MDO3000 Programmer Manual.pdf - 第68页
Command Groups T able 2-33: Measurement Commands (cont.) Command Description MEASUrement:METHod This command speci fi es the method used for calculating reference levels MEASUrement:REFLevel? Returns the current reference…

Command Groups
Table 2-33: Measurement Commands (cont.)
Command
Description
MEASUrement:IMMed:DELay? Returns information about the immediate delay measurement
MEASUrement:IMMed:DELay:
DIRection
This command specifies the search direction to use for immediate delay measurements
MEASUrement:IMMed:DELay:
EDGE<x>
This command specifies the slope of the edge used for immediate delay “from” and “to”
waveform measurements
MEASUrement:IMMed:SO Urce<x> This command specifies the “from” source for all single channel immediate measurements
This command specifies the source to measure “to” for two-channel measurements
MEASUrement:IMMed:TYPe This command specifies the type of the immediate measurement
MEASUrement:IMMed:UN Its? Returns the units of the i mmediate measurement
MEASUrement:IMMed:VALue? Returns the value of the immediate measurement
MEASUrement:INDICa tors?
Returns all measurement indicator parameters
MEASUrement:INDICa tors:
HORZ<x>?
Returns the position of the specified horizontal measurement indicator
MEASUrement:INDICa tors:
NUMHORZ?
Returns the number of horizontal measurement indicators currently being displayed
MEASUrement:INDICa tors:
NUMVERT?
Returns the number of vertical measurement indicators currently being displayed
MEASUrement:INDIC ators:STATE This command specifies the state of visible m easurement indicators
MEASUrement:INDICa tors:
VERT<x>?
Returns the value of the specified vertical measurement indicator
MEASUrement:MEAS<x>?
Returns all measurement parameters
MEASUrement:MEAS<x>:COUNt? Returns the number of values accumulated since the last statistical reset
MEASUrement:MEAS<x>:DELay? Returns the delay measurement parameters for the specified measurement
MEASUrement:MEAS<x>:DELay:
DIRection
This command specifies the search direction to use for delay measurements
MEASUrement:MEAS<x>:DELay:
EDGE<x>
This command specifies the slope of the edge to use for delay “from” and “to” waveform
measurements
MEASUrement:MEAS<x>:
MAXimum?
Returns the maximum value found since the last statistical reset
MEASUrement:MEAS<x>:MEAN?
Returns the mean value accumulated since the last statistical reset
MEASUrement:MEAS<x>:
MINImum?
Returns the minimum value found since the last statistical reset
MEASUrement:MEAS<x>:
SOUrce<x>
This command specifies the “from” source for all single channel immediate measurements
This command specifies the source to measure “to” for two-channel measurements
MEASUrement:MEAS<x>:STATE This command specifies whether the specified measurement slot is computed and displayed
MEASUrement:MEAS<x>:STDdev? Returns the standard deviation of values accumulated since the last statistical reset
MEASUrement:MEAS<x>:TYPe This command specifies the measurement<x> type
MEASUrement:MEAS<x>:UNIts?
Returns measurement<x> units
MEASUrement:MEAS<x>:VALue? Returns the value of measurement<x>
MDO4000/B/C, MSO/DPO4000B and MDO3000 Series Oscilloscopes Programmer Manual 2-37

