140 Specications of the Options Measurement accuracy DC voltage 5 V range ±0.030 % of rdg. ± 700 µV T emperature coefcient Add the following value when operating the instrument in temperature ranges of 0°C to 18°C and …

100%1 / 172
139
Specications of the Options
Output accuracy
DC voltage 5 V range ±0.035% of setting ± 800 µV
Temperature
coefcient
Add the following value when operating the instrument
in temperature ranges of 0°C to 18°C and of 28°C to
40°C.
[0.1 × (output accuracy)] /°C
Output resistance 100 m
Ω
or less
Overshoot When an output changes from 0% to 100% of a setting
value
Less than 5% (reference value with resistance load)
Output settling time Time an output signal takes to settle within 0.1% of the
range value when the signal changes from 0% to 100%
of the range
5 ms (reference value with no load)
DC current 5 mA range ±0.050% of setting ± 4.0 µA
Temperature
coefcient
Add the following value when operating
the instrument in temperature ranges of
0°C to 18°C and of 28°C to 40°C.
[0.1 × (output accuracy)] /°C
1 mA range ±0.050% of setting ± 800 nA
Temperature
coefcient
Add the following value when operating
the instrument in temperature ranges of
0°C to 18°C and of 28°C to 40°C.
[0.1 × (output accuracy)] /°C
250 µA range ±0.050% of setting ± 200 nA
Temperature
coefcient
Add the following value when operating
the instrument in temperature ranges of
0°C to 18°C and of 28°C to 40°C.
[0.1 × (output accuracy)] /°C
50 µA range ±0.050% of setting ± 40 nA
Temperature
coefcient
Add the following value when operating
the instrument in temperature ranges of
0°C to 18°C and of 28°C to 40°C.
[0.1 × (output accuracy)] /°C
Overshoot When an output changes from 0% to 100% of a setting
value
Less than 5% (reference value with resistance load)
Output settling time Time an output signal takes to settle within 0.1% of the
range value when the signal changes from 0% to 100%
of the range
5 ms (reference value with short-circuited load)
Resistance Current generation
range
Voltage measurement 0.5 V range Voltage measurement 5 V range
Percent of output
Floor [
Ω
]
Percent of output
Floor [
Ω
]
5 mA range
0.09 + 0.4000 /
I
0.10 /
I
+ 0.1 0.09 + 0.0040 /
I
0.70 /
I
+ 0.1
1 mA range
0.09 + 0.0800 /
I
0.10 /
I
+ 0.1 0.09 + 0.0800 /
I
0.70 /
I
250 µA range
0.09 + 0.0200 /
I
0.10 /
I
0.09 + 0.0200 /
I
0.70 /
I
50 µA range
0.09 + 0.0040 /
I
0.10 /
I
0.09 + 0.0040 /
I
0.70 /
I
Current values
I
are represented in milliamperes.
Output settling time Time an output signal takes to settle within 200% of
the accuracy specication centered on the nal value
(A setting resistance and an output resistance of a
connected object are in a ratio 1:1, with no capacitive
load)
When connected-object presumption is not performed:
1.5 s or less (response coefcient of resistance
generation: 0.0002)
When connected-object presumption is performed: 100
ms or less
5
Specications
140
Specications of the Options
Measurement
accuracy
DC
voltage
5 V range ±0.030 % of rdg. ± 700 µV
Temperature
coefcient
Add the following value when operating the
instrument in temperature ranges of 0°C to 18°C
and of 28°C to 40°C
[0.05 × (measurement accuracy)] /°C
5 V range
(for resistance
output function
only)
±0.030 % of rdg. ± 100 µV
Temperature
coefcient
Add the following value when operating the
instrument in temperature ranges of 0°C to 18°C
and of 28°C to 40°C
[0.05 × (measurement accuracy)] /°C
DC
current
5 mA range ±0.040 % of rdg. ± 900 nA
Temperature
coefcient
Add the following value when operating the
instrument in temperature ranges of 0°C to 18°C
and of 28°C to 40°C
[0.05 × (measurement accuracy)] /°C
1 mA range ±0.040 % of rdg. ± 180 nA
Temperature
coefcient
Add the following value when operating the
instrument in temperature ranges of 0°C to 18°C
and of 28°C to 40°C
[0.05 × (measurement accuracy)] /°C
250 µA range ±0.040 % of rdg. ± 45 nA
Temperature
coefcient
Add the following value when operating the
instrument in temperature ranges of 0°C to 18°C
and of 28°C to 40°C
[0.05 × (measurement accuracy)] /°C
50 µA range ±0.040 % of rdg. ± 10 nA
Temperature
coefcient
Add the following value when operating the
instrument in temperature ranges of 0°C to 18°C
and of 28°C to 40°C
[0.05 × (measurement accuracy)] /°C
Effect of radiated
radio-frequency
electromagnetic eld
1% f.s. (Max.) at 3 V/m (for DC voltage output function)
Effect of conducted
radio-frequency
electromagnetic eld
1% f.s. (Max.) at 3 V (for DC voltage output function)
No accuracy is specied for AC voltage output and AC current measurement.
141
Specications of the Options
Self-diagnosis
Operation
comprehensive
test
The instrument outputs the following signals with the
voltage and current functions, measuring them using the
internal measurement circuit.
If a measured value differs from the corresponding output
setting value by more than the threshold value, the
instrument gives itself a fail judgment.
The instrument returns a judgment result, output setting
value, and measured value in response to a query.
The threshold value of 0% causes the instrument to fail to
judge.
Voltage function 0 V, .0.5 V, 5 V
Current function 0 µA, 50 µA, 55 µA, 250 µA,
275 µA, 1 mA, 1.1 mA, 5 mA
Threshold
value
Setting
range
0% to 10%
Setting resolution: 1%
Default
setting
0%
Simple test Using the internal measurement circuit, the instrument
measures an signal outputted when a command is sent.
The instrument returns an output setting value and
measured value in response to a query.
Switching OUTPUT
terminals
Relay switching time
Operation
OPEN:
SHORT:
NORMAL:
Opens the circuit of the OUTPUT terminal.
Connects the OUTPUT terminal to ground. (10
Ω
or less with a
load current of 5 mA)
Outputs the specied signal across the OUTPUT terminal.
Default
setting
OPEN
Connected-object
presumption
Operation With the resistance generation function, the instrument presumes an output
resistance value of the connected object, capacitance value between the
OUTPUT and GND terminals, and voltage value across them.
Optimizes a resistance-generation response coefcient based on the
presumption.
When the time constant based on an output resistance value and inter-
terminal capacitance value is less than 10 ms, the inter-capacitance is
determined to be 0 pF.
Calculation
time
1.6 seconds (max.)
Offset cancel
Operation
Measures an output-circuit offset with the measurement circuit and outputs
a signal with the offset subtracted.
Default
setting
Off
5
Specications