MR8740T_Quick guide_eng_20191016H.pdf - 第133页
127 Specications of the Options Model U8979 Charge Unit (The rmware of ver . 2.00 or later is required to install the module.) 1. General specications Operating environment Indoors, Pollution Degree 2, Operating altit…

126
Specications of the Options
Guaranteed
accuracy period after
adjustment made by
Hioki
1 year
Measurement
accuracy
±0.3% f.s. (with the lter set at 5 Hz)
Temperature
characteristics
±0.045% f.s./°C
Effect of radiated
radio-frequency
electromagnetic eld
±5% f.s. (max.) at 3 V/m (with the lter set at 5 Hz)
Effect of conducted
radio-frequency
electromagnetic eld
±5% f.s. (max.) at 3 V (in the 10 V range, with the lter set at 5 Hz and 1 V DC inputted)
Common-mode
rejection ratio
80 dB or more (at 50 Hz / 60 Hz, a signal source resistance of 100
Ω
)
Low-pass lter Off, 5 ±50%, 500 ±50%, 5 k ±50%, 200 k ±50% (Hz); −3 dB

127
Specications of the Options
Model U8979 Charge Unit (The rmware of ver. 2.00 or later is required
to install the module.)
1. General specications
Operating
environment
Indoors, Pollution Degree 2, Operating altitude: up to 2000 m (6562 ft.)
Operating
temperature and
humidity
0°C to 40°C (32°F to 104°F), 80% RH or less (no condensation)
Storage temperature
and humidity
−10°C to 50°C (14°F to 122°F)
For −10°C or higher but lower than 40°C (14°F or higher but lower than 104°F), 80% RH
or less (no condensation)
For 40°C or higher but lower than 45°C (104°F or higher but lower than 113°F), 60% RH or
less (no condensation)
For 45°C to 50°C (113°F to 122°F), 50% RH or less (no condensation)
Standards Safety: EN61010
EMC: EN61326 Class A
Dimensions Approx. 106W × 19.8H × 196.5D mm (4.17″W × 0.78″H × 7.74″D)
Mass Approx. 250 g (8.8 oz.)
Product warranty
period
3 years
Guaranteed accuracy
period
1 year
2. Specications of input, output, and measurement
-1. Common specications
Number of channels 2 channels
Measurement mode Charge, pre-amplier, voltage (selectable for each channel)
Input type Unbalanced input (oating)
In each channel, the voltage input terminal has the same-potential ground as the charge
input terminal.
Common-mode
rejection ratio
80 dB or more (at 50 Hz / 60 Hz, a signal source resistance of 100
Ω
)
Anti-aliasing lters Cutoff frequency (fc)
20, 40, 80, 200, 400, 800, 2 k, 4 k, 8 k, 20 k, 40 k (Hz)
Switchable between on and off; the instrument automatically chooses a cutoff frequency
along with the sampling rate.
Attenuation property: −66 dB or higher at a cutoff frequency of 1.5 × fc
Maximum sampling
rate
200 kS/s
A/D resolution 16 bits (± f.s. = ±25,000 data points)
Maximum rated
voltage to earth
30 V AC, 60 V DC
(between each input channel and the enclosure, between any two of input channels)
Anticipated transient overvoltage: 330 V
-2. Voltage input
Measurement range 10, 20, 40, 100, 200, 400 mV f.s., 1, 2, 4, 10, 20, 40 V f.s.
5
Specications

