MR8740T_Quick guide_eng_20191016H.pdf - 第127页

121 Specications of the Options T emperature characteristics ±0.02% f.s./°C Effect of radiated radio-frequency electromagnetic eld ±5% f.s. (max.) at 3 V/m (with the lter set at 5 Hz) Effect of conducted radio-frequen…

100%1 / 172
120
Specications of the Options
Model U8975 4ch Analog Unit
Conditions of guaranteed accuracy
Specied under the following conditions: installed in a Memory HiCorder and operated at 23°C ±5°C (73°F ±9°F) and
80% RH or less after a half-hour (at least) warm-up and execution of zero-adjustment.
Operating
environment
Indoors, Pollution Degree 2, Operating altitude: up to 2000 m (6562 ft.)
Operating
temperature and
humidity
0°C to 40°C (32°F to 104°F), 80% RH or less (no condensation)
Storage temperature
and humidity
−10°C to 50°C (14°F to 122°F)
For −10°C or higher but lower than 40°C (14°F or higher but lower than 104°F), 80% RH
or less (no condensation)
For 40°C or higher but lower than 45°C (104°F or higher but lower than 113°F), 60% RH or
less (no condensation)
For 45°C to 50°C (113°F to 122°F), 50% RH or less (no condensation)
Standards Safety: EN61010
EMC: EN61326 Class A
Dimensions Approx. 106W × 19.8H × 196.5D mm (4.17″W × 0.78″H × 7.74″D)
Mass Approx. 250 g (8.8 oz.)
Product warranty
period
3 years
Measurement range 4, 10, 20, 40, 100, 200 V f.s.
Maximum input
voltage
200 V DC
Maximum rated
voltage to earth
300 V AC, DC, Measurement category II (between each input channel and the enclosure,
between any two of input channels)
Anticipated transient overvoltage: 2500 V
Measurement
terminal
Insulated BNC terminal
Number of channels 4 channels
Frequency
characteristics
DC to 2 MHz, −3 dB
Noise 5 mV p-p (typ., 10 mV p-p (max.) (with the highest sensitivity range with input terminals
connected with each other)
Input type Unbalanced input (oating)
Input coupling DC/GND
Input resistance 1 M
Ω
±1%
Input capacitance 30 pF ±10 pF (at 100 kHz)
A/D resolution 16 bits (± f.s. = ±32000 data points)
Maximum sampling
rate
5 MS/s
Guaranteed accuracy
period
1 year
Guaranteed
accuracy period after
adjustment made by
Hioki
1 year
Measurement
accuracy
±0.1% f.s. (with the lter set at 5 Hz)
121
Specications of the Options
Temperature
characteristics
±0.02% f.s./°C
Effect of radiated
radio-frequency
electromagnetic eld
±5% f.s. (max.) at 3 V/m (with the lter set at 5 Hz)
Effect of conducted
radio-frequency
electromagnetic eld
±5% f.s. (max.) at 3 V (in the 10 V f.s. range, with the lter set at 5 Hz and 1 V DC
inputted)
Common-mode
rejection ratio
80 dB or more (at 50 Hz / 60 Hz, a signal source resistance of 100
Ω
)
Low-pass lter Off, 5 ±50%, 500 ±50%, 5 k ±50%, 200 k ±50% (Hz); −3 dB
5
Specications
122
Specications of the Options
Model U8977 3CH Current Unit (The rmware of ver. 2.00 or later is
required to install the module.)
Conditions of guaranteed accuracy
Specied under the following conditions: installed in a Memory HiCorder and operated at 23°C ±5°C (73°F ±9°F) and
80% RH or less after a half-hour (at least) warm-up and execution of zero-adjustment.
Operating
environment
Indoors, Pollution Degree 2, Operating altitude: up to 2000 m (6562 ft.)
Operating
temperature and
humidity
0°C to 40°C (32°F to 104°F), 80% RH or less (no condensation)
Storage temperature
and humidity
−10°C to 50°C (14°F to 122°F)
For −10°C or higher but lower than 40°C (14°F or higher but lower than 104°F), 80% RH
or less (no condensation)
For 40°C or higher but lower than 45°C (104°F or higher but lower than 113°F), 60% RH or
less (no condensation)
For 45°C to 50°C (113°F to 122°F), 50% RH or less (no condensation)
Standards Safety: EN61010
EMC: EN61326 Class A
Dimensions Approx. 106W × 19.8H × 196.5D mm (4.17″W × 0.78″H × 7.74″D)
Mass Approx. 250 g (8.8 oz.)
Product warranty
period
3 years
Options Model CT9900 Conversion Cable (PL23 receptacle–ME15W plug)
Model CT9920 Conversion Cable (PL14 receptacle–ME15W plug)
Number of input
channels
3 channels