Error Me ssag es VTK - TR -BA- 072e - 1 - User Manua l MDS 358X S eries 153 Error code display: 905 Se tup Load Erro r pr. Enter Error code status menu: 905 Setup L oad Error Error handl ing: • Pre ss [enter] key to ackn…

Error Messages
152
VTK-TR-BA-072e-1 - User Manual MDS 358X Series
Error handling:
• Press [enter]-key to acknowledge error message.
• Reenter the working configuration.
Valve:
Valve remains unchanged. If the error message appears
during the start of the MDC, the valve is open, otherwise it
is closed.
902 902 EEPROM not formatted
A corrupted memory area becomes obvious when reading the EEPROM.
Error code display:
902 EEPROM not formatted Enter
Error code status menu:
902 EEPROM not Formatted
Error handling:
• Press [enter]-key to acknowledge error message – the
factory settings will be loaded
• You have to reenter your dispensing parameters
• In case the error appears again, send the MDC to
VERMES Microdispensing or to your supplier
Valve:
Valve remains unchanged. If the error message appears
during the start of the MDC, the valve is open, otherwise it
is closed.
903 903 EEPROM Write Error
The problem can occur when writing data to the EEPROM.
Error code display:
903 EEPROM Write Error pr. Enter
Error code status menu:
903 EEPROM Write Error
Error handling:
• Press [enter] key to acknowledge error message.
• If the problem arises when modifying program
parameters, the new values are not saved in the
EEPROM. After restart of the MDC, the former values of
the EEPROM will be loaded.
• In case the error appears again, send the MDC to
VERMES Microdispensing or to your supplier
Valve:
Valve remains unchanged. If the error message appears
during the start of the MDC, the valve is open, otherwise it
is closed.
904 904 Setup Save Error
The storing procedure for a setup, initiated by [save] key, fails.
Error code display:
904 Setup Save Error pr. Enter
Error code status menu:
904 Setup Save Error
Error handling:
• Press [enter] key to acknowledge error message. After
restart of the MDC, the factory settings will not be
loaded.
• You have to reenter your dispensing parameters.
• In case the error appears again, send the MDC to
VERMES Microdispensing or to your supplier.
Valve:
Valve is closed
905 905 Setup Load Error
This problem may arise during the retrieval of a setup by [recall] key.

Error Messages
VTK-TR-BA-072e-1 - User Manual MDS 358X Series
153
Error code display:
905 Setup Load Error pr. Enter
Error code status menu:
905 Setup Load Error
Error handling:
• Press [enter] key to acknowledge error message
• Repeat the procedure. If the problem persists, you have
to enter new data to the corresponding memory
position.
• In case the error appears again, send the MDC to
VERMES Microdispensing or to your supplier
Valve:
Valve remains unchanged. If the error message appears
during the start of the MDC, the valve is open, otherwise it
is closed.
999 999 Error in Errorlist
The problem cannot be assigned to any other error code from the list. This error only appears
while browsing through the error list in the status menu.
Error code display:
999 Error in Errorlist
Error code status menu:
999 Error in Errorlist
Error handling:
• Move on in the error list or leave the submenu “Error”.
Valve:
Valve is closed

Error Messages
154
VTK-TR-BA-072e-1 - User Manual MDS 358X Series
11.3 Status Messages
The following table lists all the status messages you could encounter when using this system (see
Tab. 32). Status messages can appear while you plug IN or OUT a heater or a cooling valve, as well
as in connection with the calibration of a heater (see paragraph 7.13.2, page 77).
INFORMATION
Status messages are not stored
Status messages disappear after approx. 2 s from the display. They are not saved. A special case is
a calibration gone wrong. Then you get three times in a row the alternating messages “Wrong
calibration” and “Please try again!”, separated by approx. 1.5 s.
Status message Concerns Notes
Cooler connected
Cooling
Cooling valve connected successfully
Cooler could not be activated
Cooling
Cooling valve could not be activated (e.g. because the
flow control valve is not connected correctly or because
the wrong device is connected)
Cooler is disconnected
Cooling
Cooling valve was disconnected during operation
Heater connected
Heater
Heater connected successfully
Heater could not be activated
Heater
Heater could not be activated (e.g. because the heater
is not connected correctly or because the wrong device
is connected)
Heater is disconnected
Heater
Heater was disconnected during operation
Please try again!
Calibration
Calibration was not finished successfully (e.g. because
calibrator was connected twice). Please try again.
Wrong calibration
Calibration
Calibration of the heater (or of heater
and
cooling) not
correct