128
Specications of the Options
Maximum input
voltage
40 V DC
Frequency
characteristics
DC to 50 kHz −3 dB (DC-coupled)
1 Hz to 50 kHz −3 dB (AC-coupled, at the lower cutoff frequency of 1 Hz ±50%)
Noise 80 μV p-p (typ.), 120 μV p-p (max.), with the highest sensitivity range and the terminals of
each input connector connected with each other
Input resistance 1 M
Ω
±1%
Input capacitance 200 pF or less (at 100 kHz)
Input coupling AC/DC/GND
Input terminals Insulated BNC terminal
-3. Charge input
Supported detector Charge-output acceleration detector
Measurement
sensitivity
0.1 pC/(m/s
2
) to 10 pC/(m/s
2
)
Measurement range 40, 100, 200, 400, 1 k, 2 k, 4 k, 10 k, 20 k, 40 k, 100 k, 200 k m/s
2
f.s.
Measurement sensitivity: 0.1 pC/(m/s
2
) to 0.25 pC/(m/s
2
)
20, 40, 100, 200, 400, 1 k, 2 k, 4 k, 10 k, 20 k, 40 k, 100 k m/s
2
f.s.
Measurement sensitivity: 0.251 pC/(m/s
2
) to 0.5 pC/(m/s
2
)
10, 20, 40, 100, 200, 400, 1 k, 2 k, 4 k, 10 k, 20 k, 40 k m/s
2
f.s.
Measurement sensitivity: 0.501 pC/(m/s
2
) to 1.0 pC/(m/s
2
)
4, 10, 20, 40, 100, 200, 400, 1 k, 2 k, 4 k, 10 k, 20 k m/s
2
f.s.
Measurement sensitivity: 1.001 pC/(m/s
2
) to 2.5 pC/(m/s
2
)
2, 4, 10, 20, 40, 100, 200, 400, 1 k, 2 k, 4 k, 10 k m/s
2
f.s.
Measurement sensitivity: 2.501 pC/(m/s
2
) to 5.0 pC/(m/s
2
)
1, 2, 4, 10, 20, 40, 100, 200, 400, 1 k, 2 k, 4 k m/s
2
f.s.
Measurement sensitivity: 5.001 pC/(m/s
2
) to 10.0 pC/(m/s
2
)
Frequency
characteristics
1.5 Hz to 50 kHz, −3 dB
Maximum input
charge
±500 pC (when one of the higher six ranges is chosen)
±50,000 pC (when one of the lower six ranges is chosen)
Input coupling AC/GND
Input terminals Miniature connector (#10-32UNF)
-4. Input from acceleration sensor with the built-in pre-amplier
Supported detector Acceleration detector with built-in pre-amplier
Measurement
sensitivity
0.1 mV/(m/s
2
) to 10 mV/(m/s
2
)
Measurement range 40, 100, 200, 400, 1 k, 2 k, 4 k, 10 k, 20 k, 40 k, 100 k, 200 k m/s
2
f.s.
Measurement sensitivity: 0.1 mV/(m/s
2
) to 0.25 mV/(m/s
2
)
20, 40, 100, 200, 400, 1 k, 2 k, 4 k, 10 k, 20 k, 40 k, 100 k m/s
2
f.s.
Measurement sensitivity: 0.251 mV/(m/s
2
) to 0.5 mV/(m/s
2
)
10, 20, 40, 100, 200, 400, 1 k, 2 k, 4 k, 10 k, 20 k, 40 k m/s
2
f.s.
Measurement sensitivity: 0.501 mV/(m/s
2
) to 1.0 mV/(m/s
2
)
4, 10, 20, 40, 100, 200, 400, 1 k, 2 k, 4 k, 10 k, 20 k m/s
2
f.s.
Measurement sensitivity: 1.001 mV/(m/s
2
) to 2.5 mV/(m/s
2
)
2, 4, 10, 20, 40, 100, 200, 400, 1 k, 2 k, 4 k, 10 k m/s
2
f.s.
Measurement sensitivity: 2.501 mV/(m/s
2
) to 5.0 mV/(m/s
2
)
1, 2, 4, 10, 20, 40, 100, 200, 400, 1 k, 2 k, 4 k m/s
2
f.s.
Measurement sensitivity: 5.001 mV/(m/s
2
) to 10.0 mV/(m/s
2
)