MIL-STD-202H.pdf - 第198页

MI L - S TD - 202 -2 1 3 4 .3 . Procedure . The te s t s pe c i m en s hal l b e m ount ed a s s p ec i f i ed i n t he c om pone nt s pe c i fication . W henever possible, t he test load s h al l be di s t r i bu t ed u…

100%1 / 305
MIL-STD-202-213
4.1.2.2 Transducer. The fundamental resonant frequency of the accelerometer shall be greater than 30,000 Hz,
when the accelerator is employed as the shock sensor.
4.1.2.3 Transducer calibration. Transducers shall be calibrated in accordance with ASA STD S2.2
. The accuracy
of the calibration method shall be at least ±5 percent over the frequency range of 2 to 5,000 Hz. The amplitude of the
transducer being calibrated shall also be ±5 percent over the frequency range of 4 to 5,000 Hz.
4.1.2.4 Linearity. The signal level of the system shall be chosen so that the acceleration pulse operates over the
linear portion of the system.
4.1.2.5 Transducer mounting. When conformance to 4.1.3 is required, the monitoring transducer shall be rigidly
secured and located as near as possible to an attachment point of the specimen but not on the specimen itself.
4.1.3 Application of shock measuring instrumentation. Shock measuring instrumentation shall be utilized to
determine that the correct input shock pulse is applied to the test specimen. This is particularly important where a
multi-specimen test is made. Generally, the shock pulse should be monitored whenever there is a change in the test
setup, such as a different test fixture, different component (change in physical characteristics), different weight,
different shock pulse (change in pulse shape, intensity, or duration) or different shock machine characteristics. It is
not mandatory that each individual shock be monitored, provided that the repeatability of the shock application as
specified in 4.1.1.1 has been established.
4.2 Shock pulses. Two types of shock pulses, a half-sine shock pulse and a sawtooth shock pulse, are specified.
The pulse shape and tolerances are shown on figures 1 and 2, respectively. For single degree of freedom systems, a
sawtooth shock pulse can be assumed to have a damage potential at least as great as that of the half-sine pulse if
the shock spectrum of the sawtooth pulse is everywhere at least as great as that of the half-sine pulse. This
condition will exist for two such pulses of the same duration if over most of the spectrum the acceleration peak value
of the sawtooth pulse is 1.4 times the acceleration peak value of the half-sine pulse.
4.2.1 Half-sine shock pulse. The half-sine shock pulse shall be as indicated on figure 1. The velocity change of
the pulse shall be within ±10 percent of the velocity change of the desired shock pulse. The velocity change may be
determined either by direct measurement, indirectly, or by integrating (graphically or electrically) the area (faired
acceleration pulse may be used for the graphical representation) under the measured acceleration pulse. For half-
sine acceleration pulses of less than 3 milliseconds duration the following tolerances should apply: The faired
maximum value of the measured pulse shall be within ±20 percent of the specified ideal pulse amplitude, its duration
shall be within ±15 percent of the specified ideal pulse duration, and the velocity change associated with the
measured pulse shall be within ±10 percent of V
i
= 2AD/π. See figure 1. The measured pulse will then be considered
a nominal half-sine pulse with a nominal amplitude and duration equal to respective values of the corresponding ideal
half-sine pulse. The duration of the measured pulse shall be taken as D
m
= D(.1A)/.94; where D(.1A) is the time
between points at .1A for the faired measured acceleration pulse.
4.2.1.1 The ideal half-sine pulse. An ideal half-sine acceleration pulse is given by the solid curve. See figure 1.
The measured acceleration pulse must lie within the boundaries given by the broken lines. In addition, the actual
velocity change of the shock must be within 10 percent of the ideal velocity change. The actual velocity change can
be determined by direct measurements, or from the area under the measured acceleration curve. The ideal velocity
change is equal to V
i
= 2AD/π; where A is the acceleration amplitude and D is the pulse duration of the ideal pulse.
4.2.2 Sawtooth shock pulse. The sawtooth pulse shall be as indicated on figure 2. The velocity change of the
faired measured pulse shall be within ±10 percent of the velocity change of the ideal pulse.
4.2.2.1 The ideal terminal-peak sawtooth. An ideal terminal-peak sawtooth acceleration pulse is given by the solid
line. See figure 2. The measured acceleration pulse must be within the boundaries given by the broken lines. In
addition, the actual velocity change of the shock pulse must be within 10 percent of the ideal value. The actual
velocity change can be determined from direct measurements, or from the area under the measured acceleration
curve. The ideal velocity change is equal to V
i
= PD/2; where P is the peak value of acceleration, and D is the pulse
duration.
