4 Machine - Calibrations
152 Service Manual Process Lens PL - 03/2025
Reconnect the Low Ring light connector.
Open the machine cover.
NOTICE
The mapping information is stored in C:\system\Log\TableMapping_timestamp*.log file (approx.
459KB).
The Delta of Gantry to fiducials` positions should be ≤ 20µm (0.020mm).
4.4.2 Table Mapping: Verify
For verification of positional accuracy of the camera across the work area.
Verify saves time. It takes only 5.5 minutes compared to 63 minutes in Teach.
The procedure is similar to the Teach procedure.
Verify the table mapping again if:
The machine is already calibrated but is moved.
The positional accuracy of the camera across the work area is suspect.
Maintenance is performed on the gantry axes e.g. greasing.
Step 1
Fig.232: “System Test” tab
Go to the DiagnosticPage (1).
Click System Test (2).
Go to Utility, Table Mapping Calibration
(3).
Under Option, select Verify (4).
Click Start (5).
Fig.233: Message
A message asking for the mapping data
file pops up. (1)
Click Ok (2).
4 Machine - Calibrations
Service Manual Process Lens PL - 03/2025 153
Step 2
Fig.234: Choosing the file
Choose the mp_mess.dat file (1).
Click Open (2), data loading.
Fig.235: Confirming message
A message asking to remove any support
pins pops up. Click Open door(1).
Click Confirm (2).
Wait while the conveyor adjust to a width
of 560mm. (3)
Step 3
Fig.236: Installing the Assembly
A message pops up after the conveyor
width reaches 560mm (1). Open the
machine cover.
Install the Coaxial Light Assembly and
position the Mapping Plate on the con-
veyor.
Click Ok.
Step 4
Fig.237: Setting values
Set 2D M-Ring_G (1) about 10 and all
other lighting values to 0.
The plate´s bottom right hand cross fidu-
cial comes into view. Click on the cross (2)
to centre the cross to the FOV.
4 Machine - Calibrations
154 Service Manual Process Lens PL - 03/2025
Step 5
Fig.238: “Focus” icon
Click on Focus icon (1).
Adjust Focus (2) between 50 ~ 60.
Click Continue (3) to start the teach table
mapping process.
The progress of the mapping is continu-
ously reported.
NOTICE
At 50% completion, the table towers, the stopper rises, and the plate will move forward to the stopper
to continue the second stage of the mapping. Do not be alarmed by this sudden movement. No user
intervention is required.
Step 6
Fig.239: Message”
A message pops up asking for the removal
of the coaxial light and mapping plate.
Click Ok (1) once completed.
Fig.240: Message
A successful verification is completed.
Click Ok (1).
NOTICE
If the mapping fails, clean the plate and repeat the procedure. Mapping failures are caused by a fail-
ure to read a fiducial on the plate – usually caused by fingerprint smudges on the fiducials. Verifica-
tion of the mapping results is necessary after a successful Teach.