Control Unit MDC
VTK-TR-BA-072e-1 - User Manual MDS 358X Series
35
4.5.5 Submenu “Scenario”
In the submenu “Scenario” you decide if you want to work with scenarios (see paragraph 7.9.1
"Basics about Scenarios", page 70). You can also enter parameters for pre-defined scenarios. Each
of the four potential scenarios has its own submenu, where you can set those parameters. The
second diagram explains the structure of these submenus. Check there to find the possible limits
for those parameters.
With [→] and [←] you can switch between the four scenarios.
Fig. 15: Submenu “Scenario”
To enter the parameters for a scenario, press [enter] on the respective submenu “Scenario-Def.”
The structure of these submenus is explained in the picture below (see Fig. 16, page 36).
In “Sc:1.1 Setup” you can enter “0”, “1”, “2” or “3” to select a setup or “-”, to decide that no further
segment will follow. “0” is the working configuration, “1” to “3” are the parameters of the
respective setups 1 to 3. Then the scenario will end with this segment and the menu will stop
here as well. (If you later change the “-” to the number of a setup, the following segment will be
unlocked again in the menu.) If you set the blank directly in the first segment, the system will
simply use the working configuration instead of the scenario.
After entering a setup and [→], you can select the Number of Pulses next. This NP replaces the
one from the setup, which would be otherwise used. Another [→] gives the option to enter a
scenario delay. Standard value is 10 ms.
Another [→] brings you to the second segment in your scenario (“Sc:1.2 Setup”), where you have
the same options as with the first. All in all you can combine up to ten segments (though only
four different setups, since only setups 0 to 3 are available), each with its own NP and scenario
delay, unless you stop your scenario beforehand by selecting “-” instead of a setup. After the last
segment, another [→] brings you the option to set the PLCStop “ON” or “OFF”. With “ON” a
scenario always ends automatically, once the trigger signal drops to “low”. No further segments
of the scenario will be performed. With the PLCStop “OFF”, the trigger signal only matters, if the
NP in a setup is set as “infinite”. With these parameters will be dispensed until the trigger signal
drops to “low”. If the trigger signal is already “low” when reaching the signal, only one pulse will
be shot. Then the next segment of the scenario will follow after the scenario delay (unless, of
course, it was already the last segment of the scenario).

Control Unit MDC
36
VTK-TR-BA-072e-1 - User Manual MDS 358X Series
Fig. 16: Submenu “Scenario-Def. 1”
INFORMATION
Finding PLCStop quickly
If you just want to change the PLCStop setting, but not any of the segments in your scenario, it is
quickest to use [←] once after entering the scenario, since this submenu is wrap-around, just like
all other.

Control Unit MDC
VTK-TR-BA-072e-1 - User Manual MDS 358X Series
37
4.5.6 Submenu “Service-Option”
This menu has two sections. In “Service Code” you can enter a service code (see Tab. 17).
“Baud Rate” is used to change the baud rate. There are five possible values: 9600, 19200, 38400,
57600 and 115200 (new MDCs are set to 115200).
Service Code Explanation
33
If you enter 33, you can activate or change the adjust offset (see
paragraph 7.12 "Adjust Offset", page 75).
35
If you enter 35, the adjust offset is set back to “0”.
1000
If you enter 1000, you enter a further submenu containing the options listed
in the table below (see Tab. 18).
Tab. 17: List of service codes
Option Explanation
SingleDosOK
You can select between the possibilities that the SingleDosOK signal is
switched per pulse or per setup. (Default setting is per pulse. For the pin
configuration of the PLC interface, see Page 113.)
DosOK with Delay
You can select, if before switching the signal DosOK there should be one
delay executed at the end. (Default setting is OFF.)
Auxiliary Mode
Here the auxiliary mode can be switched ON or OFF (see paragraph 7.11
"Auxiliary Mode", page 74).
Factory Settings
Here any parameters can be returned to the default factory settings (see
paragraph 7.10 "Factory Settings", page 73). You have got four alternatives:
• Setup 0 – 3 (the parameters of the working configuration and the setups
1 to 3 are returned to the factory settings)
• Reset ALL (all parameters are returned, including the settings of the
scenarios; also the heater will be turned OFF and the temperature set to
10°C, cooling will be turned OFF, set to 80 °C and the cooler offset will be
set as 0 %)
• Scenario (the parameters and settings of the scenarios are returned to the
factory settings)
• Setup ALL (the parameters of the working configuration and all setups
are returned, setups 4 – 10 receive values of setup 0)
Tab. 18: Options in the submenu of service code 1000
Fig. 17: Submenu “Service-Option”