MDO3000 Programmer Manual.pdf - 第1131页
Appendix H: List of Beta Commands and Features Examples: AFG:MODULatio n:AM:DEPth 10.0 sets the modulation depth to 10%. AFG:MODULation:AM:FREQuency Description: Sets the modulation frequency to use f or Amplitude Modula…

Appendix H: List of Beta Commands and Features
Tektronixtoco
mply with any additional obligations with respect to any Tektronix
products or services that incorporate any feedback.
Command List
TableH-1:PresentlyAvailableBetaCommands
Feature
Associated Commands
AFG run commands AFG:RUNMode
AFG trigger
commands
AFG:TRIGger
AFG:TRIGger:MODe
AFG:TRIGger:SOUrce
AFG:TRIGger:POLarity
AFG:TRIGger:COUnt
AFG:TRIGger:INTerval
AFG:TRIGger:DELay
AFG modulation
commands
AFG:MODULation:TYPeAFG:MODULation:TYPe
AFG:MODULation:AM:WAVeform
AFG:MODULation:AM:FREQuency
AFG:MODULation:AM:DEPth
AFG:MODULation:FM:WAVeform
AFG:MODULation:FM:FREQuency
AFG:MODULation:FM:DEViation
Beta Commands Listed in A lphabetical Order
AFG:MODULation:AM:DEPth
Description:
Sets the modulation depth to use for Amplitude Modulation for the AFG.
Con
ditions:
Beta Command, see conditions of use. Default setting is 50.0.
Group:
AFG
Syntax:
AFG:MODULation:AM:DEPth <NR3>
AFG:MODULation:AM:DEPth?
Arguments:
<NR3> - Depth value in percent from 0.0 to 100.0 with a resolution of 0.1%.
H-2 MDO4000/B/C, MSO/DPO4000B and MDO3000 Series Oscilloscopes Programmer Manual

Appendix H: List of Beta Commands and Features
Examples:
AFG:MODULatio
n:AM:DEPth
10.0 sets the modulation depth to 10%.
AFG:MODULation:AM:FREQuency
Description:
Sets the modulation frequency to use for Amplitude Modulation for the AFG.
Conditions:
Beta Command, see conditions of use. Default setting is 10.0E+3.
Group:
AFG
Syntax:
AFG:MODULation:AM:FREQuency <NR3>
AFG:MODULation:AM:FREQuency?
Arguments:
<NR3> - Frequency value for modulation in Hz from 200.0E-3 to 50.0E+3 with a
resolution of 100.0E-3.
Example
s:
AFG:MODULation:AM:FREQuency 10.0E+3 sets the modulation frequency to
10kHz.
AFG:MODULation:AM:WAVeform
Description:
Sets the modulation shape to use for Amplitude Modulation for the AFG.
Conditions:
Beta Command, see conditions of use. Default setting is SINE.
Group:
AFG
Syntax:
AFG:MODULation:AM:WAVeform
{SINE|SQUare|TRIangle|RAMP|NRAMP|NOISe}
AFG:MODULation:AM:WAVeform?
Arguments:
SINE -Sinewaveform
SQUare - Square waveform
TRIangle - Triangle waveform
RAMP - Positive ramp waveform
MDO4000/B/C, MSO/DPO4000B and MDO3000 Series Oscilloscopes Programmer Manual H-3

Appendix H: List of Beta Commands and Features
NRAMP - Negativ
e ramp waveform
NOISe - Noise waveform
Examples:
AFG:MODULation:AM:WAVeform SINE sets the modulation waveform to SINE.
AFG:MODULation:FM:DEViation
Description:
Sets the mod
ulation deviation to use for Frequency Modulation for the
AFG. This value is the deviation around the center frequency specified in
AFG:MODULation:FM:FREQuency
.
Conditions:
Beta Comm
and, see conditions of use
. Default setting is 50.0.
Group:
AFG
Syntax:
AFG:MODULation:FM:DEViation <NR3>
AFG:MODULation:FM:DEViation?
Arguments:
<NR3> - Deviation value in Hz from 0.0 to 25.0E+6 with a resolution of 0.1Hz.
Maximum value is half the frequency maximum specified by the shape chosen
using AFG:FUNCtion.
Examples:
AFG:MODULation:FM:DEViation 10.0E+3 sets the modulation depth to 10kHz.
AFG:MODULation:FM:FREQuency
Description:
Sets the modulation frequency to use for Frequency Modulation for the AFG.
Conditions:
Beta Command, see conditions of use. Default setting is 10.0E+3.
Group:
AFG
Syntax:
AFG:MODULation:FM:FREQuency <NR3>
AFG:MODULation:FM:FREQuency?
H-4 MDO4000/B/C, MSO/DPO4000B and MDO3000 Series Oscilloscopes Programmer Manual