Command Groups
Table 2-33: Measurement Commands (cont.)
Command
Description
MEASUrement:METHod This command specifies the method used for calculating reference levels
MEASUrement:REFLevel? Returns the current reference level parameters
MEASUrement:REFLevel:ABSolute:
HIGH
This command specifies the top reference level for rise time
MEASUrement:REFLevel:ABSolute:
LOW
This command specifies the low reference level for rise time
MEASUrement:REFLevel:ABSolute:
MID<x>
This command specifies the mid reference level for the specified channel in absolute volts
MEASUrement:REFLevel:METHod This command specifies the method for assigning high and low reference levels
MEASUrement:REFLevel:PERCent:
HIGH
This command specifies the top reference percent level for rise time
MEASUrement:REFLevel:PERCent:
LOW
This command specifies the low reference percent level for rise time
MEASUrement:REFLevel:PERCent:
MID<x>
This command specifies the mid reference level for the specified channel in percent
MEASUrement:STATIstics Clears or returns all of the statistics accumulated for all period m easurements (MEAS1
through MEAS4)
MEASUrement:STATIstics:MODe Turns measurement statistics on or off
MEASUrement:STATIstics:
WEIghting
Controls the responsiveness of the mean and standard deviation to waveform changes
Miscellaneous Command Group
Use the commands in the Miscellaneous Command Group to perform actions that
do not fit into other categories.
Table 2-34: Miscellaneous Commands
Command
Description
APPLication:LICENSE:SLOT<x>:
LOCation?
This query returns the application license l ocation. < x> can be slot number 1–4.
APPLication:LICENSE:SLOT<x>:
TRANSFER
You can use this command to transfer an application license from the module to internal
memory in the oscilloscope, and transfer it back.
APPLication:LICENSE:SLOT<x>:
TYPe?
This query returns the application license type of the module that is currently inserted in the
specified application module slot.
APPLication:TYPe
When a m ask/limit or power test application module is installed, one of the associated test
types is always selected by default. This command allows the test type to be changed
from the default.
AUTOSet Sets the vertical, horizontal and trigger controls to provide a stable display of the appropriate
waveform. This is equivalent to pressing the front panel Autoset button
AUTOSet:ENAble Enables or disables the autoset feature
2-38 MDO4000/B/C, MSO/DPO4000B and MDO3000 Series Oscilloscopes Programmer Manual

Command Groups
Table 2-34: Miscellaneous Commands (cont.)
Command
Description
AUXOut:SOUrce This command specifies the source for the auxiliary-out port.
CLEARMenu Clears the current menu from the display
DATE
This command specifies the date displayed by the oscilloscope
*DDT This command specifies the commands that will be executed by the group execute trigger
DESkew Causes the deskew values for all channels to be set to the recommended values
DISplay:INTENSITy:B A CK Light:
AUTODim:TIMe
This command specifies the state of the deskew table display
ETHERnet:LXI:LAN:PASSWord:
ENABle
This command controls whether LXI (LAN eXtensions for Instrumentation) is password
protected.
ETHERnet:LXI:LAN:PASSWord:
ESCOPEENABle
This command controls whether to use the LXI password for e*Scope (effectively equal
to enabling password protection for e*Scope). With e*Scope, you can access any
Internet-connected MSO /DPO 4000B or MDO4000/B/C Series Oscilloscope from a web
browser.
ETHERnet:LXI:LAN:RESET
This com m and resets the LXI local area network. The items which this command reset
include: DHCP/BOOTP, mDNS and DNS-SD, e*S cope password protection, LXI password
protection, and e*Scope and LXI password.
ETHERnet:LXI:LAN:SERVICENAMe This command specifies the mDNS service name used for the LXI interface.
ETHERnet:LXI:LAN:STATus?
This query returns the LXI network status.
ETHERnet:NETWORKCONFig This command specifies the Ethernet network configuration setting.
FPAnel:HOLD This command is used to emulate the push-and-hold feature of the Cursor button.
FPAnel:PRESS Simulates the action of pressing a specified front-panel button
FPAnel:TURN
Simulates the action of turning a specified front-panel control knob
GPIBUsb:ADDress? Returns the current GPIB address
GPIBUsb:ID? Returns the identification string of the connected adaptor module and firmware version
HEADer|:HDR This command specifies the R esponse Header Enable State
ID? Returns the instrument identification data similar to that returned by the *IDN? IEEE488.2
common query, including the addition of any enabled application modules. However, it does
not include the instrument serial number.
*IDN? Returns the same information as the ID? command except the data is formatted according
to Tektronix Codes & Formats
LANGuage This command specifies the user interface display language
LOCk This command specifi es the front panel lock state
*LRN? Returns a listing of oscilloscope settings
NEWpass
Changes the password for user protected data
PASSWord Enables the *PUD and NEWpass set commands
PAUSe This command causes the interface to pause the specified number of seconds before
processing any other commands.
REBOOT Performs a reboot of the instrument after a short delay.
REM
Specifies a comment, which is ignored by the oscilloscope
MDO4000/B/C, MSO/DPO4000B and MDO3000 Series Oscilloscopes Programmer Manual 2-39