTical commands.
MASK:TEMPLate:SOUrce Species the source waveform to be used to create a m ask template for limit testing.
MASK:TEMPLate:TOLerance:
HORizontal
Species the +/- horizontal limit (tolerance) for a mask template to be used for limit testing.
MASK:TEMPLate:TOLerance:
VERTical
Species the +/- vertical limit (tolerance) for a mask template to be used for limit testing.
2-32 MDO4000/B/C, MSO/DPO4000B and MDO3000 Series Oscilloscopes Programmer Manual
Command Groups
Table 2-31: Mask commands (cont.)
Command
Description
MASK:TESt:AUXout:COMPLetion Causes the instrument to send a TTL signal to the AUX:out port whenever a pass/fail
mask test completes.
MASK:TESt:AUXout:FAILure Causes the instrument to send a T TL signal to the AUX:out port whenever a pass/fail
mask test fails.
MASK:TESt:COMPLetion:CRITerion Species the criterion to be used for test completion (either by waveform or time) during
pass/fail mask testing.
MASK:TESt:DELay Species the amount of time, in seconds, the instrument should wait after the start of
pass/fail mask testing before it evaluates the waveforms.
MASK:TESt:HARDCopy Causes the instrument to generate a screen hard copy to the default printer as soon as a
pass/fail mask test fails, using the current instrument hard copy settings
MASK:TESt:REPeat Causes the mask test cycle to continuously repeat upon the completion of the previous
test cycle.
MASK:TESt:SAVEIMAG E Causes the instrument to copy the screen image to a U SB media le or a mounted network
drive if a pass/fail mask test fails.
MASK:TESt:SAVEWFM Copies the waveform data from all active channels to a le if a pass/fail m ask test fails.
MASK:TESt:SRQ:CO MPLetion Causes the instrument to send an SRQ command when a pass/fail mask test completes
MASK:TESt:SRQ:FAILure Causes the instrument to send an SRQ command when a pass/fail mask test fails.
MASK:TESt:STATE Turns the pass/fail mask test on or off.
MASK:TESt:STATus? Indicates the result of a pass/fail mask test.
MASK:TESt:STOP:FAILure Causes the instrument to stop acquiring data when a pass/fail mask test fails.
MASK:TESt:THReshold Species the number of failed tested waveforms needed in a pass/fail mask test to cause
the test status to change to ‘Failing’.
MASK:TESt:TIME Species the duration, in seconds, the instrument should run a pass/fail mask test.
MASK:TESt:WAVEform Species the number of waveforms the instrument should test during a pass/fail mask test.
MASK:USER:AMPLitude Species the nominal pulse amplitude, in volts, to be used for a user-dened custom mask.
MASK:USER:HSCAle Species the nominal timing resolution, in time/division, to be used to draw a user-dened
custom mask pulse shape.
MASK:USER:HTRIGPOS Species the nominal trigger position (pulse leading edge) to be used to draw a user-dened
custom mask, as a fraction o f the display width.
MASK:USER:LABel This command species a user-dened label for a custom mask.
MASK:USER:RECOrdlength Species the nominal record length to be used for pulse mask testing with a user-dened
custom mask.
MASK:USER{:SEG<x>|:MASK<x>} This command deletes the specied mask segment from the current mask segment. There
canbeupto8segments.
MASK:USER{:SEG<x>|:MASK<x>}:
NR_Pt?
This query returns the number of points that make up the specied mask segment of a
user-dened custom mask. There can be up to 8 segments.
MASK:USER{:SEG<x>|:MASK<x>}:
POINTS
This command species the x and y coordinates of the points that make up the segment
1–8. The units are normal waveform units. The x-coordinate is specied relative to the
trigger. The points are specied as a sequence of (x,y) points which traverse the boundary
of the s egment in a counter-clockwise direction.
MDO4000/B/C, MSO/DPO4000B and MDO3000 Series Oscilloscopes Programmer Manual 2-33
Command Groups
Table 2-31: Mask commands (cont.)
Command
Description
MASK:USER:TRIGTOSAMP Species the nominal time, in seconds, from the (leading edge) trigger position to the pulse
bit sampling position, to be used for testing with a user-dened custom mask.
MASK:USER:VOFFSet Species the nominal value, in volts, to be used to vertically offset the input channels
for a user-de ned custom m a sk.
MASK:USER:VPOS Species the nominal value, in divisions, to be used to vertically position the input channels
for a user-de ned custom m a sk.
MASK:USER:VSCAle Species the nominal value, in volts per division, to be used to vertically scale the input
channels for a user-dened custom mask.
MASK:USER:WIDth Species the nominal bit width value, in seconds, to b e used for a user-dened custom
mask.
RECAll:MASK This command recalls the mask from a specied le that was saved to the current working
directory using the SAVe:MASK command.
SAVe:MASK This command saves the current mask de nition to the le specied with a quoted string,
into the current working directory. You can recall the mask from the le by using the
command RECAll:MASK.
2-34 MDO4000/B/C, MSO/DPO4000B and MDO3000 Series Oscilloscopes Programmer Manual