Error Messages
VTK-TR-BA-171e-1 - User Manual MDS 3080-V
153
601 601 USART Buffer Overflow
This problem sometimes occurs during data transmission via serial interface. The buffer is full
and the MDC is unable to handle more inflowing data. The message “601 USART Buffer Over-
flow” is returned via serial interface. LEDs are not lit.
Error code display:
No error message
Error code status menu:
601 USART Buffer Overflow
Error message monitor:
601 USART Buffer Overflow
Error handling:
• Interrupt data transmission
• Send data again
700 700 MDC Calibr. wrong
This error message appears, if the calibration values are incorrect at the start-up of the MDC.
The values will be overwritten with the factory settings. There will be no entry in the error list.
Error code display:
700 MDC Calibr. wrong pr. Enter
Error code status menu:
700 MDC Calibr. wrong
Error handling:
• Press [enter] key to acknowledge error message.
• Send the MDC back to VERMES Microdispensing or to
your supplier.
702 702 Watchdog TimeOut
Abnormal end (crash) of the MDC.
Error code display:
702 Watchdog TimeOut pr. Enter
Error code status menu:
702 Watchdog TimeOut
Error handling:
• Press [enter] key to acknowledge error message. The
MDC restarts

Error Messages
154
VTK-TR-BA-171e-1 - User Manual MDS 3080-V
800 800 Heat. Calib. wrong
The error message “800 Heat. Calib. wrong pr. Enter” appears, if the calibration values of the
heater are incorrect at the start-up of the MDC. The values will be overwritten with the factory
settings. There will be no entry in the error list. The heater should be recalibrated or contact
our technical support.
Error code display:
800 Heat. Calib. wrong pr. Enter
Error code status menu:
No entry
Error handling:
• Press [enter] key to acknowledge error message.
• For further information, contact the Technical Support
(see Page 7).
801 801 No Heater!
Even though the heater option has been activated in the Heater Menu, this element is not
detected.
Error code display:
801 No Heater! Press Enter
Error code status menu:
No entry
Error handling:
• Press [enter] key to acknowledge error message. Heater
status will be switched “OFF” in the submenu “Heater”.
• Inspect the heater cable connection. Worn, kinked or
charred cables must be exchanged.
• Heater defect – change the heater
901 901 RAM Data Error
During the check of RAM data, a problem arises. The message has to be acknowledged by the
[enter] key. The system reformats the EEPROM, replacing current values with the factory
settings.
In case the error appears during start-up of the MDC, all setups will be reset to the factory
settings after pressing [enter]. In case the error appears after the RECALL of a specific setup,
the will be no automatic correction. Therefore, enter the parameters of the working
configuration for this setup and save them by pressing [save].
Error code display:
901 RAM Data Error pr. Enter
Error code status menu:
901 RAM Data Error
Error handling:
• Press [enter] key to acknowledge error message.
• Reenter the working configuration.
902 902 EEPROM not formatted
A corrupted memory area becomes obvious when reading the EEPROM.
Error code display:
902 EEPROM not formatted Enter
Error code status menu:
902 EEPROM not Formatted
Error handling:
• Press [enter] key to acknowledge error message – the
factory settings will be loaded
• You have to reenter your dispensing parameters
• In case the error appears again, send the MDC to
VERMES Microdispensing or to your supplier

Error Messages
VTK-TR-BA-171e-1 - User Manual MDS 3080-V
155
903 903 EEPROM Write Error
The problem can occur when writing data to the EEPROM.
Error code display:
903 EEPROM Write Error pr. Enter
Error code status menu:
903 EEPROM Write Error
Error handling:
• Press [enter] key to acknowledge error message.
• If the problem arises when modifying program
parameters, the new values are not saved in the
EEPROM. After restart of the MDC, the former values of
the EEPROM will be loaded.
• In case the error appears again, send the MDC to
VERMES Microdispensing or to your supplier
904 904 Setup Save Error
The storing procedure for a setup, initiated by [save] key, fails.
Error code display:
904 Setup Save Error pr. Enter
Error code status menu:
904 Setup Save Error
Error handling:
• Press [enter] key to acknowledge error message. After
restart of the MDC, the factory settings will not be
loaded.
• You have to reenter your dispensing parameters.
• In case the error appears again, send the MDC to
VERMES Microdispensing or to your supplier.
905 905 Setup Load Error
This problem may arise during the retrieval of a setup by [recall] key.
Error code display:
905 Setup Load Error pr. Enter
Error code status menu:
905 Setup Load Error
Error handling:
• Press [enter] key to acknowledge error message
• Repeat the procedure. If the problem persists, you have
to enter new data to the corresponding memory
position.
• In case the error appears again, send the MDC to
VERMES Microdispensing or to your supplier
999 999 Error in Errorlist
The problem cannot be assigned to any other error code from the list. This error only appears
while browsing through the error list in the status menu.
Error code display:
999 Error in Errorlist
Error code status menu:
999 Error in Errorlist
Error handling:
• Move on in the error list or leave the submenu “Error”.