MDO3000 Programmer Manual.pdf - 第786页
Commands Listed in A lphabetical Order |D5|D6|D7|D8| D9|D10|D11|D12|D 13|D14|D15|RF_A MPlitude|RF_FREQ uency |RF_PHASe|RF_ NORMal|RF_AVErag e| RF_MAXHol d|RF_MINHold} SELect:CONTR Ol? Arguments CH1–4 specifies a channel w…

Commands Listed in Alphabetical Order
Group
Vertical
Syntax
SELect:{BUS<x>|B<x>} {OFF|ON|0|1}
SELect:{BUS<x>|B<x>}?
Arguments
ON or 1 turns the specified waveform display on.
OFF or 0 turns the specified waveform display off.
SELect:CH<x>
Turns the display of the channel <x> waveform on or off, where <x > is the
channel number, which can be 1–4. This command also resets the acquisition.
The query returns whether the channel is on or off but does not indicate whether
it is the specified waveform.
Group
Vertical
Syntax
SELect:CH<x> {ON|OFF|1| 0}
SELect:CH<x>?
Arguments
ON or 1 turns on the display of the specified waveform. This waveform also
becomes the selected waveform.
OFF or 0 turns off the display of the specified waveform.
Examples
SELECT:CH2 ON turns the channel 2 waveform display on, and selects channel 2.
SELECT:CH1? might return :SELECT:CH1 1 indicating that channel 1 is being
displayed.
SELect:CONTROl
This command selects a waveform to be used with channel-related commands
such as the cursor commands. In addition to selecting the waveform, it turnson
the display of the waveform, if it wasn’t on already.
Group
Vertical
Syntax
SELect:CONTROl {CH1|CH2|CH3|CH4
|MATH|REF1|REF2|REF3|REF4|BUS1|BUS2|BUS3|BUS4|D0|D1|D2|D3|D4
MDO4000/B/C, MSO/DPO4000B and MDO3000 Series Oscilloscopes Programmer Manual 2-755

Commands Listed in Alphabetical Order
|D5|D6|D7|D8|
D9|D10|D11|D12|D13|D14|D15|RF_AMPlitude|RF_FREQuency
|RF_PHASe|RF_NORMal|RF_AVErage| RF_MAXHold|RF_MINHold}
SELect:CONTR
Ol?
Arguments
CH1–4 specifies a channel waveform as the waveform affected by the front-panel
controls.
MATH specifies the math waveform as the waveform that is affected by the
front-panel controls.
BUS1–4 specifies a bus waveform as the waveform affected by the front-panel
controls.
D0–D15 specifies a digital waveform as the waveform affected by the front-panel
controls. (MSO/MDO4000/B models only as well as MDO3000 and MDO400C
models with option MDO3MSO or MDO4MSO installed.)
|
RF_AMPlitude|RF_FREQuency|RF_PHASe (MDO4000/B/C models only)
and
RF_NORMal|RF_AVErage|RF_MAXHold| RF_MINHold specify an RF
trace as the waveform affected by the front-panel controls.
Returns
NONE if all the channels are turned off. NONE is ignored on input.
Examples
SELECT:CONTR
OL CH2
resets acquisition displays on channel 2, and causes the
selected waveform to be the implied object of waveform commands.
SELECT:CONT
ROL?
might return :SELECT:CONTROL MATH indicating that math
is the implied object of waveform commands.
SELect:DAll (No Query Form)
This command turns on or off all digital channels (D0 — D15).
NOTE. This command applies only to MSO/MDO4000/B models only as well
as MDO3000 and MDO400C models with option MDO3MSO or MDO4MSO
installe
d.
Group
Vertical
Syntax
SELect:DAll {ON|OFF|0| 1}
2-756 MDO4000/B/C, MSO/DPO4000B and MDO3000 Series Oscilloscopes Programmer Manual

Commands Listed in Alphabetical Order
Related Commands
SELect:CONTRO
l:SELect:CONTROl:SELect:D<0-15>
SELect:D<x>
Arguments
This command takes a Boolean argument, which means it can be used to either
turn all digital channels on or all digital channels off.
Examples
SELect:DAll 1 turns on all digital channels.
SELect:D<x>
Turns on the display of the digital channel <x> and resets the acquisition. <x > is
the channel number, which can be 0–15. The query returns whether the channelis
on or off but does not indicate whether it is the selected waveform.
Group
Vertical
Syntax
SELect:D<x> {<NR1>|OFF| ON}
SELect:D<x>?
Arguments
ON turns on the display of the specified waveform. This waveform also becomes
the selected waveform.
OFF turns off the display of the specified waveform.
<NR1> = 0 turns off the display of the specified waveform; any other value turns
on the display of the specified waveform.
Examples
SELECT:D2 ON turns the digital channel 2 waveform display on, and selects
digital channel 2.
SELECT:D2? might return : SELECT:D2 1 indicating that digital channel 2 is
being displayed.
SELect{:MATH|:MATH1}
Turns on and off the display of the math waveform. The query returns whether
the math waveform is on or off but does not indicate whether it is the selected
waveform.
Group
Vertical
MDO4000/B/C, MSO/DPO4000B and MDO3000 Series Oscilloscopes Programmer Manual 2-757