S2-9XXX Series Dispensing System IOM Manual Maintenance
© 2023 Nordson Corporation 6-9
6.8 Computer Battery Reset
To extend the life of the battery, it is recommended that you perform the following procedure every six
months.
To reset the battery:
1. Click
Start > Control Panel on the Windows Taskbar.
The Windows Control Panel opens.
2. Select
Power Options > Power Manager > Launch Power Manager > Battery
Information > Battery Maintenance
.
The Battery Maintenance window opens (Figure 6-4).
3. Verify the settings are as shown in Figure 6-4.
Figure 6-4 Battery Maintenance Window

S2-9XXX Series Dispensing System IOM Manual Maintenance
6-10 © 2023 Nordson Corporation
4. Click on Perform Reset.
The Battery Gauge Reset window opens (Figure 6-5).
Figure 6-5 Battery Gauge Reset Window
5. Click on Continue.
The process may take several hours to complete. You should avoid using the computer
during this period.

S2-9XXX Series Dispensing System IOM Manual Maintenance
© 2023 Nordson Corporation 6-11
6.9 Removing and Installing the Axis Covers
In order to lubricate the cables and linear guides and to tension the cables, it will be necessary to remove
the X-axis, Y-axis, and rear cable covers.
Tools and Materials Needed:
• Hex Wrench Set
WARNING! Ensure the dispensing system has been completely shutdown before attempting to
remove any panel.
6.9.1 X-Axis Cable Cover
To remove the X-axis covers (Figure 6-6):
1. Perform a service shutdown, see 2.14 Service Shutdown.
2. Open the dispensing area door.
3. Remove the four (4) screws securing the X-axis cable cover to the X-axis.
4. Remove the X-axis cable cover.
Item Description
1 X-Axis Cable Cover
2 Screws (4)
Figure 6-6 Removing the X-Axis Covers
To install the X-axis cable cover (Figure 6-1):
1. Install the four (4) screws securing the X-axis cable cover to the X-axis.
2. Torque the four (4) screws to 2.8 Nm (25 in-lbs).
3. Close the dispensing area door.