FlexTRAK Series 2MB Mater ial Handler Adden dum Installation © 2023 Nordson Corporation 3-5 3.5 Placing the System Once the FlexTRAK-2MB has been unpacked, it must b e moved to its location at the fa cility. WARNING! Haz…

FlexTRAK Series 2MB Material Handler Addendum Installation
3-4 © 2023 Nordson Corporation
3.4.2
33B33B
Storing the System
If the FlexTRAK-2MB System is to be placed in long-term storage, place all system components in
protective packaging. A desiccant should be placed in the packaging to minimize moisture exposure.
Store components in a room with humidity of less than 80 percent.
To pack for long-term storage:
1. Fill the pump reservoir with oil to the proper level.
2. Run the pump for five minutes to lubricate the seals.
3. Remove all water from the cooling system.
4. Purge all gas lines and the chamber.
DANGER! Certain process gases or liquids selected for use with this
equipment can cause fire or explosion. Do not expose equipment
to any ignition sources. Consult your safety officer and the
manufacturer’s SDS to ensure proper precautionary steps are
taken.
WARNING! Certain process gases selected for use with this equipment are
toxic and require special precautions. Consult your safety
officer and the manufacturer’s SDS to ensure proper
precautionary steps are taken.
5. Place the system components and the vacuum pump in protective packaging.
„ A desiccant should be placed in the packaging to minimize moisture exposure.
6. Store components in a room with a humidity level of less than 80 percent.
?
NOTE During the time the pump is stored, also run it for five minutes every three months to
keep the seals lubricated.

FlexTRAK Series 2MB Material Handler Addendum Installation
© 2023 Nordson Corporation 3-5
3.5 Placing the System
Once the FlexTRAK-2MB has been unpacked, it must be moved to its location at the facility.
WARNING! Hazardous voltages are found within the lower enclosure of the
FlexTRAK-2MB System. Use extreme caution to prevent injury to personnel
or damage to equipment. Only qualified technicians should install this
equipment.
WARNING! Only qualified technicians should perform system installation. It
is recommended that at least two persons guide the system while it is being
positioned on the factory floor.
WARNING! When a chiller is hooked up to the FlexTRAK, a secondary
water containment system, such as a drip pan, should be installed underneath
the FlexTRAK in order to comply with SEMI-S2 safety guidelines. This
system should include leak detection and, if possible, should have a large
enough capacity to hold all of the water used in the cooling system.
To place the system in its factory location:
1. Carefully transport the system to the installation site using a forklift or floor jack.
?
NOTE To move the system short distances, you can use the wheels attached to the bottom of the
plasma system. Then raise or lower the levelers to the desired height using a wrench.
2. Once the system has been positioned in the desired location, slowly lower the forklift until
the chamber rails are approximately at the desired height.
3. Unscrew each leveler (footpad) until it touches the floor.
4. Lower and remove the forklift (Figure 3-2).
5. Place a level across the top of the system to make sure that it is level front-to-back and side-
to-side.
6. Use an open-end wrench to turn the height adjustment nut in the desired direction to raise or
lower the system to the correct position (Figure 3-2).

FlexTRAK Series 2MB Material Handler Addendum Installation
3-6 © 2023 Nordson Corporation
Item Description
1 Height Adjustment Locking Nut (7/8 inch open-end wrench)
2 Leveler (Footpad)
3 Wheel Assembly
Figure 3-2 Leveling the System
7. Align Lane 1 by sliding a workpiece from the upstream machine's conveyor to the
FlexTRAK-2MB conveyor and then from the FlexTRAK-2MB conveyor to the downstream
machine’s conveyor.
?
NOTE Use an open-end wrench to turn the height adjustment nut in the desired
direction to maintain levelness throughout the system.
8. Align Lane 2.
„ This is a teachable setting within the FlexTRAK-2MB system software.
„ Enter into the teach mode to align and teach this lane with the upstream and
downstream machines, see 3.7 Initial Start-up Procedure.
9. If desired, install seismic brackets on each leveler or use other methods to anchor the
system in place.
1
3
2