5
北测(上海)电子科技有限公
联系方式:xuyj@beice-sh.com 13917165676
MIL-STD-202-213
4.3. Procedure. The test specimen shall be mounted as specified in the component specification. Whenever
possible, the test load shall be distributed uniformly on the test platform in order to minimize the effects of unbalanced
loads.
4.3.1 Basic design test. Three shocks in each direction shall be applied along the three mutually perpendicular
axes of the test specimen (18 shocks). If the test specimen is normally mounted on vibration isolators, the isolators
shall be functional during the test. The specified test pulse (half-sine or sawtooth pulse) shall be in accordance with
figures 1 and 2, respectively, and shall have a duration and peak value in accordance with one of the test conditions
of table 1.
TABLE 1. Test condition values.
Test
condition
Peak value
(gs)
Normal
duration (D) (ms)
Waveform
Velocity
change (V
i
) ft/sec
A
B
C
D
E
F
G
H
I
J
K
50
75
100
500
1,000
1,500
50
75
100
30
30
11
6
6
1
0.5
0.5
11
6
6
11
11
Half-sine
Half-sine
Half-sine
Half-sine 1/ 2/ 3/
Half-sine 1/ 2/ 3/
Half-sine 1/ 2/ 3/
Sawtooth 2/
Sawtooth 2/
Sawtooth 2/
Half-sine
Sawtooth
11.3
9.2
12.3
10.2
10.2
15.4
8.8
7.2
9.7
6.8
5.3
1/ For half-sine shock pulses of less than 3 milliseconds duration, it is not required that the envelope
fall within the tolerances specified on figure 1. The faired amplitude of the measured pulse shall be
within ±20 percent of the ideal amplitude. The measured duration shall be within ±15 percent of the
specified amplitude duration. The velocity change of the faired measured pulse shall be within ±10
percent of the ideal pulse. The duration of the pulse shall be measured at the 0.1A point on the
pulse. The duration of the pulse shall be the duration measured at the 0.1A point divided by .94.
Test conditions D, E, and F are principally applicable to semiconductors.
2/ Test conditions G, H, and I (sawtooth) waveforms are preferred, except for semi- conductors, for
which test conditions D, E, and F (half-sine) are preferred.
3/ For test condition D. Where the weight of multi-specimen and fixtures exceeds 150 pounds, there
is a question as to whether the shock pulse is properly transmitted to all specimens. Due
consideration should be given to the design of the test fixture to assure the proper shock input to
each specimen. This also applies to test conditions E and F except that where the weight of the
multi-specimen and fixtures exceeds 25 pounds.
6
北测(上海)电子科技有限公
联系方式:xuyj@beice-sh.com 13917165676
MIL-STD-202-213
5. DETAILED REQUIREMENTS
5.1 Measurements. Measurements are to be made before and after the required number of shocks unless
otherwise specified, and during the test if specified.
5.2 Summary. The following details are to be specified in the individual specification:
a. Mounting method and accessories (see 4.1.2.5 and 4.3).
b. Test condition letter (see 4.3.1).
c. Measurements before, during, and after the test (see 5.1).
6. NOTES
(This section contains information of a general or explanatory nature that may be helpful, but is not mandatory.)
6.1 Supersession data. The main body and 38 parts of this revision of MIL-STD-202 replace superseded MIL-STD-
202.
6.2 Cancelled Method. Methods 202 (Shock, specimens weighing not more than 4 pounds) and 205 (Medium
Impact) are cancelled, when specified use Method 213.
202 or 205
test condition
213
test condition
A 15g (pk) K 30g (sawtooth)
B 30g (pk) H 75g (sawtooth)
C 50g (pk) I 100g (sawtooth)
Custodians: Preparing activity:
Army - CR DLA CC
Navy - EC
Air Force - 85 (Project 59GP-2015-027)
DLA - CC
Review activities:
Army - AR, AT, AV, CR4, MI, SM, TE
Navy - AS, OS, SH
Air Force - 19, 99
NSA - NS
NOTE: The activities listed above were interested in this document as of the date of this document. Since
organizations and responsibilities can change, you should verify the currency of the information above using the
ASSIST Online database at https://assist.dla.mil/
7
北测(上海)电子科技有限公
联系方式:xuyj@beice-sh.com 13917165676