Packing
Packing the inspection system
247
Inspection system X7056 | Operating manual |
Version 3.1 Rev.006| 2016-01-06 | 30.009.1930a
Packing the
inspection
system airtight
Perform the following work steps:
1. Completely cover the inspection system with the aluminum
laminate bag.
2. Heat-seal the aluminum laminate bag to the bottom layer of
aluminum laminate on the wooden pallet.
Leave an opening of approx. 5 cm for pumping out the air.
3. Insert the vacuum cleaner hose into the opening and remove
the air from the foil laminate bag.
The air is completely pumped out.
4. Remove the vacuum cleaner hose.
5. Heat-seal the opening with the foil sealer.
Packing
Packing the inspection system
248
Inspection system X7056 | Operating manual |
Version 3.1 Rev.006| 2016-01-06 | 30.009.1930a
6. Affix the transport indicators, see „Transport indicators“ (Page
59).
7. Completely wrap the system with two layers of stretch wrap.
The inspection system is wrapped airtight.
Completed. The inspection system is wrapped.
Packing the inspection system in the transport case
For overseas transport, Viscom recommends having the inspection
system packed in a suitable transport case by an industrial packing
company.
NOTE
To prevent improper lifting and transporting, mark the center of
gravity on the transport case according to the center of gravity
sketches, see „Center of Mass“ (Page 260).
Transporting the inspection system
Before beginning to transport the system, read and understand the
accompanying transport manual.
Packing
Storage
249
Inspection system X7056 | Operating manual |
Version 3.1 Rev.006| 2016-01-06 | 30.009.1930a
Storage
For storage, it is important to ensure that the inspection system
and its accessories are protected from potential damage.
NOTE
Always store the system in its original packaging until it is installed
and observe the storage requirements.
Ensure appropriate protection to prevent damage due to the
following factors:
Rain, humidity
Temperature fluctuations that may cause condensation
Dust
Aggressive and soiling environments which may lead to corro-
sion
Mechanical damage or unauthorized opening of the packaging
Strong vibration (loosening of attached parts, fatigue damage,
etc.)
If the X-ray tube is not used for a longer time, a warmup
procedure has to be performed at least every 3 months to
guarantee function, see „Maintenance“ > „Warm-up procedure“
(Page 220).
Storage conditions
Permissible weight load for the
ground surface
950kg/m
2
Permissible weight load for the
ground surface (per system foot)
39 N/cm
2
Ambient temperature 0°C to 70°C
Humidity to 80%, not condensed
Average humidity <60 %