Regulating the Zero Position (Zero-Adjustment)
Re-execute zero-adjustment in the following cases:
• After replacing any modules
• After cycling the instrument
• After initializing the instrument
• After switching between DC mode and RMS mode on Model 8971 Current Unit, Model 8972 DC/
RMS Unit, or Model U8974 High Voltage Unit.
• When the ambient temperature has signicantly changed
The zero position may drift*.
*: Drift:
A phenomenon where a shift in the operating point of an operational amplier causes a false
output. Drift can result from a change in temperature and component aging over a period of use.

65
Executing Calibration (For the Instrument With Model MR8990 Installed)
2.9 Executing Calibration (For the Instrument With
Model MR8990 Installed)
Executing calibration regulates the zero position of each channel in Model MR8990 Digital
Voltmeter Unit on the reference potential of the instrument. Calibration involves all ranges of all
channels.
Before executing calibration
• Warm up the instrument for about 30 minutes after the power-on to stabilize the internal
temperature of the modules before executing calibration.
• You cannot execute calibration during measurement. However, the instrument performs
calibration at the start of measurement with the calibration setting set to on.
• No key operation is acceptable during calibration.
• The time required for calibration varies depending on the type and number of modules installed in
the instrument (it may take several seconds).
Executing calibration
> [Channel] > [MR8990]
2
1
1
Click [Zero adjust].
2
Click [Execute].
The instrument starts to calibrate Model MR8990.
Execute calibration again in the following cases:
• After replacing any modules
• After cycling the instrument
• After initializing the instrument
• When the ambient temperature has signicantly changed
The zero position may drift.