X7056_Operating manual_en.pdf - 第196页

Maintena nce Lubricating the Z-axis 196 Inspection system X7056 | Operating manual | Version 3.1 Rev.006| 2016-01-06 | 30.009.1930a Lubricating the Z-axis This section describes the lubrication of spindl es of the z-axis…

100%1 / 280
Maintenance
Lubricating the flat panel detector positioning unit
195
Inspection system X7056 | Operating manual |
Version 3.1 Rev.006| 2016-01-06 | 30.009.1930a
Lubricating the
spindles
Perform the following work steps:
1. Stand behind the inspection system.
2. Pull the X-axis toward yourself.
3. Check the spindles for contamination and clean as required.
4. Lubricate the Y-axis spindle through the lubricating nipple
attached to the spindle nut with 1.5 cm
3
grease Use the exten-
sion attachment as needed
5. Lubricate the X-axis spindle through the lubricating nipple
attached to the side of the FPD carrier on the X-axis with
1.5 cm
3
grease
6. Check if there is a visible grease film on the spindles and
guides.
The spindles are lubricated.
The runner blocks and spindles of the flat panel detector posi-
tioning unit are lubricated.
Maintenance
Lubricating the Z-axis
196
Inspection system X7056 | Operating manual |
Version 3.1 Rev.006| 2016-01-06 | 30.009.1930a
Lubricating the Z-axis
This section describes the lubrication of spindles of the z-axis.
The following sections describe how to lubricate the spindles:
Z-Axis
Lubricate z-axis
Z-Axis
Z-axis
Lubricate z-axis
Required: Lubricating grease KP2K, NLGI class 2 according to DIN 51818
Brush
Legend
Name
Spindle, Z-axis (sealed X-ray tube)
Maintenance
Lubricating the Z-axis
197
Inspection system X7056 | Operating manual |
Version 3.1 Rev.006| 2016-01-06 | 30.009.1930a
WARNING
Hazard of injury
Set the inspection system in the STOP mode before starting
lubrication, see „Operation“ > „Actuating the STOP mode“
(Page 149).
Perform the following work steps:
1. Open the rear swing doors.
The STOP mode is actuated, see „Operation“ > „Actuating the
STOP mode“ (Page 149).
2. Check the spindle for contamination and clean as required.
3. Give the lower spindle threads a thin coat of grease with the
brush (starting state: Z-axis is above, shutter is closed).
4. Close the rear swing doors and cancel the STOP mode, see
„Operation“ > „Ending the STOP mode“ (Page 150).
5. Use the software to move the Z-axis downwards.
6. Repeat steps 1-4 for the upper spindle threads.
7. Give the upper spindle threads a thin coat of grease with the
brush.
8. Use the software to move the Z-axis.
9. Check if there is a visible grease film on the spindle.
The Z-axis of the sealed X-ray tube is lubricated.