MDO3000 Programmer Manual.pdf - 第205页
Commands Listed in Alphabetical Order SEARCH:SEARCH <x>:TRIGger:A:BUS:B<x>:ETHERnet:CONDition Arguments IPv4 sets the E thernet protocol type to Internet Protocol version 4. OTHER sets the Ethernet protoco l …

Commands Listed in Alphabetical Order
Group
Bus
Syntax
BUS:B<x>:ETHERnet:PRObe {DIFFerential|SINGleended}
BUS:B<x>:ETHERnet:PRObe?
Related Commands
BUS:B<x>:ETHERnet:SOUrce:DIFFerential
BUS:B<x>:ETHERnet:SOUrce:DMINus
BUS:B<x>:ETHERnet:SOUrce:DPLUs
Arguments
DIFFerential
SINGleended
Examples
BUS:B1:ETHERnet:PRObe DIFFerential specifies that
a differential probe is connected to the input channels
specified by
BUS:B<x>:ETHERnet:SOUrce:DPLUs and
BUS:B<x>:ETHERnet:SOUrce:DMINus.
BUS:B1:ETHERnet:PRObe?
might return SINGL EENDED, indicating that the
probe type has been set to single-ended.
BUS:B<x>:ETHERnet:PROTOcol
Use this command to set the Ethernet protocol type to TCP/IPv4, or to OTHER.
The default is IPV4.
NOTE. The commands TRIGger:A:BUS:B<x>:ETHERnet:CONDition DATa and
SEARCH:SEARCH<x
>:TRIGger:A:BUS:B<x>:ETHERnet:CONDition DATa
are impacted by the setting of this command.
Conditions
This command req
uires a DPO4ENET application module for DPO/MSO4000B
and MDO4000/B/C models.
Group
Bus
Syntax
BUS:B<x>:ETHERnet:PROTOcol {IPv4|OTHER}
BUS:B<x>:ETHERnet:PROTOcol?
Related Commands
TRIGger:A:BUS:B<x>:ETHERnet:CONDition
2-174 MDO4000/B/C, MSO/DPO4000B and MDO3000 Series Oscilloscopes Programmer Manual

Commands Listed in Alphabetical Order
SEARCH:SEARCH
<x>:TRIGger:A:BUS:B<x>:ETHERnet:CONDition
Arguments
IPv4 sets the E
thernet protocol type to Internet Protocol version 4.
OTHER sets the Ethernet protocol type to other than IPv4.
Examples
BUS:B1:ETHERnet:PROTOcol IPv4 specifies IPv4 as the Ethernet protocol.
BUS:B1:ETHERnet:PROTOcol? might return OTHER.
BUS:B<x>:ETHERnet:SOUrce:DIFFerential
This command specifies the Ethernet data source for differential input. The
supported source waveforms are channels 1–4, math waveform, and reference
waveforms 1–4. The default is channel 1.
Conditions
This command requires a DPO4ENET application module for DPO/MSO4000B
and MDO4000/B/C models.
Group
Bus
Syntax
BUS:B<x>:ETHERnet:SOUrce:DIFFerential {CH1|CH2|CH3|CH4|MATH
|REF1|REF2|REF3|REF4}
BUS:B<x>:ETHERnet:SOUrce:DIFFerential?
Related Commands
BUS:B<x>:ETHERnet:PRObe
TRIGger:A:BUS:B<x>:ETHERnet:CONDition
SEARCH:SEARCH<x>:TRIGger:A:BUS:B<x>:ETHERnet:CONDition
Arguments
CH1–4 specifies to use one of the channels 1–4 as the Ethernet data source for
differential input
.
MATH specifies to use the math waveform as the source for Ethernet data
differential inpu
t
REF1–4 specifies to use one of the reference waveforms 1–4 as the source for
Ethernet data dif
ferential input.
Examples
BUS:B1:ETHERnet:SOUrce:DIFFerential ref4 specifies to use reference
waveform 4 as the source for the Ethernet data differential input.
MDO4000/B/C, MSO/DPO4000B and MDO3000 Series Oscilloscopes Programmer Manual 2-175

Commands Listed in Alphabetical Order
BUS:B1:ETHERn
et:SOUrce:DIFFerential?
might return CH2, indicating
that channel 2 is the currently specified source for Ethernet data differential input.
BUS:B<x>:ETHERnet:SOUrce:DMINus
This command specifies the Ethernet data source for D- input for single-ended
probing. The default is Channel 2.
Conditions
This command requires a DPO4ENET application module for DPO/MSO4000B
and MDO4000/B/C models.
Group
Bus
Syntax
BUS:B<x>:ETHERnet:SOUrce:DMINus {CH1|CH2|CH3|CH4|D0|D1|D2|D3
|D4|D5|D6|D7|D8|D9|D10|D11|D12|D13|D14|D15}
BUS:B<x>:ETHERnet:SOUrce:DMINus?
Related Commands
BUS:B<x>:ETH
ERnet:PRObe
BUS:B<x>:ETHERnet:SOUrce:DPLUs
Arguments
CH1–4 specifies to use one of the analog channels as the Ethernet data source
for the D- input.
D0–D15 specifies to use one of the digital channelsD0–D15astheEthernetdata
source for the D- input. (MSO/MDO4000/B/C models only. )
Examples
BUS:B1:ETHERnet:SOUrce:DMINus D5 specifies the Ethernet data source
for the D- input as D5.
BUS:B1:ETHERnet:SOUrce:DMINus? might return CH3, indicating that
channel 3 is the Ethernet data source for the D- input.
BUS:B<x>:ETHERnet:SOUrce:DPLUs
This command specifies the Ethernet data source for the D+ input for single-ended
probing. The default is channel 1.
Conditions
This command requires a DPO4ENET application module for DPO/MSO4000B
and MDO4000/B/C models.
2-176 MDO4000/B/C, MSO/DPO4000B and MDO3000 Series Oscilloscopes Programmer Manual