45
Copying Settings (Copy Function)
3.5 Copying Settings (Copy Function)
You can copy settings of other channels, as well as the trigger settings.
The following procedure explains how to copy settings of another channel.
> [Func] > [Copy] > [Channel]
3
2
1
4
1
In the [Contents] area, click one or more of items that you want to copy.
Depending on module types, some items may not be able to be copied.
Basic Copies the mode, measurement range, coupling, L.P.F., division ratio, and module-
specic settings.
Display Copies the display setting (excluding comments).
Comment Copies a comment.
Scaling Copies the scaling setting.
2
Click the [Source] box, and then choose a source channel from the list.
3
Click the [Destination] box, and then choose a destination channel from the list.
Otherwise, click [All].
CH1-1
(Channel
selection)
Click this option when you would like to copy the settings to any one of the channels.
Select a destination channel from the list.
All Copies settings to all channels of UNIT 1 through UNIT 27.
4
Click [Copy].
After copying the settings, check that levels and setting values, which includes upper and lower values, and
the measurement range are appropriately set.
3
Advanced Functions

46
Conguring Measuring-Module-Specic Settings
3.6 Conguring Measuring-Module-Specic
Settings
You can congure advanced settings for each module.
Conguring Model 8968 High Resolution Unit settings
> [Channel] > [8968]
1
Click the area that includes [A.A.F].
The [A.A.F.] setting dialog box appears.
2
Click the [A.A.F.] box, and then choose [On] or [Off] from the list for the anti-aliasing lter
setting.
Enabling the anti-aliasing lter prevents aliasing distortion from inuencing FFT calculations.
The cutoff frequency changes automatically according to the sampling rate setting.
Off
Disables the anti-aliasing lter.
On Enables the anti-aliasing lter.
(Disabled when the external sampling is used, or the sampling rate is set at 100 kS/s
or faster)
3
Click [Close].
The setting dialog box closes.

47
Conguring Measuring-Module-Specic Settings
Conguring Model 8967 Temp Unit settings
> [Channel] > [8967]
1
2
3
4
5
6
1
Click the [Mode] box, and then choose a thermocouple type from the list.
Choose an option depending on the type of a thermocouple to be used.
Mode Measurable range Mode Measurable range
K
−200°C to 1350°C R 0°C to 1700°C
J −200°C to 1100°C S 0°C to 1700°C
E −200°C to 800°C B 400°C to 1800°C
T −200°C to 400°C W 0°C to 2000°C
N −200°C to 1300°C
2
Click the area that includes [RJC].
The setting dialog box appears.
3
Click the [RJC] box, and then choose between [Int.] and [Ext.] from the list for reference
junction compensation.
Int.
Executes the reference junction compensation inside the module.
(Measurement accuracy: The sum of the accuracy of the temperature measurement
and that of the reference junction compensation)
Ext. Does not execute the reference junction compensation inside the module.
(Measurement accuracy: The accuracy of the temperature measurement only)
When connecting a thermocouple directly to the module, choose [Int.].
When connecting a thermocouple via a reference junction device that includes a zero-point bath, choose [Ext.].
3
Advanced Functions