MDO3000 Programmer Manual.pdf - 第239页

Commands Listed in Alphabetical Order Examples BUS:UPPerthre shold:REF2 TTL sets the upper threshold for reference waveform 2 to 1.4000. BUS:UPPerthre shold:REF3? might return 0.0E+0, which is the default. BUSY? (Query O…

100%1 / 1158
Commands Listed in Alphabetical Order
Group
Bus
Syntax
BUS:UPPerthreshold{:MATH|:MATH1} {<NR3>|ECL|TTL}
BUS:UPPerthreshold{:MATH|:MATH1}?
Related Commands
BUS:LOWerthreshold{:MATH|:MATH1}
Arguments
<NR3> is a floating point number that specifies the upper threshold of the math
waveform, in volts.
ECL ECL (-1.3 volts). Note that this setting is constrained, depending upon the
vertical scale for the specified math waveform.
TTL TTL (1.4 volts). Note that this setting is constrained, depending upon the
vertical scale for the specified math waveform.
Examples
BUS:UPPerhreshold:MATH TTL sets the upper threshold for the math
waveform to 1.4000 volts.
BUS:UPPerhreshold:MATH? might return 0.0E+0, which is the default.
BUS:UPPerthreshold:REF<x>
This command sets the upper threshold for each reference waveform. This applies
to all search and trigger types that use that reference waveform.
Group
Bus
Syntax
BUS:UPPerthreshold:REF<x> {<NR3>|ECL|TTL}
BUS:UPPerthreshold:REF<x>?
Related Commands
BUS:LOWerthreshold:REF<x>
Arguments
<NR3> is a floating point number that specifies the upper threshold for the
reference waveform, in volts.
ECL ECL (-1.3 volts). Note that this setting is constrained, depending upon the
vertical scale for the specified reference waveform.
TTL TTL (1.4 volts). Note that this setting is constrained, depending upon the
vertical scale for the specified reference waveform.
2-208 MDO4000/B/C, MSO/DPO4000B and MDO3000 Series Oscilloscopes Programmer Manual
Commands Listed in Alphabetical Order
Examples
BUS:UPPerthre
shold:REF2 TTL
sets the upper threshold for reference
waveform 2 to 1.4000.
BUS:UPPerthreshold:REF3? might return 0.0E+0, which is the default.
BUSY? (Query Only)
Returns the status of the oscilloscope. This command allows you to synchronize
the operation of the oscilloscope with your application program. (See page3-7,
Synchronization Methods.)
Certain oscilloscope operations can affect the
BUSY? response. (See Table 3-3
on page 3-8.)
Group
Status and Error
Syntax
BUSY?
Related Commands
*OPT?, *WAI
Returns
<NR1> = 0 means the oscilloscope is not busy processing a command whose
execution time is extensive.
<NR1> = 1 means the oscilloscope is busy processing a command whose execution
time is extensive. (See Table 3-3 on page 3-8.)
Examples
BUSY? might return :BUSY 1 indicating that the oscilloscope is currently busy.
*CAL? (Query Only)
Performs an internal self-calibration and returns the oscilloscope calibration status.
NOTE. Disconnect or otherwise remove all input signals prior to starting
self-calibration. The self-calibration can take several minutes to complete .
Group
Calibration and Diagnostic
Syntax
*CAL?
Returns
<NR1> = 1 indicates the calibration did not complete successfully.
MDO4000/B/C, MSO/DPO4000B and MDO3000 Series Oscilloscopes Programmer Manual 2-209
Commands Listed in Alphabetical Order
<NR1> = 0 indica
tes the calibration completed without errors.
Examples
*CAL? starts the internal signal path calibration and might return 0 to indicate that
the calibration was successful.
CALibrate:FACtory:STATus? (Query Only)
Returns the factory calibration status value saved in nonvolatile memory.
Group
Calibra
tion and Diagnostic
Syntax
CALibrate:FACtory:STATus?
Examp
les
CAL:FAC:STAT? might return CALIBRATE:FACTORY:STATUS PASS indicating
that factory calibration passed.
CALibrate:FACtory:STATus:AFG? (Query Only)
Thi
s query returns the factory calibration status for the Arbitrary Function
Generator portion of the instrument, if present.
Th
is query is synonymous with CALibrate:RESults:FACtory:AFG?
Group
Calibration and Diagnostic
Syntax
CALibrate:FACtory:STATus:AFG?
Related Commands
CALibrate:FACtory:STATus:SCOPE?, CALibrate:FACtory:STATus:RF?
Returns
PASS - factory calibration has passed
INIT - factory calibration is unadjusted
RUNNING - factory calibration is running
Examples
CALIBRATE:FACTORY:STATUS:AFG? might return RUNNING, indicating that the
factory calibration for the AFG portion of the instrument is in progress.
2-210 MDO4000/B/C, MSO/DPO4000B and MDO3000 Series Oscilloscopes Programmer Manual