VTK-TR-BA-072e-1 - User Manual MDS 358X Series
14.6 Overview of the Command Functions
Details of the commands listed below are explained in paragraph 8.1.2.2, page 88.
RS-232C commands Functions
1
*ESR? (e.g. 50 errors) Check the latest error codes
2
*ESR2? (e.g. 50 errors) Check the latest error codes with parameters
3
*IDN? Check device information
4
*OPC? Check the last trigger impulses
5
ADJUST:? Check the adjust status
6
ADJUST:START Initiate the adjust
7
ADJUST:OFFSET:? Check current adjust offset value
8
ADJUST:OFFSET:<offset value> Set adjust offset value
9
ADJUST:CALIBRATION:? Check the status of (remote) adjust
10
ADJUST: CALIBRATION:ON Activate the adjust
11
HEATER:? Check the heater status
12
HEATER:ID? Check the information of the connected heater
13
HEATER:OFF Turn off the connected heater
14
HEATER:ON Turn on the connected heater
15
HEATER:TEMP:<target temperature> Set the target temperature for the connected heater
16
HEATER:LIMITS:<min>,<max> Set the limits for the temperature regulation
17
COOLER:? Check the cooling status
18
COOLER:ID? Check the information of the connected flow control valve
19
COOLER:OFF Turn off the connected flow control valve
20
COOLER:ON Turn on the connected flow control valve
21
COOLER:TEMP:<target temperature> Set the target temperature for the cooling
22
COOLER:OFFSET:<offset> Set the cooler offset for the connected flow control valve
23
KEY:ENTER Send ENTER signal to confirm error(s)
24
KEY:ESCAPE Send ESCAPE signal
25
HELP Check all RS-232C commands
26
LCD? Check current content of the screen
27
MAINT:STATUS Check accumulated amount of pulses (in %)
28
MAINT:MESSAGE:OFF Deactivate the maintenance message
29
MAINT:MESSAGE:ON Activate the maintenance message
30
SYSTEM:KLOCK:OFF Unlock the keypad
31
SYSTEM:KLOCK:ON Lock the keypad
32
SYSTEM:SHOW:CYCLES Check the current cycle counter
33
SYSTEM:SHOW:VALVEID Check the valve ID
34
SYSTEM:SHOW:CONTROLLERID Check the MDC ID
35
SYSTEM:SHOW:STATUS
Check the status of KeyLock, DosOK with Delay, SingleDosOK
and Auxiliary Mode
36
SYSTEM:SHOW:ACTTEMP Check the actuator temperature
37
SYSTEM:DOSOKDELAY:OFF Deactivate the DOSOK-Delay

Attachments
VTK-TR-BA-072e-1 - User Manual MDS 358X Series
171
RS-232C commands Functions
38
SYSTEM:DOSOKDELAY:ON Activate the DOSOK-Delay
39
SYSTEM:SINGLEDOSOK:SETUP Set the Single-DOSOK signal to “Setup”
40
SYSTEM:SINGLEDOSOK:PULSE Set the Single-DOSOK signal to “Pulse”
41
SYSTEM:PASSWORD:<your password> Send the password for unlocking the keypad
42
SYSTEM:PASSWORD:OFF Deactivate the password for unlocking the keypad
43
SYSTEM:PASSWORD:ON Activate the password for unlocking the keypad
44
SYSTEM:PASSWORD:SET:<your password> Set the password for unlocking the keypad
45
SYSTEM:AUXILIARYMODE:OFF Deactivate the auxiliary mode
46
SYSTEM:AUXILIARYMODE:ON Activate the auxiliary mode
47
TRIGGER:ASET:? Check the pulse parameters in the RAM
48
TRIGGER:ASET:<RI>,<OT>,<FA>,<NL>,<NP>,<D
L>
Change the pulse parameters in the RAM
(MDC returns “OK”)
49
TRIGGER:ASET:<RI>,<OT>,<FA>,<NL>,<NP>,<D
L>,1
Change the pulse parameters in the EEPROM
(MDC returns “OK”)
50
STRIGGER:ASET:<RI>,<OT>,<FA>,<NL>,<NP>,<
DL>
Change the pulse parameters in the RAM
(MDC returns the saved parameters)
51
STRIGGER:ASET:<RI>,<OT>,<FA>,<NL>,<NP>,<
DL>,1
Change the pulse parameters in the EEPROM
(MDC returns the saved parameters)
52
VALVE:UP Open the valve
53
VALVE:DOWN Close the valve
54
VALVE:AOPEN Initiate a dispensing cycle with current parameters
(MDC returns “OK”)
55
VALVE:AOPEN:<RI>,<OT>,<FA>,<NL>,<NP>,<D
L>
Initiate a dispensing cycle with given parameters
(MDC returns “OK”)
56
VALVE:AOPENS<setup no.> Initiate a dispensing cycle with selected setup
( MDC returns “OK”)
57
SVALVE:AOPEN Initiate a dispensing cycle with current parameters
(MDC returns the saved parameters)
58
SVALVE:AOPEN:<RI>,<OT>,<FA>,<NL>,<NP>,<
DL>
Initiate a dispensing cycle with given parameters
(MDC returns the saved parameters)
59
SVALVE:AOPENS<setup no.> Initiate a dispensing cycle with selected setup
(MDC returns the saved parameters)
60
WRITE:LCD:<text> Write an ASCII text on the display
61
TAPPET:SET:<value> Set tappet maintenance counter
62
TAPPET:CLEAR Clear tappet maintenance counter
63
NOZZLE:SET:<value> Set nozzle maintenance counter
64
NOZZLE:CLEAR Clear nozzle maintenance counter
65
SCENARIO:STATUS Check the status of scenarios and PLCstops
66
SCENARIO:OFF Deactivate the use of scenarios
67
SCENARIO:ON Activate the use of scenarios
68
SCENARIO:PLCSTOP:1:OFF Deactivate the PLC-Stop
69
SCENARIO:PLCSTOP:1:ON Activate the PLC-Stop
70
SCENARIO:SAVE:<scenario no.>:<values> Save the parameters for selected scenario