Control Unit MDC
VTK-TR-BA-171e-1 - User Manual MDS 3080-V
21
4.2 Front Side
Fig. 1: Front Side
1 Mounting hole
2 LC display
3 Keypad with 12 soft keys
4 Control lamp for First Drop
(“function”) “red”
5 Control lamp for adjust “green“
6 Control lamp for adjust „red“
7 Control lamp for status of
heating „red“
LC display:
Data, service intervals and menu options are shown in the two lines of the illuminated LC display
(LCD = “liquid crystal display”). The current menu option appears in the upper line. The lower line
contains the current parameter value for dispensing, according to numerical data to be modified.
Keypad with 12 softkeys:
Menu options to control the system are selected by means of these keys.
Details for changing parameters etc. are explained in a following chapter (see paragraph 4.4,
page 25).
Control lamp for First Drop (“function”):
This red LED indicates, if there is a problem during the first drop adjust. It is also ON during the
start-up of the MDC, while the power supply unit is initialized.
Control lamps for adjust (“green” and “red”):
During the adjust, these two control lamps indicates the adjust result that is internally monitored:
• Green: the adjust value is OK. Press [Enter]-key to confirm it.
• Red: the adjust value is too high. You have to lower it according to the instructions of this
manual (see paragraph 6.5 "The Adjust Process", page 52).

Control Unit MDC
22
VTK-TR-BA-171e-1 - User Manual MDS 3080-V
You can de-/activate the control lamps for adjust. It makes no difference during the adjust (as
above). But it affects the lamps outside of the adjust (as below).
Outside of the adjust: (if the control lamps for adjust are activated)
• Red: the MDC detects an error.
• Green: the valve detects that the adjust value falls inside the adjust range.
Outside of the adjust: (if the control lamps for adjust are deactivated)
• Red: the MDC detects an error (except for an adjust error).
• The green adjust-LED stays OFF.
To de-/activate the control lamps for adjust, enter the service code 1100 (see paragraph 4.5.6
"Submenu “Service-Option”", page 33). Alternatively, you can use the serial command
SETADJLED:OFF/ON (see paragraph 8.1.2.2 "Explanations", page 88).
Control lamp for status of heating:
This red LED indicates the activity of the heating.
Heating activated – LED ON
Heating deactivated – LED OFF
Heat-up phase – LED flashes

Control Unit MDC
VTK-TR-BA-171e-1 - User Manual MDS 3080-V
23
4.3 Back Side
Fig. 2: Back side
1 Thermocouple „Type K“ socket
2 Mains switch
3 AUX socket
4 Socket for heating
5 RS-232C interface (9-pin)
6 Sensor socket
7 Actuator socket
8 PLC-interface (15-pin)
9 Mains connector
Thermocouple socket (TCK):
This element is intended for connection of thermocouples “Type K“.
Mains switch:
By means of this illuminated switch, the unit is switched ON and OFF.
AUX socket:
This connector can be used for supply of an external device (e.g. an optocoupler) or to control
certain parameter setups or scenarios (see paragraph 7.9.3 "Scenario Selection with Select Pins",
page 67). It is one of the three interfaces of the control unit (see paragraph 8.3, page 128 for the
pin assignments).
Actuator socket:
For the connection of the actuator cable.