MR8740T_Quick guide_eng_20191016H.pdf - 第130页

124 Specications of the Options Input resistance 1 M Ω ±1% Maximum input current Depends on a sensor . Maximum rated voltage to earth Non-isolated Maximum sampling rate 5 MS/s A/D resolution 16 bits Guaranteed accuracy …

100%1 / 172
123
Specications of the Options
Supported current
sensors
Directly connectable current sensors
Model 9272-05 Clamp on Sensor (20 A/200 A AC)
Model CT6841-05 AC/DC Current Probe (20 A)
Model CT6843-05 AC/DC Current Probe (200 A)
Model CT6844-05 AC/DC Current Probe (500 A,
φ
20 mm)
Model CT6845-05 AC/DC Current Probe (500 A,
φ
50 mm)
Model CT6846-05 AC/DC Current Probe (1000 A)
Model CT6862-05 AC/DC Current Sensor (50 A)
Model CT6863-05 AC/DC Current Sensor (200 A)
Model 9709-05 AC/DC Current Sensor (500 A)
Model CT6904 AC/DC Current Sensor (500 A)
Model CT6865-05 AC/DC Current Sensor (1000 A)
Model CT6875 AC/DC Current Sensor (500 A)
Model CT6876 AC/DC Current Sensor (1000 A)
Model CT6877 AC/DC Current Sensor (2000 A)
Current sensors that require Model CT9920 for connection
Model CT7631 AC/DC Current Sensor (100 A)
Model CT7636 AC/DC Current Sensor (600 A)
Model CT7642 AC/DC Current Sensor (2000 A)
Model CT7731 AC/DC Auto-Zero Current Sensor (100 A)
Model CT7736 AC/DC Auto-Zero Current Sensor (600 A)
Model CT7742 AC/DC Auto-Zero Current Sensor (2000 A)
Model CT7044 AC Flexible Current Sensor (6000 A,
φ
100 mm)
Model CT7045 AC Flexible Current Sensor (6000 A,
φ
180 mm)
Model CT7046 AC Flexible Current Sensor (6000 A,
φ
254 mm)
Measurement range For the directly connectable current sensors, the instrument automatically recognizes the
rating of the current sensor.
2 A, 4 A, 10 A, 20 A, 40 A, 100 A (rating: 20 A)
4 A, 10 A, 20 A, 40 A, 100 A, 200 A (rating: 50 A)
20 A, 40 A, 100 A, 200 A, 400 A, 1000 A (rating: 200 A)
40 A, 100 A, 200 A, 400 A, 1000 A, 2000 A (rating: 500 A)
100 A, 200 A, 400 A, 1000 A, 2000 A, 4000 A (rating: 1000 A)
For the current sensors that require Model CT9920 for connection, choose a conversion
rate or a model number.
200 A (Model CT7631, Model CT7731)
200 A, 400 A, 1000 A (Model CT7636, Model CT7736)
2000 A, 4000 A (Model CT7642, Model CT7742)
2000 A, 4000 A, 10000 A (Model CT7044, Model CT7045, Model CT7046)
2000 A, 4000 A, 10000 A, 20000 A, 40000 A, 100000 A (0.1 mV/A)
200 A, 400 A, 1000 A, 2000 A, 4000 A, 10000 A (1 mV/A)
20 A, 40 A, 100 A, 200 A, 400 A, 1000 A (10 mV/A)
2 A, 4 A, 10 A, 20 A, 40 A, 100 A (100 mV/A)
0.2 A, 0.4 A, 1 A, 2 A, 4 A, 10 A (1000 mV/A)
Frequency
characteristics
DC to 2 MHz (±3 dB)
Noise 10 mA p-p (max.) (using the 20 A sensor, with the 2 A f.s. range, with input terminals
connected with each other)
Low-pass lter Off, 5 Hz, 500 Hz, 5 kHz, 200 kHz ±50%; −3 dB
Input type Current sensor
Measurement
terminal
LR10-DC12BR (Hirose connector)
Input coupling DC/GND
5
Specications
124
Specications of the Options
Input resistance 1 M
Ω
±1%
Maximum input
current
Depends on a sensor.
Maximum rated
voltage to earth
Non-isolated
Maximum sampling
rate
5 MS/s
A/D resolution 16 bits
Guaranteed accuracy
period
1 year
Guaranteed
accuracy period after
adjustment made by
Hioki
1 year
Measurement
accuracy
±0.3% f.s. + (accuracy of current sensor)
Temperature
characteristics
±0.045% f.s./°C
Effect of radiated
radio-frequency
electromagnetic eld
±5% f.s. (max.) at 3 V/m (with the lter set at 5 Hz)
Effect of conducted
radio-frequency
electromagnetic eld
±5% f.s. (max.) at 3 V (using the 20 A sensor, with the 20 A f.s. range, with the lter set at
5 Hz, with a current of 2 A DC inputted)
125
Specications of the Options
Model U8978 4CH Analog Unit (The rmware of ver. 2.00 or later is
required to install the module.)
Conditions of guaranteed accuracy
Specied under the following conditions: installed in a Memory HiCorder and operated at 23°C±5°C (73°F±9°F) and
80% RH or less after a half-hour (at least) warm-up and execution of zero-adjustment.
Operating
environment
Indoors, Pollution Degree 2, Operating altitude: up to 2000 m (6562 ft.)
Operating
temperature and
humidity
0°C to 40°C (32°F to 104°F), 80% RH or less (no condensation)
Storage temperature
and humidity
−10°C to 50°C (14°F to 122°F)
For −10°C or higher but lower than 40°C (14°F or higher but lower than 104°F), 80% RH
or less (no condensation)
For 40°C or higher but lower than 45°C (104°F or higher but lower than 113°F), 60% RH or
less (no condensation)
For 45°C to 50°C (113°F to 122°F), 50% RH or less (no condensation)
Standards Safety: Model EN61010
EMC: EN61326 Class A
Dimensions Approx. 106W × 19.8H × 196.5D mm (4.17″W × 0.78″H × 7.74″D)
Mass Approx. 250 g (8.8 oz.)
Product warranty
period
3 years
Measurement range 100, 200, 400 mV f.s., 1, 2, 4, 10, 20, 40 V f.s.
Maximum input
voltage
(Direct input) 40 V DC
(In combination with Model 9665) 400 V DC
Maximum rated
voltage to earth
(Direct input)
30 V AC, 60 V DC (between each input channel and the enclosure, between any two of
input channels)
(In combination with Model 9665)
300 V AC, DC, Measurement category II (between each input channel and the enclosure,
between any two of input channels)
Measurement
terminal
Insulated BNC terminal
Number of channels 4 channels
Frequency
characteristics
DC to 2 MHz, −3 dB
Noise 500 μV p-p (typ.), 1 mV p-p (max.) (with the highest sensitivity range and the terminals of
each input connector connected with each other)
Input type Unbalanced input (oating)
Input coupling DC/GND
Input resistance 1 M
Ω
±1%
Input capacitance 30 pF ±10 pF (at 100 kHz)
A/D resolution 16 bits (± f.s. = ±32000 data points)
Maximum sampling
rate
5 MS/s
Guaranteed accuracy
period
1 year
5
Specications