MDO3000 Programmer Manual.pdf - 第543页

Commands Listed in Alphabetical Order Syntax POWer:SOA:MAS K:MAXVolts <NR3> POWer:SOA:MAS K:MAXVolts? Arguments <NR3> is a floating point number that spec ifies the maximum voltage applied to SOA mas k testing.…

100%1 / 1158
Commands Listed in Alphabetical Order
Arguments
<NR3> is a floati
ng point number that represents the SOA mask coordinates.
Examples
POWER:SOA:MASK:DEFINE 10 specifies the SOA mask coordinates as
0.0E+0,0.0E+0,0.0E+0,0.0E+0,0.0E+0,30.0000,25.0000,30.0000,300.0000,2.5000,
300.0000,2.4000,300.0000,2.2000,300.0000,1.9000,300.0000,1.3000,
300.0000,0.0E+0.
POWER:SOA:MASK:DEFINE ? might return :POWER:SOA:MAS K:DEFINE
0.0E+0,30.0000,25.0000,30.0000,300.0000,2.5000,300.0000,0.0E +0
indicating that the SOA mask coordinates are set to
0.0E+0,30.0000,25.0000,30.0000,300.0000,2.5000,300.0000,0.0E+0.
POWer:SOA:MASK:MAXAmps
This command specifies the maximum current applied to SOA mask testing.
Conditions
This command requires a DPO4PWR application module for 4000 series models,
or an MDO3PWR application module for MDO3000 series models.
Group
Power
Syntax
POWer:SOA:MASK:MAXAmps <NR3>
POWer:SOA:MASK:MAXAmps?
Arguments
<NR3> is a floating point number that specifies the maximum current applied
to SOA mask testing.
Examples
POWER:SOA:MASK:MAXAMPS 40 sets the SOA mask testing current to 40 A.
POWER:SOA:MASK:MAXAMPS ? might return :POWER:SOA:MASK:MAXAMPS
30.0000
indicating that the SOA mask testing maximum current is set to 30 A.
POWer:SOA:MASK:MAXVolts
This command specifies the maximum voltage applied to SOA mask testing.
Conditions
This command requires a DPO4PWR application module for 4000 series models,
or an MDO3PWR application module for MDO3000 series models.
Group
Power
2-512 MDO4000/B/C, MSO/DPO4000B and MDO3000 Series Oscilloscopes Programmer Manual
Commands Listed in Alphabetical Order
Syntax
POWer:SOA:MAS
K:MAXVolts <NR3>
POWer:SOA:MASK:MAXVolts?
Arguments
<NR3> is a floating point number that specifies the maximum voltage applied
to SOA mask testing.
Examples
POWER:SOA:MASK:MAXVOLTS 200 sets the SOA mask testing voltage to 200 V.
POWER:SOA:MASK:MAXVOLTS ? might return :POWER:SO A:MASK:MAXVOLTS
300.0000
indicating that the SOA mask testing maximum voltage is set to 300 V.
POWer:SOA:MASK:MAXWatts
This command specifies the maximum power applied to SOA mask testing.
Conditions
This command requires a DPO4PWR application module for 4000 series models,
or an MDO3PWR application module for MDO3000 series models.
Group
Power
Syntax
POWer:SOA:MASK:MAXWatts <NR3>
POWer:SOA:MASK:MAXWatts?
Arguments
<NR3> is a floating point number that specifies the maximum power applied to
SOA mask testing.
Examples
POWER:SOA:MASK:MAXWATTS 1000 sets the SOA mask testing power to
1.0000E+3 W.
POWER:SOA:MASK:MAXWATTS ? might return :POWER:SO A:MASK:MAXWATTS
750.0000
indicating that the SOA mask testing maximum power is set to 750 W.
POWer:SOA:MASK:NR_Pt? (Query Only)
Returns the number of mask points defined.
Conditions
This command requires a DPO4PWR application module for 4000 series models,
or an MDO3PWR application module for MDO3000 series models.
Group
Power
MDO4000/B/C, MSO/DPO4000B and MDO3000 Series Oscilloscopes Programmer Manual 2-513
Commands Listed in Alphabetical Order
Syntax
POWer:SOA:MAS
K:NR_Pt?
Examples
POWER:SOA:MASK:NR_PT? might return :POWER:SOA:MASK: NR_PT 4
indicating that the number of mask points defined is 4.
POWer:SOA:MASK:STATE
This command specifies the state of the mask for SOA calculations.
Conditions
This command requires a DPO4PWR application module for 4000 series models,
or an MDO3PWR application module for MDO3000 series models.
Group
Power
Syntax
POWer:SOA:MASK:STATE {OFF|LIMITS|POINTS}
POWer:SOA:MASK:STATE?
Arguments
OFF disables mask testing.
LIMITS enables mask testing based on limits specified using
POWer:SOA:MASK:MAXAmps, POWer:SOA:MASK:MAXVolts,and
POWer:SOA:MASK:MAXWatts commands
POINTS enables mask testing based on masks points defined.
Examples
POWER:SOA:MASK:STATEPOINTS enables mask testing power based on points
defined.
POWER:SOA:MASK:STATE? might return :POWER:SOA:MASK:STATE LIMITS
indicating that the mask testing is enabled based on limits.
POWer:SOA:MASK:STOPOnviol
This command specifies the enabled state of the mask stop on violation condition.
Conditions
This command requires a DPO4PWR application module for 4000 series models,
or an MDO3PWR application module for MDO3000 series models.
Group
Power
2-514 MDO4000/B/C, MSO/DPO4000B and MDO3000 Series Oscilloscopes Programmer Manual