MDO3000 Programmer Manual.pdf - 第245页
Commands Listed in Alphabetical Order common, howeve r , to use uncalibrated probes (pa rticularly when the oscilloscope inputs are connected into a tes t system with coaxial cables). Group Calibration and Diagnostic Syn…

Commands Listed in Alphabetical Order
Returns
This query will
return one of the following:
INIT indicates the RF portion of the instrument has not been calibrated.
PASS indicates the RF internal calibration completed successfully.
RUNNING indicates the oscilloscope internal calibration is currently running.
FAIL indicates the RF internal calibration did not complete successfully.
Examples
CAL:INTERN:STAT:RF? might return INIT indicating that the RF portion of the
instrument has not been internally calibrated.
CALibrate:INTERNal:STATus:SCOPE? (Query Only)
This query returns the status of the last SPC run for the oscilloscope portion of
the instrument (doesn't include the RF portion). To query the status of the RF
portions, use CALibrate:INTERNal:STATus:RF?
This query is synonymous to CALibrate:RESults:SPC:SCOPE?
Group
Calibration and Diagnostic
Syntax
CALibrate:INTERNal:STATus:SCOPE?
Returns
Th
is query will return one of the following:
INIT indicates the oscilloscope portion of the instrument has not been calibrated.
PASS indicates the oscilloscope internal calibration completed successfully.
RUNNING indicates the oscilloscope internal calibration is currently running.
FAIL indicates the oscilloscope internal calibration did not complete successfully.
Examples
CAL:INTERN:STAT:SCOPE? might return INIT indicating that the oscilloscope
portion of the instrument has not been calibrated.
CALibrate:RESults? (Query Only)
Returns the status of internal and factory calibrations, without performing any
calibration operations. The results returned do not include the calibration status
of attached probes. The
CALibrate:RESults? query is intended to support
GO/NoGO testing of the oscilloscope calibration readiness: all returned results
should indicate PASS status if the oscilloscope is "fit for duty". It is quite
2-214 MDO4000/B/C, MSO/DPO4000B and MDO3000 Series Oscilloscopes Programmer Manual

Commands Listed in Alphabetical Order
common, howeve
r, to use uncalibrated probes (particularly when the oscilloscope
inputs are connected into a test system with coaxial cables).
Group
Calibration
and Diagnostic
Syntax
CALibrate:RESults?
Related Commands
*CAL?
CALibrate:RESults:FACtory? (Query Only)
Returns the status of internal and factory calibration, without performing any
calibration operations.
Group
Calibration and Diagnostic
Syntax
CALibrate:RESults:FACtory?
CALibrate:RESults:FACtory:AFG? (Query Only)
This query returns the factory calibration status for the Arbitrary Function
Generator portion of the instrument, if present.
This query is synonymous with
CALibrate:FACtory:STATus:AFG?
Conditions
Available for MDO3000 and MDO4000C series only.
Group
Calibration and Diagnostic
Syntax
CALibrate:RESults:FACtory:AFG?
Returns
PASS - factory calibration has passed
INIT - factory calibration is unadjusted
RUNNING - factory calibration is running
Examples
CALIBRATE:RESULTS:FACTORY:AFG? might return RUNNING, indicating that
the factory calibration for the AFG portion of the instrument is in progress.
MDO4000/B/C, MSO/DPO4000B and MDO3000 Series Oscilloscopes Programmer Manual 2-215

Commands Listed in Alphabetical Order
CALibrate:RE
Sults:FACtory:RF? (Query Only)
This query returns the factory calibration status for the RF portion of the
instrument, if present.
This query is synonymous with
CALibrate:FACtory:STATus:RF?
Group
Calibration and Diagnostic
Syntax
CALibrate:RESults:FACtory:RF?
Returns
PASS - factory calibration has passed
INIT - factory calibration is unadjusted
RUNNING - factory calibration is in progress
Examples
CALIBRATE:RESULTS:FACTORY:RF? might return RUNNING, indicating that the
factory calibration for the RF portion of the instrument is in progress.
CALibrate:RESults:FACtory:SCOPE? (Query Only)
This query returns the factory calibration status for the oscilloscope (doesn't
include RF or AFG) of the instrument.
This query is synonymous with the following query:
:CALibrateFACtory:STATus:SCOPE?
Group
Calibration and Diagnostic
Syntax
CALibrate:RESults:FACtory:SCOPE?
Returns
This query will return one of the following:
INIT indicates the oscilloscope portion of the instrument has not been calibrated.
PASS indicates the oscilloscope internal calibration completed successfully.
RUNNING indicates the oscilloscope internal calibration is currently running.
Examples
CAL:RESULTS:SPC:SCOPE? might return INIT indicating that the oscilloscope
portion of the instrument has not been calibrated.
2-216 MDO4000/B/C, MSO/DPO4000B and MDO3000 Series Oscilloscopes Programmer Manual