MIL-STD-202-112
4.7.2 Apparatus.
4.7.2.1 Chamber. A vacuum/pressure chamber capable of the evacuation and subsequent pressure bombing of
devices up to 75 psig and maintaining pressure up to 12.5 hours.
4.7.2.2 Filter. A filtration system capable of removing particles greater than 1 micron in size from the fluorocarbon
fluid.
4.7.2.3 Detector. A fluorocarbon vapor detection system with a nominal sensitivity of 10
-5
atm cm
3
/s. The
sensitivity of the detector shall be better than 14 micrograms of FC84 or D/80. The detector shall have a linear
response to 280 micrograms and a response time of better than 1 second.
4.7.2.4 Instruments. Suitable calibrated instruments to indicate that test, pressure times and temperatures are as
specified. The detection system shall be calibrated using a calibration source of fluorocarbon at least once each shift
to an accuracy of at least ±15 percent.
4.7.3 Precautions. The following precautions shall be observed in conducting the fluorocarbon gross leak test.
a. Fluorocarbons shall be filtered through a filter system capable of removing particles greater than 1
micrometer prior to use. Bulk filtering and storage is permissible. Liquid that has accumulated observable
quantities of particulate matter during use shall be discarded or reclaimed by filtration for re-use.
Precautions shall be taken to prevent contamination.
b. Precautions should be taken to prevent operator injury due to package rupture or violent evolution of bomb
fluid when testing large packages.
4.7.4 Procedure. The devices shall be placed in a vacuum/pressure chamber and the pressure reduced to 5 torr
or lower and maintained for 30 minutes minimum. A sufficient amount of FC-84 or D/80 detector fluid shall be
admitted to the pressure chamber to completely cover the devices. The fluid shall be admitted after the 30 minute
minimum vacuum period but before breaking the vacuum. The devices shall then be pressurized in accordance with
table II. Upon completion of the pressurization period, the pressure shall be released and the devices removed from
the pressure chamber without being removed from the detector fluid for more than 20 seconds. When the devices
are removed from the fluid for testing, they shall be air dried for a minimum of 20 seconds and a maximum of 5
minutes prior to the test cycle. The devices shall then be tested with a fluorocarbon vapor detector in accordance
with the system operating instructions and as follows. Devices shall be admitted to the purge chamber and purged
(heated to remove any remaining surface fluorocarbon and to initiate vaporization of any internal fluorocarbon fluid).
Purge time shall be in accordance with table III. Test time shall commence immediately after purge either in the
same chamber (single chamber system) or an adjoining test chamber (dual chamber system). Test time shall be a
minimum of 3.5 seconds (unless the device is rejected earlier) with the fluorocarbon vapor detector purge and test
chambers at a temperature of 125°C ±5°C or 2.5 seconds minimum with the purge and test chambers at 150°C ±5°C.
During test time, the amount of fluorocarbon outgassing is measured and the total integrated value is compared to
the alarm level.
NOTE: Air dry, purge, and test time limits for each device shall be complied with in all cases, whether the device is
handled individually or in plastic or metal tubes.
4.7.4.1 Failure criteria. A device shall be rejected if the detector instrumentation indicates more than 280
micrograms of fluorocarbon gas.
12
北测(上海)电子科技有限公司
联系方式:xuyj@beice-sh.com 13917165676

MIL-STD-202-112
TABLE II. Test pressure and duration for condition F.
Test pressure Duration
75 psig .5 hr
60 psig 1.0 hr
30 psig 2.5 hr
15 psig 12.5 hr
TABLE III. Purge time for condition F.
Package with internal
free volume (cm
3
)
Purge time 1/
(seconds)
≤ 0.01
3 1/3-5
>0.01 ≤0.10
6-9
>0.10 8 2/3-13
1/ With purge chamber at 150°C, minimum and maximum
purge times should be reduced by 1/3.
NOTE: If necessary, due to device construction or materials, maximum purge time can be determined by cycling a
device with a 0.02 to 0.05 inch diameter hole and measuring the maximum time that can be used without permitting
the device to escape detection during the test cycle.
4.8. Measurements. Upon completion of the test, measurements shall be made as specified.
5. DETAILED REQUIREMENTS
5.1 Summary. The following details are to be specified in the individual specification.
a. Test condition letter (see 4.1.1).
(1) For test condition C.
(a) Procedure number and accept and reject leak rate.
(b) Backfilling pressure when verification of seal is required (see 4.4.3, 4.4.4.1 and .4.4.2).
(c) Number of pressure cycles if applicable (see 4.4.4.2).
(d) For procedure IV.
1. Reduced pressure of the chamber and duration of pressurization (see 4.4.4.4).
2. Whether perforation of specimen is required (see 4.4.4.4).
(e) For procedures III and IV.
1. Applicable gross leak conditions.
(2) For test conditions D and E.
(a) Test mounting fixtures (see 4.5.4 and 4.6.4).
1. Measurements after test (see 4.8).
13
北测(上海)电子科技有限公司
联系方式:xuyj@beice-sh.com 13917165676

MIL-STD-202-112
6. NOTES
(This section contains information of a general or explanatory nature that may be helpful, but is not mandatory.)
6.1 Supersession data. The main body and 38 parts of this revision of MIL-STD-202 replace superseded MIL-STD-
202.
Custodians: Preparing activity:
Army - CR DLA – CC
Navy - EC
Air Force - 85 (Project 59GP-2015-016)
DLA - CC
Review activities:
Army - AR, AT, AV, CR4, MI, SM, TE
Navy - AS, OS, SH
Air Force - 19, 99
NSA - NS
NOTE: The activities listed above were interested in this document as of the date of this document. Since
organizations and responsibilities can change, you should verify the currency of the information above using the
ASSIST Online database at https://assist.dla.mil/