Adjustment HLGB-10424-01 Supplied A ir Sensor Setup SHEET 1/2 Supplied Air Sensor Setup This section describes s etting procedure f or the Supplied Air sensor located on the on the interface panel on the back of the unit…

100%1 / 712
Adjustment
HLGB-10423-01
Vacuum Pump Sensor Setup
SHEET
2/2
6 Press the U and V button to display the
G-ON.
7 Press the MODE button one time.
1
kPa
2
8 Press the U and V button to display the
-48.0.
Setting is completed.
1
kPa
2
Adjustment
HLGB-10424-01
Supplied Air Sensor Setup
SHEET
1/2
Supplied Air Sensor Setup
This section describes setting procedure for the Supplied Air sensor located on the on the interface
panel on the back of the unit.
[Procedure]
1 Press the MODE button for longer than 2
seconds.
HYS on the display screen.
1
kPa
2
2 Press the MODE button one time.
OFF on the display screen.
1
kPa
2
3 Press the MODE button one time.
N.o on the display screen.
1
kPa
2
4 Press the MODE button one time.
2.5 on the display screen.
1
kPa
2
5 Press the MODE button one time.
R-ON on the display screen.
1
kPa
2
Supplied air sensor
MODE button
Adjustment
HLGB-10424-01
Supplied Air Sensor Setup
SHEET
2/2
6 Press the U and V button to display the
G-ON.
1
kPa
2
7 Press the MODE button one time.
MPA on the display screen.
1
kPa
2
8 Press the MODE button two time.
Lo-1 on the display screen.
1
kPa
2
9 Press the U and V button to display
the .400.
1
kPa
2
10 Press the MODE button one time.
Hi-1 on the display screen.
1
kPa
2
11 Press the U and V button to display
the .450.
Setting is completed.
1
kPa
2