Commands Listed in Alphabetical Order
TRIGger:A:BU
S:B<x>:FLEXray:DATa:SIZe
This command specifies the length of the data string, in bytes, when triggering
on the FlexRay bus data field. The trigger condition needs to be set to ID or
IDANDDATA (u
sing TRIGger:A:BUS:B<x>:FLEXray:CONDition).
B<x>
Group
Trigger
Syntax
TRIGger:A:BUS:B<x>:FLEXray:DATa:SIZe <NR1>
TRIGger:A:BUS:B<x>:FLEXray:DATa:SIZe?
Arguments
<NR1> is the FlexRay data string length, in bytes.
Examples
TRIGGER:A:BUS:B1:FLEXRAY:DATA:SIZE 8 sets the data string size to 8
bytes.
TRIGGER:A:BUS:B1:FLEXRAY:DATA:SIZE? might return
TRIGGER:A:BUS:B1:FLEXRAY:DATA:SIZE 1 indicating the data
size is 1 byte.
TRIGger:A:BUS:B<x>:FLEXray:DATa:VALue
This command specifies the low value when triggering on the FlexRay bus data
field. (Use TRIGger:A:BUS:B<x>:FLEXray:DATa:HIVALue to set the upper
value.) The trigger condition needs to be set to ID or IDANDDATA (using
TRIGger:A:BUS:B<x>:FLEXray:CONDition).
B<x>
Conditions
This command requires a DPO4AUTOMAX application module for
DPO/MSO4000B and MDO4000/B/C models or MDO3FLEX application module
for MDO3000 models.
Group
Trigger
Syntax
TRIGger:A:BUS:B<x>:FLEXray:DATa:VALue <QString>
TRIGger:A:BUS:B<x>:FLEXray:DATa:VALue?
Arguments
<QString> is a quoted string.
Examples
TRIGGER:A:BUS:B1:FLEXRAY:DATA:VALU
E “1100 1101”
sets the FlexRay
data value for triggering to 11001101.
2-814 MDO4000/B/C, MSO/DPO4000B and MDO3000 Series Oscilloscopes Programmer Manual

Commands Listed in Alphabetical Order
TRIGGER:A:BUS
:B1:FLEXRAY:DATA:VALUE?
might return
TRIGGER:A:BUS:B1:FLEXRAY:DATA:VALUE "XXXXXXXX" indicating the
FlexRay data value is don't care.
TRIGger:A:B
US:B<x>:FLEXray:EOFTYPE
This command specifies the end of file type (static, dynamic or any) when
triggering on the FlexRay bus EOF field. The trigger condition needs to be setto
EOF (using
TRIGger:A:BUS:B<x>:FLEXray:CONDition).
B<x>
Conditions
This command requires a DPO4AUTOMAX application mo
dule for
DPO/MSO4000B and MDO4000/B/C models or MDO3FLEX application module
for MDO3000 models.
Group
Trigger
Syntax
TRIGger:A:BUS:B<x>:FLEXray:EOFTYPE {STATic|DYNAMic|ANY}
TRIGger:A:BUS:B<x>:FLEXray:EOFTYPE?
Arguments
STATic specifies triggering on the STATIC end of file type.
DYNAMic specifies triggering on the DYNAMIC end of file type.
ANY specifies triggering on a STATIC or DYNAMIC end of file type.
Examples
TRIGGER:A:BUS:B1:FLEXRAY:EOFTYPE ANY sets the FlexRay end of file
type to ANY.
TRIGGER:A:BUS:B1:FLEXRAY:EOFTYPE? might return
TRIGGER:A:BUS:B1:FLEXRAY:EOFTYPE STATIC indicating the
FlexRay end of file type is STATIC
TRIGger:A:BUS:B<x>:FLEXray:ERRTYPE
This command specifies the error type when triggering on the FlexRay
bus signal. The trigger condition needs to be set to ERROR (using
TRIGger:A:BUS:B<x>:FLEXray:CONDition).
B<x>
Conditions
This command requires a DPO4AUTOMAX application module for
DPO/MSO4000B and MDO4000/B/C models or MDO3FLEX application module
for MDO3000 models.
MDO4000/B/C, MSO/DPO4000B and MDO3000 Series Oscilloscopes Programmer Manual 2-815

Commands Listed in Alphabetical Order
Group
Trigger
Syntax
TRIGger:A:BUS:B<x>:FLEXray:ERRTYPE
{CRCHeader|CRCTrailer|SYNCFrame|STARTupnosync|NULLFRStatic|
NULLFRDynamic}
TRIGger:A:BUS:B<x>:FLEXray:ERRTYPE?
Arguments
CRCHeader sets the error type to CRCHeader.
CRCTrailer sets the error type to CRCTrailer.
SYNCFrame sets the error type to SYN
CFrame.
STARTupnosync sets the error type to STARTupnosync.
NULLFRStatic sets the error type to NULLFRStatic.
NULLFRDynamic sets the error type to NULLFRDynamic.
Examples
TRIGGER:A:BUS:B1:FLEXRAY:ERRTYPE SYNCFRAME sets the trigger type
is SYNCFRAME.
TRIGGER:A:BUS:B1:FLEXRAY:ERRTYPE? might return
TRIGGER:A:BUS:B1:FLEXRAY:ERRTYPE CRCHEADER indicating the
FlexRay trigger type is CRCHeader.
TRIGger:A:BUS:B<x>:FLEXray:FRAMEID:HIVALue
This command specifies the high value when triggering on the FlexRay bus
frame ID field. (Use TRIGger:A:BUS:B<x>:FLEXray:FRAMEID:VALue to
set the low value.) The trigger condition needs to be set to IDentifier (using
TRIGger:A:BUS:B<x>:FLEXray:CONDition).
B<x>
Conditions
This command requires a DPO4AUTOMAX application module for
DPO/MSO4000B and MDO4000/B/C models or MDO3FLEX application module
for MDO3000 models.
Group
Trigger
Syntax
TRIGger:A:BUS:B<x>:FLEXray:FRAMEID:HIVALue <QString>
TRIGger:A:BUS:B<x>:FLEXray:FRAMEID:HIVALue?
Arguments
<QString> is a quoted string that is the binary frame ID high value.
2-816 MDO4000/B/C, MSO/DPO4000B and MDO3000 Series Oscilloscopes Programmer Manual