MDO3000 Programmer Manual.pdf - 第423页
Commands Listed in Alphabetical Order <NR1> is an inte ger . 0 turns off this feature; any other integer turns it on. Examples MASK:TEST:SRQ :FAILURE? might return :M ASK:TEST:SRQ:FA ILURE 1 , indicating that the p…

Commands Listed in Alphabetical Order
Related Commands
MASK:TESt:STA
TE,
MASK:TESt:COMPLetion:CRITerion
Arguments
ON turns on this feature, so that upon the completion of a mask test, the instrument
will send an SRQ command (if registers are set to send SRQ when OPC is
asserted. (See page 3-1, Registers.))
OFF turns off this feature.
<NR1> is an integer. 0 turns off this feature; any other integer turns it on.
Examples
MASK:TESt:SRQ:COMPLetion ON sends an SRQ command upon test
completion.
MASK:TESt:SRQ:COMPLetion? might return 0, indicating the instrument is not
settosendanSRQcommandupon test completion.
MASK:TESt:SRQ:FAILure
This command causes the instrument to send an SRQ command when a pass/fail
mask test fails.
A series of examples showing how to use mask commands for typical tasks
is included in an appendix. (See page E-1, Mask/Limit Command Sequence
Examples.)
Conditions
This command requires a DPO4LMT application module for 4000 series models,
andanMDO3LMTapplicat
ion module for MDO3000 series models.
Group
Mask
Syntax
MASK:TESt:SRQ:FAILure {ON|OFF|<NR1>}
MASK:TESt:SRQ:FAILure?
Related Commands
MASK:TESt:STATE,
MASK:TESt:COMPLetion:CRITerion
Arguments
ON turns on this feature, so that when a mask test fails, the instrument will send
an SRQ command (if registers are set to send SRQ when OPC is asserted. (See
page 3-1, Registers.)).
OFF turns off this feature.
2-392 MDO4000/B/C, MSO/DPO4000B and MDO3000 Series Oscilloscopes Programmer Manual

Commands Listed in Alphabetical Order
<NR1> is an inte
ger. 0 turns off this feature; any other integer turns it on.
Examples
MASK:TEST:SRQ:FAILURE? might return :MASK:TEST:SRQ:FAILURE 1,
indicating that the pass/fail SRQ on failure is turned on.
MASK:TESt:STATE
This command turns the pass/fail mask test on or off. Most of the other
MASK:TEST commands need to be executed before this command.
NOTE. To ensure that Mask commands are enabled, use the command
APPLication:TYPe LIMITMask.
A series of examples showing how to use mask commands for typical tasks
is included in an appendix. (See page E-1, Mask/Limit Command Sequence
Examples.)
Conditions
This command requires a DPO4LMT application module for 4000 series models,
and an MDO3LMT application module for MDO3000 series models.
Group
Mask
Syntax
MASK:TESt:STATE {ON|OFF|<NR1>}
MASK:TESt:STATE?
Related Commands
MASK:TESt:WAVEform
Arguments
ON turns the mask test on.
OFF turns the mask test off.
<NR1> is an integer. 0 turns the mask test off; any other integer turns it on
.
Examples
MASK:TESt:STATE ON starts the mask test.
MASK:TESt:STATus? (Query Only)
This command indicates the result of a pass/fail mask test.
MDO4000/B/C, MSO/DPO4000B and MDO3000 Series Oscilloscopes Programmer Manual 2-393

Commands Listed in Alphabetical Order
Aseriesofexam
ples showing how to use mask commands for typical tasks
is included in an appendix. (See page E-1, Mask/Limit Command Sequence
Examples.)
Conditions
This command requires a DPO4LMT application module for 4000 series models,
and an MDO3LMT application module for MDO3000 series models.
Group
Mask
Syntax
MASK:TESt:STATus?
Related Commands
MASK:TESt:STATE,
Returns
This query returns one of: OFF, DELAY, PASSING, FAILING, PASSED, o r
FAILED,
When the violation count exceeds the violation threshold during the
test, the status changes from
PASSING to FAILING. When the test is complete,
it changes from
FAILING to FAILED.
Examples
MASK:TESt:STAT? might r
eturn PASSED indicating the test passed and is
complete.
MASK:TESt:STOP:FAILure
This command causes the instrument to stop acquiring data when a pass/fail mask
test fails.
A series of examples showing how to use mask commands for typical tasks
is included in an appendix. (See page E-1, Mask/Limit Command Sequence
Examples.)
Conditions
This command requires a DPO4LMT application module for 4000 series models,
and an MDO3LMT application module for MDO3000 series models.
Group
Mask
Syntax
MASK:TESt:STOP:FAILure {|ON|OFF|<NR1>}
MASK:TESt:STOP:FAILure?
Related Commands
MASK:TESt:STATE,
2-394 MDO4000/B/C, MSO/DPO4000B and MDO3000 Series Oscilloscopes Programmer Manual