QKN= j~åÇ~íçêó=péÉÅáÑáÅ~íáçåë=Ñçê=ÅçãéêÉëëÉÇ=~áê=ëìééäáÉë=
Inlet pressure to filter/mist
separator/regulator unit:
6 bar (90 psig) minimum.
OUTLET TO SYSTEM:
Oil content:
Less than 10 ppm
Maximum Moisture Content: -3ºC (25ºF)
Filtration: Maximum particle size of 0.3 microns
Regulator outlet pressure: Adjustable from 3 bar to 6 bar (45 psig to 90
psig). 6 bar must be the maximum provided.
Maximum flow rate:
135 litres/minute (5 scfm).
Pressure monitoring: 0 to 10 bar (0 to 150 psi) pressure gauge.
Services Specifications
Issue 16: April 05 Page 12 of 22 Printed: 6-Jan-06, 8:44
System Manual lñÑçêÇ=fåëíêìãÉåíë=mä~ëã~=qÉÅÜåçäçÖó== lfmq=póëíÉãë
RK= káíêçÖÉå=
Nitrogen is required to vent and purge process chambers, load locks and pumps.
2 MICRON
FILTER
REGULATOR
1/4"
SWAGELOK
CONNECTOR
TO PLASMALAB
OR IONFAB
SYSTEM
TO PUMP
PURGE
CONNECTOR
0 TO 1 SLM
FLOW CONTROL
ROTAMETER
(IF REQUIRED)
Fig 5.1: Recommended nitrogen supply installation
Semiconductor grade fittings and pressure regulators, together with electropolished stainless
steel tube must be used to ensure that gas quality is not degraded. Purges to turbo pumps are
supplied by OIPT as part of the system. The customer will usually need to fit a purge to the
rotary pump. See below.
RKN= j~åÇ~íçêó=ëéÉÅáÑáÅ~íáçåë=Ñçê=káíêçÖÉå=ëìééäáÉë=
Pipework fittings and
pressure regulators:
Semiconductor grade
Gas handling tubing: Electropolished stainless steel
Purity: At least 99.99% or higher to satisfy process
requirements.
Filtration: 2 micron filter mounted adjacent to the
system.
Regulation: 0.5 to 5 bar (7.5 to 75 psig)
Minimum pressure at input
to system:
3 bar (45 psig). Certain pumps, for example
Edwards Drystar pumps, may need up to 5 bar
(75psig) to ensure satisfactory purging. Check
with the vendor's instructions.
Rotary pump purging
It is the customer’s responsibility to ensure that a rotary pump purge
connection is fitted and used correctly. This is needed to ensure the protection
of the pumping system from the customer’s process, and may also be required
by local safety regulations. Because customers’ requirements vary, components
are not supplied automatically by OIPT. Kits of parts are available if required.
Contact OIPT for further information.
It may be safe to omit this feature on certain systems, such as Ionfab machines
running inert gas processes. However, unless the customer has written
agreement on this point from OIPT, any damage caused by the omission cannot
be covered by the system warranty.
Services Specifications
Printed: 6-Jan-06, 8:44 Page 13 of 22 Issue 16: April 05
lfmq=póëíÉãë lñÑçêÇ=fåëíêìãÉåíë=mä~ëã~=qÉÅÜåçäçÖó== System Manual
SK= mêçÅÉëë=Ö~ëÉë=
REGULATOR 0.5 BAR TO 5 BAR
(7 TO 75 PSIG) LOCATED
ADJACENT TO THE SYSTEM
SUPPLY TO
GAS POD
OR INTERNAL
GAS LINE
MINIMUM PRESSURE
3 BAR (45 PSIG)
ALL FITTINGS,
REGULATOR AND
FILTER TO BE
SEMICONDUCTOR
GRADE
A
LL TUBING TO BE
ELECTROPOLISHED
STAINLESS STEEL
NOTES:
Fig 6.1: Recommended process gas supply installation
Semiconductor grade fittings and pressure regulators, together with electropolished stainless
steel tube must be used to ensure that gas quality is not degraded.
SKN= j~åÇ~íçêó=ëéÉÅáÑáÅ~íáçåë=Ñçê=éêçÅÉëë=Ö~ë=ëìééäáÉë=
Pipework fittings and pressure
regulators:
Semiconductor grade
Gas handling tubing: Electropolished stainless steel
Purity: At least 99.99% or higher to satisfy
process requirements.
Filtration: A 2-micron filter is fitted to each
gas line supplied as part of the
system. For other grades of filter,
please consult OIPT.
Regulation: 0.5 to 5 bar (7.5 to 75 psig)
Minimum pressure at input to system: 2 bar (30 psig)
SKNKN= póëíÉãë=ìëáåÖ=eÉäáìã=ï~ÑÉê=ÅççäáåÖ=
The specifications for the Helium gas supply are as for those of the process gases given in sub-
sections 6 and 6.1 with the exception that the maximum pressure at the inlet to the
pressure controller must not exceed 3.5 bar (43 psig). The design of the Helium
pressure controller is such that it can be destroyed by higher pressures.
Services Specifications