13. " DETERM INE C AMERA NO ZZLE OFFSE T" follows automatically: NOTE If, when you ar e about to individually selec t this menu option, y ou are no longer sur e whether you have in the meantime exchanged or tur…

100%1 / 138
9. The procedure "DETERMINE CALIBRATION FACTORS OF HEAD CAMERA" follows automatically:
The calibration plate is picked up, centered and put down again, the camera moves over the calibrati-
on plate, the calibration factors are determined, checked and displayed:
CAL. FACTORS CAMERA NO...
X1: -.0088 X2: 1.1783 X3: -.0149
Y2: -.0021 Y2: 1.4934 Y3: -.4704
VALUES X3 AND Y3 ARE NOT STORED BUT OCCUPIED WITH 0.
If the error message "CHECK MEASUREMENT NOT SUCCESSFUL" is displayed, the values deter-
mined are ambiguous.
NOTE
The next menu option can only be performed if the check measurement was correct.
Troubleshooting:
Press "RETURN" (values are not saved in this case) press "END" (PCB together with calibration plate
is conveyed to the output conveyor) remove PCB visual inspect calibration plate for oxidation or con-
tamination use new plate, if necessary repeat the complete measurement (from point 1 onwards).
10. After the measurement has been successfully completed, "DETERMINE ANGLE CORRECTION OF
D-AXIS" follows automatically.
The plate is picked up, centered, turned, put down and measured. The values determined incl. the de-
viation are displayed:
DETERMINE ANGLE OFFSET
ZERO POINT CORRECTION OLD : .......
ZERO POINT CORRECTION NEW : .......
ANGLE OFFSET MEASURED AND REFERENCE POINT FOUND
If the discrepancy (old / new) is within the acceptable tolerance, the reference point run is carried out
using the new values. Otherwise, the display appears as under point 11.
11.The measurement is repeated with the new values. If the placement head is not within the tolerance when the
measurement is repeated, an error message is released and a display of the difference determined, as for example:
HEAD OUT OF TOLERANCE
MEASURED VALUE DIFFERENCE AT 0 AND 180 DEGREES: ....... DIGIT
THE VALUE OF THE ZERO POINT CORRECTION WILL NOT BE STORED
PLEASE CORRECT MANUALLY
Locate the cause of the error (see listing below), eliminate the error and repeat the calibration, starting
with "DETERMINE CALIBRATION FACTORS".
Possible causes:
Rotation error of D-axisperform troubleshooting (see Service Manual HS-180, chapter 9,
29mm Head or Service Manual 30mm Head on HS-180)!
Circular fiducials of the calibration plate are poor / incomplete use a new calibration plate!
Calibration factors are incorrect Redetermine the calibration factors, then repeat the above
measurement!
12. If no error message is issued, the values determined are automatically transferred to the main memory.
User’s Manual AutoCal Program 7 AutoCal Measuring Sequence: Routine Checks on HS-180 Placement Stations
Software Version 114.xxx 7.13 EXECUTE NEXT 4 POINTS (Calibration of PCB Vision System)
7 - 27
13. "DETERMINE CAMERA NOZZLE OFFSET" follows automatically:
NOTE
If, when you are about to individually select this menu option, you are no longer sure whether you have in
the meantime exchanged or turned the calibration plate, you should in any case repeat the calibration,
starting with "DETERMINE CALIBRATION FACTORS".
Measuring procedure:
The calibration plate is picked up, centered, turned and deposited. The distance between the nozzle
center and camera center is measured.
The values determined including their algebraic signs are automatically displayed and saved:
CAMERA NOZZLE OFFSET
CAMERA OFFSET OLD X,Y ....... .......
CAMERA OFFSET NEW X,Y ....... .......
CAMERA NOZZLE OFFSET COMPLETED
14. "DETERMINE JAW OFFSET" follows automatically:
The calibration plate is picked up, centered and put down, several times.
The values determined are displayed and automatically transferred to the main memory.
JAW OFFSET
Z-JAW CORRECTION OLD : -.05 H-JAW CORRECTION OLD : -.07
Z-JAW CORRECTION NEW: -.05 H-JAW CORRECTION NEW: -.08
JAW OFFSET COMPLETED
15. Check the new jaws correction value displayed.
NOTE
The jaws correction value is displayed in mm (incl. sign).
The max. acceptable Z and/or H-jaws correction value is ± 0.20 mm.
If the Z or H-jaws offset value is larger than ± 0.20 mm: adjust the symmetry of the jaws (see Service
Manual or adjusting instructions for jaws) repeat the camera calibration, starting with "DETERMINE
CALIBRATION FACTORS".
16. After completion of all four measuring routines, all determined measurement results are displayed.
Confirm the display with "RETURN".
17. The calibration plate is deposited, the PCB conveyed to the output conveyor and the main menu displayed.
Remove the PCB, together with the calibration plate, from the output conveyor.
18. From the main menu, select "STORING". The values are transferred to the work diskette.
7 AutoCal Measuring Sequence: Routine Checks on HS-180 Placement Stations User’s Manual AutoCal Program
7.13 EXECUTE NEXT 4 POINTS (Calibration of PCB Vision System) Software Version 114.xxx
7 - 28
7.14 COMPONENT CAMERA MENU
For the procedure, note:
First
determine the "COMPONENT CAMERA HEIGHT"
and then
"EXECUTE THE FOLLOWING POINTS".
Correct calibration factors are a
prerequisite
for the further calibration (remember this when selecting the
menu options individually).
After the measuring routine "DETERMINE CALIBRATION FACTORS" and acceptance of the new calibra-
tion factors (= after successful measurement!) the menu options "DETERMINE CAMERA CENTRE" and
"DETERMINE ROTATION ANGLE" on the component camera menu must
also
be carried out in any case!
Moreover, when
individually selecting
the menu options, note:
The calibration plate must not be
exchanged or turned, not even
180 degrees,
between the individual
measuring procedures.
Testing aids required
Steel nozzle in perfect condition, type 104, item no. 00301435
Vacuum nozzle adapter, type 117, item no. 00300659
VISION 2 calibration plate, item no. 00132070
Teach panel (external), item no. 00132207
7.14.1 Procedure: "COMPONENT CAMERA HEIGHT"
NOTE
With machines up to and including
batch identity no. 3
, width adjustment is
now
to be performed using
the handwheel. However, this is not necessary if you only wish to carry out the measuring routine "COM-
PONENT CAMERA HEIGHT" as no PCB needs to be placed into the conveyor in this case.
1.
From the main menu, select the
"COMPONENT CAMERA MENU"
.
The "COMPONENT CAMERA MENU" (see Fig. 1.4) is displayed on the screen.
Now, if you have not already done so in the course of individual checks, install a steel nozzle in perfect
condition (normal size) not a
pointed
nozzle!
2.
First select
"CENTERING CAMERA HEIGHT"
. The following display appears on the screen:
’TEACH’ PLACEMENT HEAD OVER COMPONENT CAMERA
TEACHING IS ON, ’END’ TO FINISH
NOTE
In addition the current ACTUAL position of the placement head in X, Y and Z-directions is displayed.
3.
Teach the nozzle in X/Y-direction approximately in the
center
above the glass disk of the component vision
module and then press "END" on the teach panel.
The Z-axis is lowered by means of the sensor stop.
User’s Manual AutoCal Program 7 AutoCal Measuring Sequence: Routine Checks on HS-180 Placement Stations
Software Version 114.xxx 7.14 COMPONENT CAMERA MENU
7 - 29