a3369b.pdf - 第165页
Transport, Storage and Disposal 156 VTK - TR -BA- 171e - 1 - User Man u al MDS 3080 -V 12 Transport, St orag e and Disp osal 12.1 Transport For sh ipme nt, the s ystem is p acked in an ap propr iate case. Preserve it for…

Error Messages
VTK-TR-BA-171e-1 - User Manual MDS 3080-V
155
903 903 EEPROM Write Error
The problem can occur when writing data to the EEPROM.
Error code display:
903 EEPROM Write Error pr. Enter
Error code status menu:
903 EEPROM Write Error
Error handling:
• Press [enter] key to acknowledge error message.
• If the problem arises when modifying program
parameters, the new values are not saved in the
EEPROM. After restart of the MDC, the former values of
the EEPROM will be loaded.
• In case the error appears again, send the MDC to
VERMES Microdispensing or to your supplier
904 904 Setup Save Error
The storing procedure for a setup, initiated by [save] key, fails.
Error code display:
904 Setup Save Error pr. Enter
Error code status menu:
904 Setup Save Error
Error handling:
• Press [enter] key to acknowledge error message. After
restart of the MDC, the factory settings will not be
loaded.
• You have to reenter your dispensing parameters.
• In case the error appears again, send the MDC to
VERMES Microdispensing or to your supplier.
905 905 Setup Load Error
This problem may arise during the retrieval of a setup by [recall] key.
Error code display:
905 Setup Load Error pr. Enter
Error code status menu:
905 Setup Load Error
Error handling:
• Press [enter] key to acknowledge error message
• Repeat the procedure. If the problem persists, you have
to enter new data to the corresponding memory
position.
• In case the error appears again, send the MDC to
VERMES Microdispensing or to your supplier
999 999 Error in Errorlist
The problem cannot be assigned to any other error code from the list. This error only appears
while browsing through the error list in the status menu.
Error code display:
999 Error in Errorlist
Error code status menu:
999 Error in Errorlist
Error handling:
• Move on in the error list or leave the submenu “Error”.

Transport, Storage and Disposal
156
VTK-TR-BA-171e-1 - User Manual MDS 3080-V
12 Transport, Storage and Disposal
12.1 Transport
For shipment, the system is packed in an appropriate case. Preserve it for further transport
procedures in the future. The necessity can never be excluded, e.g. for purposes of maintenance.
Observe the following measures.
– If the original packaging is no longer available, select a suitable material for the package.
– Pack in a way that protects the delivery against shock and vibrations.
– Fill empty spaces with appropriate material (e.g. paper, insulating air cushion, Styrofoam).
– Decontaminate the system completely before shipping it.
– Fill out the corresponding form completely (see paragraph 14.9, page 176). Attach it, clearly
visible, at the exterior of the package.
CAUTION
Decontamination
The customer is liable for all consequences resulting from insufficient decontamination of the
system (health of transport personnel)!
12.2 Storage
The environmental conditions in the room intended for storage should fulfill the requirements of
the system.
Observe the following storage conditions.
• The room should be sufficiently ventilated and free of dust
• Admissible temperature: Between +5 °C and +30 °C
• Rel. humidity < 50 %
• The system should not be stored together with fuel, solvents, lubricants, acids and other
chemicals able to develop vapors, which may aggress to the surface of the components.
Air-tight foil sheathing with included desiccants may be useful.
12.3 Recycling and Disposal
The entire packaging consists of 100 % recyclable material.
At the end of its lifetime, the product itself should be discarded in conformity
with local regulations. Prescriptions with respect to handling of electrical
scrap have to be observed.

Spare Parts and Tools
VTK-TR-BA-171e-1 - User Manual MDS 3080-V
157
13 Spare Parts and Tools
Here are some of the most important spare and extra parts as well as tools. For the latest and
complete list, please check our homepage at www.vermes.com.
13.1 Nozzle Fixation Nuts
Nozzle Fixation Nut NFN-TA-HC
Order no. 1014181
Double Channel Nozzle Fixation Nut
NFN-TA
Order no. 1016198
Nozzle Fixation Nut NFN-TA-IB
Order no. 1014494
Nozzle Fixation Nut NFN-TA-HC-IB
Order no. 1014493
Nozzle Fixation Nut NFN-TA-
HC+10mm
Order no. 1015451
Tab. 33: Nozzle Fixation Nuts