7 Alarm Messages
7.27 Cooling Water failure (option)
Operating Instructions
Version 1.5
7.27 Cooling Water failure (option)
Fig. 7-18 Cooling System Components
Fig. 7-19 Cooling System Components CN
Possible causes:
The cooling water pump on the
water tank cannot be started or is
defective.
The pressure sensor is defective.
Procedure:
Inspect all tubing and connec-
tions to the pump.
Inspect the sensor and the cool-
ing water pump and replace if
necessary.
Page 190 VISION XP+ VAC
7 Alarm Messages
7.28 Water amount is too low (option)
Operating Instructions
Version 1.5
7.28 Water amount is too low (option)
7.29 Water Leak (option)
Possible causes:
The water system has a leak.
There’s no water in the tank.
The sensor is defective.
Procedure:
Inspect all tubing and connections within the water system, and repair if
necessary.
Fill the water tank.
Inspect the sensor and replace if necessary.
Fig. 7-20 Tank with Water Level Sensor
Fig. 7-21 Tank with Water Level Sensor CN
Possible causes:
There is too little water in the tank.
The sensor is defective.
Procedure:
Fill in the mixture of cooling water
in (E) by means of a watering can.
Inspect the sensor.
E
E
VISION XP+ VAC Page 191
7 Alarm Messages
7.30 Cooling system failure (option)
Operating Instructions
Version 1.5
7.30 Cooling system failure (option)
External cooling
Procedure:
Check external cooling equipment. Switch on or repair external cooling
equipment.
Check temperature setpoint and hysteresis. Correct selected values.
Check coolant water level and replenish if necessary.
7.31 N2 consumption OK / too high (option)
Only with N2 consumption measuring. In the „Consumption Parameter“
mask a limit value is set. If it is exceeded, the alarm is triggered off (not in
the flushing phase).
7.32 N2/O2 Pressure Too Low (option)
Fig. 7-22 Servomotor, Main N2 Supply Line
The alarm is only activated in the
event of operation with nitrogen.
Possible causes:
Gas pressure is too low or nonex-
istent.
Supply pressure has fallen below
the selected alarm value.
The heat is turned off. The Belimo
servomotor in the main N2/O2
supply line is closed.
Procedure:
Inspect the N2/O2 supply line.
5bar Control