Communication Interfaces VTK - TR -BA- 171e - 1 - User Manua l MDS 3080 -V 93 20 SETADJLED:OFF Descript ion: This command deactivates t he control la mps f or adju st. This affect s the lamps ou tsid e of th e adjust: • …

Communication Interfaces
92
VTK-TR-BA-171e-1 - User Manual MDS 3080-V
15 KEY:ENTER
Description:
The acknowledgement signal is transferred to the control unit. This option
is only usable in case of errors, which produce a message on the display of
the MDC. You can respond with a serial command. Then the function is
identical to pressing the [enter]-key on the keypad.
Example:
Input:
KEY:ENTER
Result:
The ENTER signal is send.
Return:
OK (no other reaction of the MDC)
16 KEY:ESCAPE
Description:
The ESCAPE signal is transferred to the control unit. This option is only
usable in case of errors, which produce a message on the display of the
MDC. You can respond with a serial command. Then the function is
identical to pressing the [esc]-key on the keypad.
Example:
Input:
KEY:ESCAPE
Result:
The ESCAPE signal is send.
Return:
OK (no other reaction of the MDC)
17 KEY:ADJUST
Description:
The adjust signal is transferred to the control unit.
The function is identical to pressing the [adj]-key on the keypad. For the
remote first drop adjust, this command has the same effect as pressing the
[adj]-key of the MDC during the F2 adjust.
Example:
Input:
KEY:ADJUST
Result:
The first drop position is increased by 1 %.
Return:
OK
18 HELP
Description:
Shows a list with all RS-232C commands.
Information!
The listed commands for heaters do not work with this system. Please
ignore them!
Example:
Input:
HELP
Result:
List with all RS-232C commands.
Return:
List with all commands
19 LCD? LCD? = Liquid-Crystal Display Query
Description:
Use this command to externally inspect the current content of the screen.
Example:
Input:
LCD?
Result:
When sending this command the information on the
display will be returned.
E.g. immediately after switching the MDC ON, the content
of the screen could be “3090+ 200 Hz“.
Return:
“3090+ 200 Hz”

Communication Interfaces
VTK-TR-BA-171e-1 - User Manual MDS 3080-V
93
20 SETADJLED:OFF
Description:
This command deactivates the control lamps for adjust. This affects the
lamps outside of the adjust:
• The red adjust-LED is ON, if the MDC detects an error (except for an
adjust error)
• The green adjust-LED stays OFF
During the adjust, the adjust-LEDs work as always.
Alternatively, you can deactivate the control lamps for adjust by means of
the service code 1100 (see paragraph 4.5.6 "Submenu “Service-Option”
",
page 33).
Example:
Input:
SETADJLED:OFF
Result:
The control lamps for adjust are deactivated.
Return:
OK
21 SETADJLED:ON
Description:
This command activates the control lamps for adjust. This affects the lamps
outside of the adjust:
• Red: if the MDC detects an error
• Green: the valve detects that the adjust value falls inside the
adjust range.
During the adjust, the adjust-LEDs work as always.
Alternatively, you can activate the control lamps for adjust by means of the
service code 1100 (see paragraph 4.5.6 "Submenu “Service-Option”
",
page 33).
Example:
Input:
SETADJLED:ON
Result:
The control lamps for adjust are activated.
Return:
OK
22 SYSTEM:KLOCK:OFF (KLOCK = Key Lock)
Description:
Access to keypad is permitted, the locking function disabled.
Example:
Input:
SYSTEM:KLOCK:OFF
Result:
The keypad of the control unit can be used.
Return:
OK
23 SYSTEM:KLOCK:ON (KLOCK = Key Lock)
Description:
This command locks the keypad of the control unit. This way, unauthorized
modification of parameters can be prevented.
Example:
Input:
SYSTEM:KLOCK:ON
Result:
The keypad is locked.
Return:
OK
24 SYSTEM:SHOW:CYCLES
Description:
The current value of the cycle counter is indicated.
Example:
Input:
SYSTEM:SHOW:CYCLES
Result:
Current value of the cycle counter.
Return:
1235000

Communication Interfaces
94
VTK-TR-BA-171e-1 - User Manual MDS 3080-V
25 SYSTEM:SHOW:VALVEID
Description:
The valve ID is displayed.
In case the system is in auxiliary mode, the return is “Auxiliary Mode”.
Example:
Input:
SYSTEM:SHOW:VALVEID
Result:
ID of the connected valve.
Return:
Valve ID: 10PEA001
26 SYSTEM:SHOW:CONTROLLERID
Description:
The ID of the control unit is displayed.
Example:
Input:
SYSTEM:SHOW:CONTROLLERID
Result:
ID of the control unit.
Return:
Controller ID: 12060
27 SYSTEM:SHOW:STATUS
Description:
This command sends the current status of KeyLock, First Drop, Fixed
Adjust, DosOK with Delay, SingleDosOK and Auxiliary Mode.
If the status for First Drop or Fixed Adjust is ON, the system sends some
extra information like TNL or adjust value.
Example:
Input:
SYSTEM:SHOW:STATUS
Result:
Settings of the above listed items
Return:
KeyLock: OFF
FirstDrop: OFF
FixedAdjust: OFF
DosOK with Delay: OFF
SingleDosOK: per pulse
Auxiliary Mode: OFF