MR8740T_Quick guide_eng_20191016H.pdf - 第138页
132 Specications of the Options Model MR8790 W aveform Generator Unit Conditions of guaranteed accuracy Specied under the following conditions: operated at 23°C±5°C (73°F±9°F) and 80% RH or less after a half-hour (at l…

131
Specications of the Options
Effect of radiated
radio-frequency
electromagnetic eld
±0.1% f.s. (max.) at 3 V/m (in the 1 V f.s. range)
Effect of conducted
radio-frequency
electromagnetic eld
±0.1% f.s. (max.) at 3 V (in the 1 V f.s. range)
Common-mode
rejection ratio
100 dB or more (at 50 Hz / 60 Hz, a signal source resistance of 100
Ω
)
5
Specications

132
Specications of the Options
Model MR8790 Waveform Generator Unit
Conditions of
guaranteed accuracy
Specied under the following conditions: operated at 23°C±5°C (73°F±9°F) and 80% RH
or less after a half-hour (at least) warm-up (when installed in a Memory HiCorder)
Product warranty
period
3 years
Guaranteed accuracy
period
1 year
Guaranteed
accuracy period after
adjustment made by
Hioki
1 year
Number of output
channels
4 channels (isolated between enclosure and output, and between every channel)
Self-diagnosis
function
Available (with the voltage and current monitors)
Switchable between
the voltage and
current monitors
Resolution: 5 µA (current monitor), 10 mV (voltage monitor)
Monitor accuracy: ±3.0% f.s. (f.s. = 10 V: voltage monitor, f.s. = 5 mA: current monitor)
Maximum output
current
±5 mA
Allowable load
resistance
2 k
Ω
or more
Output terminal SMB terminal
Output conguration Waveform output / open / short-circuit
Output relay
switching time
5 ms or less
Output protection The output current is limited to 40 mA (when the output terminals are are connected with
each other)
Maximum rated
voltage to earth
33 V rms AC or 70 V DC (between each output channel and enclosure, between any two
of output channels)
Anticipated transient overvoltage: 330 V
Withstanding voltage 350 V AC (sensitivity current: 1 mA) (between each output channel and enclosure,
between any two of output channels)
Operating
temperature and
humidity
Consistent with the specications of Memory HiCorder in which Model MR8790 is installed.
Operating
environment
Consistent with the specications of Memory HiCorder in which Model MR8790 is installed.
Storage temperature
and humidity
−20°C to 50°C (−4°F to 122°F), 90% RH or less (no condensation)
Dimensions Approx. 106W × 19.8H × 196.5D mm (4.17″W × 0.78″H × 7.74″D) (excluding protrusions)
Mass Approx. 230 g (8.1 oz.)
Standards Safety: EN61010
EMC: EN61326 Class A
Effect of radiated
radio-frequency
electromagnetic eld
±3% f.s. (max.) at 3 V/m (f.s. = 10 V)
Effect of conducted
radio-frequency
electromagnetic eld
±1% f.s. (max.) at 3 V (f.s. = 10 V)

133
Specications of the Options
Options Model L9795-01 Connection Cable (terminal type: SMB terminal - alligator clip)
Model L9795-02 Connection Cable (terminal type: SMB terminal - BNC terminal)
Maximum output
voltage
±10 V
Resolution 16 bits
Output frequency Output frequency: DC, 0 Hz to 20 kHz (sine wave)
Setting resolution: 1 Hz
Frequency accuracy: ±0.01% of setting
Amplitude Measurable range: 0 V p-p to 20 V p-p
Setting resolution: 1 mV
Amplitude accuracy: ±0.25% of setting ±2 mV p-p (1 Hz to 10 kHz),
±0.6% of setting ±2 mV p-p (more than 10 kHz to 20 kHz)
DC offset Setting range: −10 V to 10 V (The peak value, which is the total of amplitude and DC
offset, is limited to ±10 V)
Setting resolution: 1 mV, offset accuracy: ±3 mV
DC output Output accuracy: ±0.6 mV
5
Specications