MDO3000 Programmer Manual.pdf - 第412页
Commands Listed in A lphabetical Order As e r i e so fe x a m ples showing how to use m ask commands for typical tasks is included in an appendix. (See page E-1, Mask/Limit Command Sequence Examples .) NOTE. T o ensur e …

Commands Listed in Alphabetical Order
SAVEWFM, MASK:
TESt:SAVEIMAGE, MASK:TESt:AUXout:FAILure,
MASK:TESt:HARDCopy,orMASK:TESt:SRQ:FAILure.
Aseriesofexa
mples showing how to use mask commands for typical tasks
is included in an appendix. (See page E-1, Mask/Limit Command Sequence
Examples.)
Conditions
This command requires a DPO4LMT application module for 4000 series models,
and an MDO3LMT application module for MDO3000 series models.
Group
Mask
Syntax
MASK:STOPOnviolation {ON|OFF|<NR1>}
MASK:STOPOnviolation?
Related Commands
MASK:TESt:SAVEWFM,
MASK:TESt:SAVEIMAGE,
MASK:TESt:AUXout:FAILure,
MASK:TESt:HARDCopy,
MASK:TESt:SRQ:FAILure
Arguments
ON stops waveform acquisition
on the first occurrence of a mask violation.
OFF turns this feature off.
<NR1>=0 turns this feature off ; any other value turns it on.
Examples
MASK:STOPONVIOLATION ON sets the instrument to stop waveform acquisition
on the first occurrence of a waveform violation.
MASK:STOPONVIOLATION? might return :MASK:STOPONVIOLATION ON,
indicating that the waveform acquisitions will be stopped upon the first occurrence
of a waveform violation.
MASK:TEMPLate:CREATEmask (No Query Form)
This command, used for limit mask testing, causes a template mask to be created
based on the settings of the MASK:TEMPLate:SOUrce, MASK:TEMPLate:
TOLerance:HORizontal,andMASK:TEMPLate:TOLerance:VERTical
commands. If you do not specify horizontal or vertical tolerances, the default
tolerance of one minor division will be used.
MDO4000/B/C, MSO/DPO4000B and MDO3000 Series Oscilloscopes Programmer Manual 2-381

Commands Listed in Alphabetical Order
Aseriesofexam
ples showing how to use mask commands for typical tasks
is included in an appendix. (See page E-1, Mask/Limit Command Sequence
Examples.)
NOTE. To ensure that Mask commands are enabled, use the command
APPLication:TYPe LIMITMask.
Conditions
This comma
nd requires a DPO4LMT application module for 4000 series models,
and an MDO3LMT application module for MDO3000 series models.
Group
Mask
Syntax
MASK:TEMPLate:CREATEmask
Related Commands
MASK:TEMPLate:SOUrce,
MASK:TEMPLate:TOLerance:HORizontal,
MASK:TEMPLate:TOLerance:VERTical,
SAVe:MASK
MASK:TEMPLate:SOUrce
This command, typically used for limit mask testing, specifies the source
waveform to be used to create a template mask. The source can be either
one of four channels, or one of four saved reference waveforms. See the
SAVe:WAVEform commands for more information on creating reference
waveforms.
A series of examples showing how to use mask commands for typical tasks
is included in an appendix. (See pag
eE-1,Mask/Limit Command Sequence
Examples.)
NOTE. To ensure that Mask commands are enabled, use the command
APPLication:TYPe LIMITMask.
Conditions
This command requires a DPO4LMT application module for 4000 series models,
andanMDO3LMTapplicationmodu
le for MDO3000 series models.
Group
Mask
2-382 MDO4000/B/C, MSO/DPO4000B and MDO3000 Series Oscilloscopes Programmer Manual

Commands Listed in Alphabetical Order
Syntax
MASK:TEMPLate
:SOUrce {C H1|CH2|CH3|CH4|REF1|REF2|REF3|REF4}
MASK:TEMPLate:SOUrce?
Related Commands
MASK:TEMPLate:CREATEmask,
MASK:TEMPLate:TOLerance:HORizontal,
MASK:TEMPLate:TOLerance:VERTical,
SAVe:MASK
MASK:TEMPLate:TOLerance:HORizontal
This command specifies the horizontal limit (tolerance) for a template maskto
be used for limit testing. This indicates how far horizontally from the template
source to create the mask.
A series of examples showing how to use mask commands for typical tasks
is included in an appendix. (See page E-1, Mask/Limit Command Sequence
Examples.)
Conditions
This command requires a DPO4LMT application module for 4000 series models,
and an MDO3LMT application module for MDO3000 series models.
Group
Mask
Syntax
MASK:TEMPLate:TOLerance:HORizontal <NR3>
MASK:TEMPLate:TOLerance:HORizontal?
Related Commands
MASK:TEMPLate:CREATEmask,
MASK:TEMPLate:TOLerance:VERTical,
SAVe:MASK
Arguments
<NR3> is a floating point value that specifies the horizontal limit (tolerence) in
milli-divisions. Range: 0 to 500 mDiv, resolution: 1mDiv, default: 40mDiv.
This value cannot be negative.
Examples
If you create a limit mask after a default setup, then query:
:MASK:TEMPLate:TOLerance:HORizontal?
you get the response: 200.000E-3
Now send the commands:
MDO4000/B/C, MSO/DPO4000B and MDO3000 Series Oscilloscopes Programmer Manual 2-383