4OM-1733 1-4-7 4.2 V acuum System V acuum Pressure Every Day Inspection Required T ime: 1 minute Fig. 4A4-8 1208-001 4.2 V acuum System

100%1 / 153
4OM-1733
1-4-6
(2) While rotating the lter holder counterclockwise, pull it out.
Filter Holder
Fig. 4A4-7
(3) Pull out the lter and blow air to it for cleaning.
(4) Insert the lter back in place and turn the lter holder clockwise to attach.
(5) Attach the lter cover.
Note
When the lter has become defective, replace it with a new one.
1208-001
4.1 Air Source
4OM-1733
1-4-7
4.2 Vacuum System
Vacuum Pressure
Every Day Inspection
Required Time: 1 minute
Fig. 4A4-8
1208-001
4.2 Vacuum System
4OM-1733
1-4-8
4.2.1 Vacuum Pressure Check
Cycle : Every Day (Required Time: 1 minute)
Vacuum Pressure
Fig. 4A4-9
Procedure
Check the vacuum pressure visually.
Vacuum Valve Fully Closed : Min. -93 kPa (Min. 70 cmHg) or more
Running : Min. -60 kPa (Min. 45 cmHg) or more
Note
When the vacuum pressure is insufcient, air may be leaking from the vacuum
line or the vacuum pump may have broken down.
1211-002
4.2 Vacuum System