Commands Listed in Alphabetical Order Conditions MDO3000 and MDO 4000C only; requires DVM option key to be installed. Group DVM Synt ax DVM:MEASUreme nt:HIStory:MINIm um? Examples DVM:MEASU:HIS :MINI? might return 100.00…

100%1 / 1158
Commands Listed in Alphabetical Order
DVM:MEASUrem
ent:HIStory:AVErage? (Query Only)
Returns the average readout value for the Digital Voltmeter function over the
history period.
Any qualifier, such as "Measurement warning, Clipping positive", is automatically
set as an error event.
Conditions
MDO3000 and MDO4000C only; requires DVM option key to be installed.
Group
DVM
Syntax
DVM:MEASUrement:HIStory:AVErage?
Examples
DVM:MEASU:HIS:AVE? might return 100.0000E+3, which represents the
average DVM readout value over the history period.
DVM:MEASUrement:HIStory:MAXimum? (Query Only)
Returns the maximum readout value for the DVM function over the history period.
Any qualifier, such as "Measurement warning, Clipping positive", is automatically
set as an error event.
Conditions
MDO3000 and MDO4000C only; requires DVM option key to be installed.
Group
DVM
Syntax
DVM:MEASUrement:HIStory:MAXimum?
Examples
DVM:MEASU:HIS:MAX? might return 100.0000E+3, which represents the
maximum readout value for the DVM function over the history period.
DVM:MEASUrement:HIStory:MINImum? (Query Only)
Returns the minimum readout value for the DVM function over the history period.
Any qualifier, such as "Measurement warning, Clipping positive", is automatically
set as an error event.
2-298 MDO4000/B/C, MSO/DPO4000B and MDO3000 Series Oscilloscopes Programmer Manual
Commands Listed in Alphabetical Order
Conditions
MDO3000 and MDO
4000C only; requires DVM option key to be installed.
Group
DVM
Synt
ax
DVM:MEASUrement:HIStory:MINImum?
Examples
DVM:MEASU:HIS:MINI? might return 100.0000E+3, which represents the
minimum readout value for the DVM function over the history period.
DVM:MEASUrement:INFMAXimum? (Query Only)
Returns the maximum DVM readout value over the entire time that the DVM has
been on since the last change using the DVM:MODe or DVM:SOUrce commands
or
DVM RESE T.
Any qualifier, such as "Measurement warning, Clipping positive", is automatically
set as an error event.
Conditions
MDO3000 and MDO4000C only; requires DVM option key to be installed.
Group
DVM
Syntax
DVM:MEASUrement:INFMAXimum?
Examples
DVM:MEASU:INFMAX? might return 100.0000E+3, which represents the
maximum readout value of the DVM function over the entire time that the DVM
has been on since the last change using the DVM:MODe or DVM:SOUrce
commands or
DVM RESET.
DVM:MEASUrement:INFMINimum? (Query Only)
Returns the minimum readout value of the DVM function over the entire time
that the DVM has been on since the last change using the DVM:MODe or
DVM:SOUrce commands or
DVM RESE T.
Any qualifier, such as "Measurement warning, Clipping positive", is automatically
set as an error event.
Conditions
MDO3000 and MDO4000C only; requires DVM option key to be installed.
MDO4000/B/C, MSO/DPO4000B and MDO3000 Series Oscilloscopes Programmer Manual 2-299
Commands Listed in Alphabetical Order
Group
DVM
Syntax
DVM:MEASUrement:INFMINimum?
Examples
DVM:MEASU:INMIN? might return 100.0000E+3, which represents the
minimum readout value of the DVM function over the entire time that the DVM
has been on since the last change using the DVM:MODe or DVM:SOUrce
commands or
DVM RESE T.
DVM:MEASUrement:VALue? (Query Only)
Returns the Digital Voltmeter value.
Any qualifier, such as "Measurement warning, Clip
ping positive", is automatically
set as an error event.
Conditions
MDO3000 and MDO4000C only; requires DVM option key to be installed.
Group
DVM
Syntax
DVM:MEASUrement:VALue?
Examples
DVM:MEASU:VAL? might return 100.0000E+3, which represents the DVM value.
DVM:MODe
Specifies (or queries) the mode to use for the Digital Voltmeter (ACRMS,
ACDCRMS, DC, Frequency, or OFF).
Conditions
MDO3000 and MDO4000C only; requires DVM option key to be installed.
Group
DVM
Syntax
DVM:MODe {ACRMS|ACDCRM S|DC|FREQuency|OFF}
DVM:MODe?
Arguments
ACRMS displays the root-mean-square value of the acquired data, with the DC
component removed.
2-300 MDO4000/B/C, MSO/DPO4000B and MDO3000 Series Oscilloscopes Programmer Manual