Commands Listed in Alphabetical Order
OFF sets the val
ue to 0.
ON sets the value to 1.
Examples
TRIGger:A:BUS:B1:MIL1553B:STATus:BIT:ME NOCARE sets the status
word message error bit value to be used in the trigger to X.
TRIGger:A:BUS:B1:MIL1553B:STATus:BIT:ME? might return 1.
TRIGger:A:BUS:B<x>:MIL1553B:STATus:BIT:SRQ
When the MIL-STD-1553 bus trigger condition is set to STATus, this command
specifies the status word service request (SRQ) bit value (bit 11) to be used in the
trigger. Returned values are 0, 1, or X (don’t care, which is the default).
B<x>
Conditions
This command requires a DPO4AERO application module for DPO/MSO4000B
and MDO4000/B/C models or MDO3AERO application module for
MDO3000 models.
Group
Trigger
Syntax
TRIGger:A:BUS:B<x>:MIL1553B:STATus:BIT:SRQ
{0|1|X|ZERo|ONE|NOCARE|OFF|ON}
TRIGger:A:BUS:B<x>:MIL1553B:STATus:BIT:SRQ?
Related Commands
TRIGger:A:BUS:B<x>:MIL1553B:CONDition
Arguments
0
1
X
sets the value to X (“don’t care”) which is the default.
ZERO sets the value to 0.
ONE sets the value to 1.
NOCARE sets the value to X (“don’t care”) which is the default.
OFF sets the value to 0.
ON sets the value to 1.
Examples
TRIGger:A:BUS:B1:MIL1553B:STAT:BIT:SRQ NOCARE sets the status word
SRQ bit value to be used in the trigger to X.
2-848 MDO4000/B/C, MSO/DPO4000B and MDO3000 Series Oscilloscopes Programmer Manual
Commands Listed in Alphabetical Order
TRIGger:A:BUS
:B1:MIL1553B:STATUS:BIT:SRQ?
might return 1.
TRIGger:A:BUS:B<x>:MIL1553B:STATus:BIT:SUBSF
When the MIL-STD-1553 bus trigger condition is set to STATus, this command
specifies the status word subsystem flag bit value (bit 17) to be used in the trigger.
Returned values are 0, 1, or X (don’t care, which is the default).
B<x>
Conditions
This command requires a DPO4AERO application module for DPO/MSO4000B
and MDO4000/B/C models or MDO3
AERO application module for
MDO3000 models.
Group
Trigger
Syntax
TRIGger:A:BUS:B<x>:MIL1553B:STATus:BIT:SUBSF
{0|1|X|ZERo|ONE|NOCARE|OFF|ON}
TRIGger:A:BUS:B<x>:MIL1553B:STATus:BIT:SUBSF?
Related Commands
TRIGger:A:BUS:B<x>:MIL1553B:CONDition
Arguments
0
1
X
sets the value to X (“don’t care”) which is the default.
ZERO sets the value to 0.
ONE sets the value to 1.
NOCARE sets the value to X (“don’t care”) which is the default.
OFF sets the value to 0.
ON sets the value to 1.
Examples
TRIGger:A:BUS:B1:MIL1553B:STAT:BIT:SUBSF NOCARE sets the status
word subsystem flag bit value to be used in the trigger to X.
TRIGger:A:BUS:B1:MIL1553B:STAT:BIT:SUBSF? might return 1.
MDO4000/B/C, MSO/DPO4000B and MDO3000 Series Oscilloscopes Programmer Manual 2-849
Commands Listed in Alphabetical Order
TRIGger:A:BU
S:B<x>:MIL1553B:STATus:BIT:TF
When the MIL-STD-1553 bus trigger condition is set to STATus, this command
specifies the status word terminal flag bit value (bit 19) to be used in the trigger.
Returned val
ues are 0, 1, or X (don’t care, which is the default).
B<x>
Conditions
This command requires a DPO4AERO application module for DPO/MSO4000B
and MDO4000/B/C models or MDO3AERO application module for
MDO3000 models.
Group
Trigger
Syntax
TRI
Gger:A:BUS:B<x>:MIL1553B:STATus:BIT:TF
{0|1|X|ZERo|ONE|NOCARE|OFF|ON}
TR
IGger:A:BUS:B<x>:MIL1553B:STATus:BIT:TF?
Related Commands
TRIGger:A:BUS:B<x>:MIL1553B:CONDition
Arguments
0
1
X
sets the value to X (“don’t care”) which is the default.
ZERO sets the value to 0.
ONE sets the value to 1.
NOCARE sets the value to X (“don’t care”) which is the default.
OFF sets the value to 0.
ON sets the value to 1.
Examples
TRIGger:A:BUS:B1:MIL1553B:STAT:BIT:TF NOCARE sets the status word
terminal flag bit value to be used in the trigger to X.
TRIG:A:BUS:B1:MIL1553B:STAT:BIT:TF? might return 1.
TRIGger:A:BUS:B<x>:MIL1553B:STATus:PARity
When the MIL-STD-1553 bus trigger condition is set to STATus, this command
specifies the status parity bit value to be used in the trigger. Returned values are 0,
1, or X (don’t care, which is the default).
B<x>
2-850 MDO4000/B/C, MSO/DPO4000B and MDO3000 Series Oscilloscopes Programmer Manual