Communication Interfaces 96 VTK - TR -BA- 072e - 1 - User Manua l MDS 358X S eries 29 SYSTEM:KLOCK:OFF (KLOCK = Key Lock) Descript ion: Access to keypad is p ermitted, the lock ing function disabled. Example: Inpu t: SYS…

100%1 / 195
Communication Interfaces
VTK-TR-BA-072e-1 - User Manual MDS 358X Series
95
26 MAINT:STATUS MAINT = Maintenance
Description:
This command provides the number of pulses in percent of the preset
limits. It is useful, if you want to estimate the date of the next exchange or
maintenance.
The answer will be send in one line.
Example:
Input:
MAINT:STATUS
Result:
Current percentage of the preset number of cycles (limit).
Return:
Maintenance: 10 %,Nozzle: 20 %,Tappet: 30 %,Maint.
Message: ON
27 MAINT:MESSAGE:OFF (MAINT = Maintenance)
Description:
This command deactivates the maintenance message.
The message “Maint.” will no longer be shown in line 2 of the display. Also
the red maintenance LED will not be switched on, if the maintenance,
nozzle or tappet limits are reached.
By default, the maintenance message is activated.
Example:
Input:
MAINT:MESSAGE:OFF
Result:
The maintenance message is deactivated.
Return:
OK
28 MAINT:MESSAGE:ON (MAINT = Maintenance)
Description:
This command activates the maintenance message.
The message “Maint.” will be shown in line 2 of the display. Also the red
maintenance LED will be switched on, if the maintenance, nozzle or tappet
limits are reached.
By default, the maintenance message is activated.
Example:
Input:
MAINT:MESSAGE:ON
Result:
The maintenance message is activated.
Return:
OK
Communication Interfaces
96
VTK-TR-BA-072e-1 - User Manual MDS 358X Series
29 SYSTEM:KLOCK:OFF (KLOCK = Key Lock)
Description:
Access to keypad is permitted, the locking function disabled.
Example:
Input:
SYSTEM:KLOCK:OFF
Result:
The keypad of the control unit can be used.
Return:
OK
30 SYSTEM:KLOCK:ON (KLOCK = Key Lock)
Description:
This command locks the keypad of the control unit. This way, unauthorized
modification of parameters can be prevented.
Example:
Input:
SYSTEM:KLOCK:ON
Result:
The keypad is locked.
Return:
OK
31 SYSTEM:SHOW:CYCLES
Description:
The current value of the cycle counter is indicated.
Example:
Input:
SYSTEM:SHOW:CYCLES
Result:
Current value of the cycle counter.
Return:
1235000
32 SYSTEM:SHOW:VALVEID
Description:
The valve ID is displayed.
In case the system is in auxiliary mode, the return is Auxiliary Mode”.
Example:
Input:
SYSTEM:SHOW:VALVEID
Result:
ID of the connected valve.
Return:
Valve ID: 10PEA001
33 SYSTEM:SHOW:CONTROLLERID
Description:
The ID of the control unit is displayed.
Example:
Input:
SYSTEM:SHOW:CONTROLLERID
Result:
ID of the control unit.
Return:
Controller ID: 13050
34 SYSTEM:SHOW:STATUS
Description:
This command sends the current status of KeyLock, DosOK with Delay,
SingleDosOK and Auxiliary Mode.
Example:
Input:
SYSTEM:SHOW:STATUS
Result:
Settings of the above listed items
Return:
KeyLock: OFF
DosOK with Delay: OFF
SingleDosOK: per pulse
Auxiliary Mode: OFF
Communication Interfaces
VTK-TR-BA-072e-1 - User Manual MDS 358X Series
97
35 SYSTEM:SHOW:ACTTEMP
Description:
The current temperature of the actuator (piezo) is displayed.
NOTE:
In case an MFC is connected to your MDC, you cannot use this command,
since it would respond with a wrong value.
Example:
Input:
SYSTEM:SHOW:ACTTEMP
Result:
The value appears in °C.
Return:
70
36 SYSTEM:DOSOKDELAY:OFF
Description:
This command deactivates the DOSOK-Delay.
When this is true, the length of a delay is not added to the length of the
DOSOK signal.
Example:
Input:
SYSTEM:DOSOKDELAY:OFF
Result:
Deactivates the DOSOK-delay.
Return:
OK
37 SYSTEM:DOSOKAYDELAY:ON
Description:
This command activates the DOSOK-Delay.
When this is true, the length of a delay is added to the length of the DOSOK
signal.
Example:
Input:
SYSTEM:DOSOKDELAY:ON
Result:
Activates the DOSOK-delay.
Return:
OK
38 SYSTEM:SINGLEDOSOK:SETUP
Description:
This command sets the Single-DOSOK signal to “Setup”. The length of the
Single-DOSOK signal is that of the setup. (Default setting is “Pulse”.)
Example:
Input:
SYSTEM:SINGLEDOSOK:SETUP
Result:
The Single-DOSOK is set to “setup”.
Return:
OK
39 SYSTEM:SINGLEDOSOK:PULSE
Description:
This command sets the Single-DOSOK signal to “Pulse”. The length of the
Single-DOSOK signal is that of a pulse. (Default setting is “Pulse”.)
Example:
Input:
SYSTEM:SINGLEDOSOK:PULSE
Result:
The Single-DOSOK is set to “pulse”.
Return:
OK