4OM-1733-005w_F8.pdf - 第80页
4OM-1733 1-4-7 4.2 V acuum System V acuum Pressure Every Day Inspection Required T ime: 1 minute Fig. 4A4-8 1208-001 4.2 V acuum System

4OM-1733
1-4-6
(2) While rotating the lter holder counterclockwise, pull it out.
Filter Holder
Fig. 4A4-7
(3) Pull out the lter and blow air to it for cleaning.
(4) Insert the lter back in place and turn the lter holder clockwise to attach.
(5) Attach the lter cover.
Note
When the lter has become defective, replace it with a new one.
1208-001
4.1 Air Source

4OM-1733
1-4-7
4.2 Vacuum System
Vacuum Pressure
Every Day Inspection
Required Time: 1 minute
Fig. 4A4-8
1208-001
4.2 Vacuum System

4OM-1733
1-4-8
4.2.1 Vacuum Pressure Check
Cycle : Every Day (Required Time: 1 minute)
Vacuum Pressure
Fig. 4A4-9
Procedure
Check the vacuum pressure visually.
Vacuum Valve Fully Closed : Min. -93 kPa (Min. 70 cmHg) or more
Running : Min. -60 kPa (Min. 45 cmHg) or more
Note
When the vacuum pressure is insufcient, air may be leaking from the vacuum
line or the vacuum pump may have broken down.
1211-002
4.2 Vacuum System