MDO3000 Programmer Manual.pdf - 第399页
Commands Listed in Alphabetical Order DS4XNA –D S 4 N AM ax Ou tput 13 8.26 Mb /s STS1Pulse – STS–1 Pulse 51.84 Mb/s STS1Eye – STS–1 Eye 51.84 Mb/s STS3 – STS–3 155.52 Mb/s STSX3 – STS–3 Max Output 155.52 Mb/s ITU-T : CL…

Commands Listed in Alphabetical Order
MASK:COPy:SO
Urce
This command specifies the mask that is to be copied to a user-defined custom
mask. The specified mask is not copied until after the
MASK:COPy:USER
command is ex
ecuted. Note that in addition to a standard mask,
LIMit, US ER
and NONe can also be specified.
NOTE. The only argument enumerations available for the MDO3000 series
models are
CUSTom | LIMit | NONe | USER
.
A series of examples showing how to use mask commands for typical tasks
is included in an appendix. (See page E-1, Mask/Limit Command Sequence
Examples.)
Conditions
This command requires a DPO4LMT application module for 4000 series models,
and an MDO3LMT application module for MDO3000 series models.
Group
Mask
Syntax
MASK:COPy:SOUrce
{DS1|DS1A|DS1C|DS2|DS3|DS4NA|DS4XNA|STS1Pulse|STS1Eye|STS3
|STSX3|CLOCKCoax|CLOCKSymmetrical|DSOContra|DSODouble
|DSOSingle|DSOTiming|G703DS1|DS1Rate|DS2RATECoax
|DS2RATESymmetrical|G703DS3|DS3Rate|E1Coax
|E1Symmetrical|E2|E3|HST1|HST2|HST3|HST4|HST5|HST6|NONe|USER
|LIMit}
MASK:COPy:SOUrce?
Related Commands
MASK:CUSTom
MASK:COPy:USER
Arguments
ANSI T1.102:
DS1 –DS11.544Mb/s
DS1A – DS1A 2.048 Mb/s
DS1C – DS1C 3.152 Mb/s
DS2 – DS2 6.312 Mb/s
DS3 –DS344.736Mb/s
DS4NA – DS4NA 139.26 Mb/s
2-368 MDO4000/B/C, MSO/DPO4000B and MDO3000 Series Oscilloscopes Programmer Manual

Commands Listed in Alphabetical Order
DS4XNA –DS4NAM
ax Output 138.26 Mb/s
STS1Pulse – STS–1 Pulse 51.84 Mb/s
STS1Eye – STS–1 Eye 51.84 Mb/s
STS3 – STS–3 155.52 Mb/s
STSX3 – STS–3 Max Output 155.52 Mb/s
ITU-T:
CLOCKCoax – Clock Interface Coaxial Pair 2.048 Mb/s
CLOCKSymmetrical – Clock Interface Symmetrical Pair 2.048 Mb/s
DSOContra – DSO Data Contradirectional 64 kb/s
DSODouble – DSO Double 64 kb/s
DSOSingle – DSO Single 64 kb/s
DSOTiming – DSO Timing 64 kb/s
G703DS1 – DS1 G.7031.544 Mb/s
DS1Rate –DS1OldRate1.544Mb/s
DS2RATECoax – DS2 Rate Coaxial Pair 6.312 Mb/s
DS2RATESymmetrical – DS2 Rate Symmetrical Pair 6.312 Mb/s
G703DS3 – DS3 G.703 44.736 Mb/s
DS3Rate – DS3 Old Rate 44.736 Mb/s
E1Coax –E1CoaxialPair2.048Mb/s
E1Symmetrical – E1 Symmetric Pair 2.048 Mb/s
E2 – E2 8.448 Mb/s
E3 – E3 34.368 Mb/s
USB:
HST1 – HS:T1 480 Mb/s
HST2 – HS:T1 480 Mb/s
HST3 – HS:T1 480 Mb/s
HST4 – HS:T1 480 Mb/s
HST5 – HS:T1 480 Mb/s
HST6 – HS:T1 480 Mb/s
In addition to the standard masks,
LIMit, U SER, and NONe can be input as well.
MDO4000/B/C, MSO/DPO4000B and MDO3000 Series Oscilloscopes Programmer Manual 2-369

Commands Listed in Alphabetical Order
Examples
MASK:COPy:SOU
rce
E1Coax specifies that the source for a user-defined custom
mask (to be created with the command
MASK:COPy:USER)istobeastandard
ITU-T E1 Coaxial Pair mask.
MASK:COPy:SOUrce? might return NONE, indicating that no source has been
specified.
MASK:COPy:
USER (No Query Form)
This creates a user-defined custom mask by making a copy of the source mask
that was specified prior with the command .MASK:COPy:SOUrce The source
mask coul
d be a standard, limit or another custom mask.
A series of examples showing how to use mask commands for typical tasks
is inclu
ded in an appendix. (See page E-1, Mask/Limit Command Sequence
Examples.)
Conditions
This command requires a DPO4LMT application module for 4000 series models,
and an MDO3LMT application module for MDO3000 series models.
Group
Mask
Syntax
MASK:COPy:USER
Related Commands
MASK:CUSTom
MASK:COPy:SOUrce
Examples
MASK:COPy:USER creates a user-defined custom mask from a copy of the source
mask.
MASK:COUNt (No Query Form)
This command resets to zero the number of hits and failures for all mask segments.
A series of examples showing how to use mask commands for typical tasks
is included in an appendix. (See page E-1, Mask/Limit Command Sequence
Examples.)
Conditions
This command requires a DPO4LMT application module for 4000 series models,
and an MDO3LMT application module for MDO3000 series models.
2-370 MDO4000/B/C, MSO/DPO4000B and MDO3000 Series Oscilloscopes Programmer Manual