MDO3000 Programmer Manual.pdf - 第730页
Commands Listed in A lphabetical Order Group Search Syntax SEARCH:SEARCH <x>:TRIGger:A:BU S:B<x>:SPI:DATa {:MOSI|:OUT}: VALue < bin > SEARCH:SEARCH <x>:TRIGger:A:BU S:B<x>:SPI:DATa {:MOSI|:O…

Commands Listed in Alphabetical Order
Arguments
SS specifies a se
arch based on the Slave Selection condition.
STARTofframe is applicable when BUS: B<x>:SPI:FRAMING is set to
IDLEtime. When the search condition is set to STARTofframe,theinstrument
searches on the first SPI clock after an idle time when there are no clocks.
MISO specifies a search based on the Master-In Slave-Out condition.
MOSI specifies a search based on the Master-Out Slave-In condition.
MISOMOSI specifies a search based on the Master-In Slave-Out and Master-Out
Slave-In conditions.
SEARCH:SEARCH<x>:TRIGger:A:BUS:B<x>:SPI:DATa{:MISO|:IN}:VALue
This command specifies the data in value to be used in an SPI search, if the search
condition is set to
MISO or MISOMOSI. SEARCH<x> is the search number, which
is always 1, and
B<x>
Conditions
This command requires a DPO4EMBD application module for DPO/MSO4000B
and MDO4000/B/C models or MDO3EMBD application module for
MDO3000 models.
Group
Search
Syntax
SEARCH:SEARCH<x>:TRIGger:A:BUS:B<x>:SPI:DATa{:MISO|:IN}:
VALue < bin>
SEARCH:SEARCH<x>:TRIGger:A:BUS:B<x>:SPI:DATa{:MISO|:IN}:
VALue?
Related Commands
SEARCH:SEARCH<x>:TRIGger:A:BUS:B<x>:SPI:CONDition
Arguments
<bin> is the data string in binary format.
SEARCH:SEARCH<x>:TRIGger:A:BUS:B<x>:SPI:DATa{:MOSI|:OUT}:VALue
This command specifies the data out value to be used in an SPI search if the
search the condition is set to
MOSI,orMISOMOSI. SEARCH<x> is the search
number, which is always 1, and
B<x>
Conditions
This command requires a DPO4EMBD application module for DPO/MSO4000B
and MDO4000/B/C models or MDO3EMBD application module for
MDO3000 models.
MDO4000/B/C, MSO/DPO4000B and MDO3000 Series Oscilloscopes Programmer Manual 2-699

Commands Listed in Alphabetical Order
Group
Search
Syntax
SEARCH:SEARCH<x>:TRIGger:A:BU
S:B<x>:SPI:DATa{:MOSI|:OUT}:
VALue < bin>
SEARCH:SEARCH<x>:TRIGger:A:BUS:B<x>:SPI:DATa{:MOSI|:OUT}:
VALue?
Related Commands
SEARCH:SEARCH<x>:TRIGger:A:BUS:B<x>:SPI:CONDition
Arguments
<bin> is the data in binary format.
SEARCH:SEARCH<x>:TRIGger:A:BUS:B<x>:SPI:D
ATa:SIZe
ThiscommandspecifiesthelengthofthedatastringtobeusedinanSPIsearch,
if the search condition is
MISO, MOSI,orMISO MOSI. SEARCH<x> is the search
number, which is always 1, and
B<x>
Conditions
This command requires a DPO4EMBD application module for DPO/MSO4000B
and MDO4000/Bmodels or MDO3EMBD application module for
MDO3000 models.
Group
Search
Syntax
SEARCH:SEARCH<x>:TRIGger:A:BUS:B<x>:SPI:DATa:SIZe <NR1>
SEARCH:SEARCH<x>:TRIGger:A:BUS:B<x>:SPI:DATa:SIZe?
Related Commands
SEARCH:SEARCH<x>:TRIGger:A:BUS:B<x>:SPI:CONDition
Arguments
<NR1> is the data string length in bytes.
SEARCH:SEARCH<x>:TRIGger:A:BUS:B<x>:USB:ADDRess:HIVALue
This command specifies the upper limit of the address string for the inside-of-range
and outside-of-range qualifiers to be used in aUSB search. Use the command
SEARCH:SEARCH<x>:TRIGger:A:BUS:B<x>:USB:ADDRess:VALue to set
the lower limit.
SEARCH<x> is the search number, which is always 1, and B<x>
2-700 MDO4000/B/C, MSO/DPO4000B and MDO3000 Series Oscilloscopes Programmer Manual

Commands Listed in Alphabetical Order
Conditions
This command re
quires the DPO4USB application module to be installed, except
for MDO3000 models, which require the MDO3USB. Note that 1 GHz bandwidth
models are recommended for high-speed (HS) USB.
Group
Search
Syntax
SEARCH:SEARCH<x>:TRIGger:A:BUS:B<x>:USB:ADDRess:HIVALue
<QString>
SEARCH:SEARCH<x>:TRIGger:A:BUS:B<x>:USB:ADDRess:HIVALue?
Related Commands
SEARCH:SEARCH<x>:TRIGger:A:BUS:B<x>:USB:CONDition,
SEARCH:SEARCH<x>:TRIGger:A:BUS:B<x>:USB:ADDRess:VALue
Arguments
<QString> within the range 0000000 to 1111111 (00 hex to 7F hex).
Examples
SEARCH:SEARCH1:TRIGGER:A:BUS:B1:USB:ADDRESS:HIVALUE
“0001000”
sets the upper limit to binary 0001000 (08 hex).
SEARCH:SEARCH1:TRIGGER:A:BUS:B1:USB:ADDRESS:HIVALUE? might
return
:SEARCH:SEARCH1:TRIGGER:A:BUS:B1:USB:ADDRESS:HIVALUE
“1111111”
, which indicates that the upper limit is 1111111 (7F hex).
SEARCH:SEARCH<x>:TRIGger:A:BUS:B<x>:USB:ADDRess:VALue
This command specifies the address value for the lower limit for inside-of-range
and outside-of-range qualifiers to be used in a USB search.
SEARCH<x> is the
search number, which is always 1, and
B<x>
Conditions
This command requires the DPO4USB application module to be installed, except
for MDO3000 models, which require the MDO3USB. Note that 1 GHz bandwidth
models are recommended for high-speed (HS) USB.
Group
Search
Syntax
SEARCH:SEARCH<x>:TRIGger:A:BUS:B<x>:USB:ADDRess:VALue
<QString>
SEARCH:SEARCH<x>:TRIGger:A:BUS:B<x>:USB:ADDRess:VALue?
Related Commands
SEARCH:SEARCH<x>:TRIGger:A:BUS:B<x>:USB:CONDition
MDO4000/B/C, MSO/DPO4000B and MDO3000 Series Oscilloscopes Programmer Manual 2-701