Maintenanc e Lubricating the flat panel detector positioning unit 195 Inspection system X7056 | Operating manual | Version 3.1 Rev.006| 2016-01-06 | 30.009.1930a Lubricating the spindles Perform the following work steps:…

100%1 / 280
Maintenance
Lubricating the flat panel detector positioning unit
194
Inspection system X7056 | Operating manual |
Version 3.1 Rev.006| 2016-01-06 | 30.009.1930a
4. Grease the two front runner blocks on the Y-axis with 0.4 cm
3
grease.
5. Grease the two rear runner blocks on the Y-axis with 0.4 cm
3
grease.
6. Move the runner blocks back and forth with three double
strokes; the distance should be equal to at least three times
the length of a runner block.
7. Check if there is a visible grease layer on the guide rails.
The Y-axis guide carriages are greased.
Maintenance
Lubricating the flat panel detector positioning unit
195
Inspection system X7056 | Operating manual |
Version 3.1 Rev.006| 2016-01-06 | 30.009.1930a
Lubricating the
spindles
Perform the following work steps:
1. Stand behind the inspection system.
2. Pull the X-axis toward yourself.
3. Check the spindles for contamination and clean as required.
4. Lubricate the Y-axis spindle through the lubricating nipple
attached to the spindle nut with 1.5 cm
3
grease Use the exten-
sion attachment as needed
5. Lubricate the X-axis spindle through the lubricating nipple
attached to the side of the FPD carrier on the X-axis with
1.5 cm
3
grease
6. Check if there is a visible grease film on the spindles and
guides.
The spindles are lubricated.
The runner blocks and spindles of the flat panel detector posi-
tioning unit are lubricated.
Maintenance
Lubricating the Z-axis
196
Inspection system X7056 | Operating manual |
Version 3.1 Rev.006| 2016-01-06 | 30.009.1930a
Lubricating the Z-axis
This section describes the lubrication of spindles of the z-axis.
The following sections describe how to lubricate the spindles:
Z-Axis
Lubricate z-axis
Z-Axis
Z-axis
Lubricate z-axis
Required: Lubricating grease KP2K, NLGI class 2 according to DIN 51818
Brush
Legend
Name
Spindle, Z-axis (sealed X-ray tube)